510(k) · Cleared
Recalled · Class II · 2025Raystation
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Raysearch Laboratories AB · MUJ System, Planning, Radiation Therapy Treatment
Cleared on Oct 23, 2014 as substantially equivalent to K140187 (Raysearch Laboratories AB, 2014) after a 105-day review. Its family tree reaches back 9 generations to 2001. 17 later devices descend from it. It has been recalled 5 times.
- Decided
- Oct 23, 2014
- Review
- 105 days
- Class
- Class II
- Product code
- MUJ
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 9 generations to K010975 (2001). 64 ancestors in all, the oldest K802809 (1981). 17 later devices descend from it, 1 citing it directly.
- Root · 9 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
- 5 generations between
- 3 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
- 2 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
- 1 generation backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
- This deviceK141860 RaystationRaysearch Laboratories AB · 2014
- Descendant · cites it directly · 16 of its ownK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016 · Recalled
- Descendant · 2 generations on · 15 of its ownK170355 RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · Recalled
- Descendant · 3 generations on · 13 of its ownK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled
- Descendant · 4 generations on · 12 of its ownK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
- Descendant · 5 generations on · 11 of its ownK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
Show all 9 generations
- Root · 9 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
- 8 generations backK030981 EclipseVarian Medical Systems, Inc. · 2003
- 7 generations backK041403 Eclipse with AaaVarian Medical Systems · 2004
- 6 generations backK050296 Eclipse Treatment Planning SystemVarian Medical Systems · 2005
- 5 generations backK071873 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
- 4 generations backK073020 Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
- 3 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
- 2 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
- 1 generation backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
- This deviceK141860 RaystationRaysearch Laboratories AB · 2014
Showing 5 of 17 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 17 · See all- K160093RayStationRaySearch Laboratories AB (PUBL) · 2016 · cites it directly16descendants
- K170355RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · 2 generations on15descendants
- K171536RayStationRaySearch Laboratories AB (PUBL) · 2017 · 3 generations on13descendants
- K180379RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · 4 generations on12descendants
- K190387RayStationRaySearch Laboratories AB (PUBL) · 2019 · 5 generations on11descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K010975 | Eclipse 6.5 | Varian Medical Systems, Inc. | 2001 | 3 |
| −8 | K030981 | Eclipse | Varian Medical Systems, Inc. | 2003 | — |
| −7 | K041403 | Eclipse with Aaa | Varian Medical Systems | 2004 | — |
| −6 | K050296 | Eclipse Treatment Planning System | Varian Medical Systems | 2005 | — |
| −5 | K071873 | Modification to Eclipse Treatment Planning System | Varian Medical Systems, Inc. | 2007 | 5 |
| −4 | K073020 | Eclipse Treatment Planning System | Varian Medical Systems, Inc. | 2007 | 6 |
| −3 | K091492 | Modification to Eclipse Treatment Planning System | Varian Medical Systems, Inc. | 2009 | 2 |
| −2 | K102216 | Xio RTP System | Computerized Medical Systems, Inc. | 2010 | 7 |
| −1 | K140187 | Ray Station | Raysearch Laboratories AB | 2014 | 1 |
| 0 (this device) | K141860 | Raystation | Raysearch Laboratories AB | 2014 | — |
| +1 | K160093 | RayStation | RaySearch Laboratories AB (PUBL) | 2016 | 7 |
| +2 | K170355 | RayStation 6 | RaySearch Laboratories AB (PUBL) | 2017 | 1 |
| +3 | K171536 | RayStation | RaySearch Laboratories AB (PUBL) | 2017 | 6 |
| +4 | K180379 | RayStation 7 | RaySearch Laboratories AB (PUBL) | 2018 | 7 |
| +5 | K190387 | RayStation | RaySearch Laboratories AB (PUBL) | 2019 | 32 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1986 | 1 | 0 |
| 1987 | 1 | 0 |
| 1989 | 2 | 0 |
| 1990 | 2 | 0 |
| 1991 | 1 | 0 |
| 1992 | 4 | 0 |
| 1994 | 2 | 0 |
| 1996 | 5 | 0 |
| 1997 | 2 | 0 |
| 1999 | 4 | 0 |
| 2000 | 5 | 0 |
| 2001 | 5 | 0 |
| 2002 | 3 | 0 |
| 2003 | 2 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
| 2006 | 2 | 0 |
| 2007 | 1 | 0 |
| 2008 | 3 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 1 | 0 |
| 2017 | 3 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 3 | 0 |
| 2022 | 2 | 0 |
| 2023 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MUJ device
- #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K140187 | Ray StationSummary page 4…assification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) 7.7 Predicate device RayStation 4.0.2 K140187 7.8 Device description RayStation 4.5 is a radiation therapy treatment planning system, i.e. a software program for plan… | 2014 |
What descends from it? (17)
See all 171 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K160093 | RayStation | 2016 | 16 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 16, 2025 | Class II | Z-1195-2025 | Under Investigation by firm | Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in… |
| Feb 10, 2016 | Class II | Z-1712-2016 | Other | A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right… |
| May 29, 2015 | Class II | Z-2785-2015 | Software design | An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images |
| May 27, 2015 | Class II | Z-2457-2015 | Software design | Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction… |
| Mar 4, 2015 | Class II | Z-1310-2015 | Under Investigation by firm | An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
- 19 of the 64 ancestors have a recall of any class.
- Of its descendants, K160093 (RaySearch Laboratories AB (PUBL), 2016) has gone furthest, with 16 descendants of its own.
- 1 other device also cites its predicate K140187.
How does its category compare?
All MUJ devicesMUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · MUJ
10in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MUJ
186in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MUJ
2in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MUJ
21in 2025
MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 67 | 4 | 24 |
| 2008 | 9 | 63 | 10 | 51 |
| 2009 | 16 | 84 | 11 | 24 |
| 2010 | 11 | 56 | 23 | 78 |
| 2011 | 9 | 103 | 17 | 73 |
| 2012 | 8 | 59 | 4 | 39 |
| 2013 | 5 | 64 | 6 | 33 |
| 2014 | 9 | 108 | 2 | 35 |
| 2015 | 13 | 78 | 7 | 30 |
| 2016 | 5 | 80 | 13 | 27 |
| 2017 | 7 | 30 | 4 | 16 |
| 2018 | 7 | 53 | 3 | 25 |
| 2019 | 9 | 56 | 4 | 25 |
| 2020 | 9 | 104 | 2 | 26 |
| 2021 | 11 | 87 | 4 | 15 |
| 2022 | 8 | 168 | 3 | 22 |
| 2023 | 10 | 140 | 2 | 76 |
| 2024 | 9 | 122 | 0 | 9 |
| 2025 | 10 | 186 | 0 | 21 |
| 2026 | 9 | 233 | 2 | — |
What is it indicated for?
RayStation is a software system designed for treatment planning and analysis of
radiation therapy. The treatment plans provide treatment unit set-up parameters and
estimates of dose distributions expected during the proposed treatment, and may be
used to administer treatments after review and approval by the intended user.
The system functionality can be configured based on user needs.
The intended users of RayStation shall be clinically qualified radiation therapy staff
trained in using the system.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K141860 · printed Nov 6, 2026
Cite this page
510(k) Report. “K141860: Raystation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141860
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026