510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Raystation

K130617510(k) TraditionalCleared May 10, 201364-day reviewClass II21 CFR 892.5050Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Raysearch Laboratories AB · MUJ System, Planning, Radiation Therapy Treatment

Cleared on May 10, 2013 as substantially equivalent to K120387 (Raysearch Laboratories AB, 2012) and 2 other predicates after a 64-day review. Its family tree reaches back 7 generations to 1987. 21 later devices descend from it. It has been recalled 6 times.

Decided
May 10, 2013
Review
64 days
Class
Class II
Product code
MUJ
Recalls
6
Predicates
3

Its longest line of descent runs back 7 generations to K862643 (1987). 48 ancestors in all, the oldest K802809 (1981). 21 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2000: 3 other ancestors2011: 1 other ancestor2001: 5 other ancestors2008: 2 other ancestors1999: 4 other ancestors1996: 5 other ancestors2006: 2 other ancestors2005: 1 other ancestor2004: 2 other ancestors1992: 4 other ancestors1997: 1 other ancestor1991: 1 other ancestor1990: 2 other ancestors1989: 2 other ancestors1986: 1 other ancestor1981: 1 other ancestor1994: 2 other ancestors2015: 1 descendant cleared2014: 2 descendants cleared2018: 2 descendants cleared2016: 1 descendant cleared2017: 3 descendants cleared2019: 1 descendant cleared2020: 1 descendant cleared2021: 3 descendants cleared2022: 2 descendants cleared2023: 2 descendants cleared2026: 1 descendant cleared2025: 2 descendants clearedK862643 · Radiation Oncology Computer Sys Treatment Planning · Pacific Physics, Inc. · 1987K970236 · Theraplan Plus Version 1.2 · Theratronics, Inc. · 1997K031349 · OTP 1.2 · Nucletron Corp. · 2003K081281 · Ocentra Masterplan 3.1 (Ocentra Brachy) · Nucletron Corporation · 2008 · recalledK083264 · Rayautoplan 1.0 · Raysearch Laboratories AB · 2009 · recalledK100552 · Raystation Version 1.0 · Raysearch Laboratories AB · 2010 · recalledK120387 · Raystation · Raysearch Laboratories AB · 2012 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 6 other ancestors4 generations back: 11 other ancestors5 generations back: 10 other ancestors6 generations back: 9 other ancestors 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 2 descendants6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 2 descendants9 generations forward: 2 descendants10 generations forward: 2 descendants11 generations forward: 2 descendants12 generations forward: 1 descendant13 generations forward: 1 descendant14 generations forward: 1 descendant K120387 · Raystation · Raysearch Laboratories AB · 2012 · recalledK100552 · Raystation Version 1.0 · Raysearch Laboratories AB · 2010 · recalledK083264 · Rayautoplan 1.0 · Raysearch Laboratories AB · 2009 · recalledK081281 · Ocentra Masterplan 3.1 (Ocentra Brachy) · Nucletron Corporation · 2008 · recalledK031349 · OTP 1.2 · Nucletron Corp. · 2003K970236 · Theraplan Plus Version 1.2 · Theratronics, Inc. · 1997K862643 · Radiation Oncology Computer Sys Treatment Planning · Pacific Physics, Inc. · 1987−50+5+10
  1. Root · 7 generations backK862643 Radiation Oncology Computer Sys Treatment PlanningPacific Physics, Inc. · 1987
  2. 6 generations backK970236 Theraplan Plus Version 1.2Theratronics, Inc. · 1997
  3. 5 generations backK031349 OTP 1.2Nucletron Corp. · 2003
  4. 4 generations backK081281 Ocentra Masterplan 3.1 (Ocentra Brachy)Nucletron Corporation · 2008 · Recalled
  5. 3 generations backK083264 Rayautoplan 1.0Raysearch Laboratories AB · 2009 · Recalled
  6. 2 generations backK100552 Raystation Version 1.0Raysearch Laboratories AB · 2010 · Recalled
  7. 1 generation backK120387 RaystationRaysearch Laboratories AB · 2012 · Recalled
  8. This deviceK130617 RaystationRaysearch Laboratories AB · 2013
  9. Descendant · cites it directly · 19 of its ownK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
  10. Descendant · 2 generations on · 17 of its ownK141860 RaystationRaysearch Laboratories AB · 2014 · Recalled
  11. Descendant · 3 generations on · 16 of its ownK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016 · Recalled
  12. Descendant · 4 generations on · 15 of its ownK170355 RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · Recalled
  13. Descendant · 5 generations on · 13 of its ownK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2014: 22015: 12016: 12017: 32018: 22019: 12020: 12021: 32022: 22023: 22025: 22026: 1
2014high 3 (2017) · low 0 (2024)2026

Most influential descendants

Showing 5 of 21 · See all
  • K140187Ray StationRaysearch Laboratories AB · 2014 · cites it directly19descendants
  • K141860RaystationRaysearch Laboratories AB · 2014 · 2 generations on17descendants
  • K160093RayStationRaySearch Laboratories AB (PUBL) · 2016 · 3 generations on16descendants
  • K170355RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · 4 generations on15descendants
  • K171536RayStationRaySearch Laboratories AB (PUBL) · 2017 · 5 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K862643Radiation Oncology Computer Sys Treatment PlanningPacific Physics, Inc.1987—
−6K970236Theraplan Plus Version 1.2Theratronics, Inc.1997—
−5K031349OTP 1.2Nucletron Corp.2003—
−4K081281Ocentra Masterplan 3.1 (Ocentra Brachy)Nucletron Corporation20082
−3K083264Rayautoplan 1.0Raysearch Laboratories AB20092
−2K100552Raystation Version 1.0Raysearch Laboratories AB20104
−1K120387RaystationRaysearch Laboratories AB20126
0 (this device)K130617RaystationRaysearch Laboratories AB2013—
+1K140187Ray StationRaysearch Laboratories AB20141
+2K141860RaystationRaysearch Laboratories AB20145
+3K160093RayStationRaySearch Laboratories AB (PUBL)20167
+4K170355RayStation 6RaySearch Laboratories AB (PUBL)20171
+5K171536RayStationRaySearch Laboratories AB (PUBL)20176

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198920
199020
199110
199240
199420
199650
199710
199940
200030
200150
200220
200420
200510
200620
200820
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201510
201610
201730
201820
201910
202010
202130
202220
202320
202520
202610
FIG. 1 — Lineage of K130617, 1987–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120387Raystation
Summary page 1
…on name Medical charged-particle radiation therapy system (21 CER 892,5050, Product Code MtJJ) 5.7 Predicate devices RaySlation 2.5 K120387 DCM 2.0 K020971 Proton Vision 7.0 K000922 5.8 Device description…
Raysearch Laboratories AB20126
*K020971DCM 2.0
Summary page 1
…n therapy system (21 CER 892,5050, Product Code MtJJ) 5.7 Predicate devices RaySlation 2.5 K120387 DCM 2.0 K020971 Proton Vision 7.0 K000922 5.8 Device description RayStation 3.5 is a treatment planning systet…
Mds Nordion AB2002—
*K000922Proton Vision
Summary page 1
…tJJ) 5.7 Predicate devices RaySlation 2.5 K120387 DCM 2.0 K020971 Proton Vision 7.0 K000922 5.8 Device description RayStation 3.5 is a treatment planning systetn. i.c. a softwvare program for planning and anal…
Varian Assoc., Inc.2000—

What descends from it? (21)

See all 21

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140187Ray StationRaysearch Laboratories AB201419
K150547.decimal Astroid Dosimetry App.Decimal, Inc.20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 8, 2015Class IIZ-1679-2015Software Design ChangeRaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
Apr 25, 2014Class IIZ-1806-2014Software designThis notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such…
Feb 14, 2014Class IIZ-1481-2014Software designRaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.
Feb 6, 2014Class IIZ-1308-2014Software Design ChangeRaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy.
Feb 6, 2014Class IIZ-1307-2014Software Design ChangePotential for dose errors due to software program errors.
Oct 23, 2013Class IIZ-0281-2014Software designRay Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 2.1 years on median.
  • 9 of the 48 ancestors have a recall of any class.
  • Of its descendants, K140187 (Raysearch Laboratories AB, 2014) has gone furthest, with 19 descendants of its own.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130617: Raystation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130617

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026