RayStation

K160093510(k) TraditionalCleared Apr 8, 201680-day reviewClass II21 CFR 892.5050Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

RaySearch Laboratories AB (PUBL) · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Apr 8, 2016 as substantially equivalent to K141860 (Raysearch Laboratories AB, 2014) after an 80-day review. Its family tree reaches back 10 generations to 2001. 16 later devices descend from it. It has been recalled 7 times.

Decided
Apr 8, 2016
Review
80 days
Class
Class II
Product code
MUJ
Recalls
7
Predicates
1

Its longest line of descent runs back 10 generations to K010975 (2001). 65 ancestors in all, the oldest K802809 (1981). 16 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2012: 1 other ancestor2009: 2 other ancestors2002: 3 other ancestors2000: 5 other ancestors2011: 1 other ancestor2010: 1 other ancestor2008: 3 other ancestors2007: 1 other ancestor2003: 2 other ancestors2001: 5 other ancestors1999: 4 other ancestors1997: 2 other ancestors1996: 5 other ancestors2006: 2 other ancestors2005: 1 other ancestor2004: 2 other ancestors1992: 4 other ancestors1991: 1 other ancestor1994: 2 other ancestors1990: 2 other ancestors1989: 2 other ancestors1987: 1 other ancestor1986: 1 other ancestor1981: 1 other ancestor2017: 3 descendants cleared2018: 1 descendant cleared2019: 1 descendant cleared2020: 1 descendant cleared2021: 3 descendants cleared2022: 2 descendants cleared2023: 2 descendants cleared2026: 1 descendant cleared2025: 2 descendants clearedK010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalledK030981 · Eclipse · Varian Medical Systems, Inc. · 2003K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005×2: 2 generations cleared close together, K071873 (2007), K073020 (2007) K091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalled×2: 2 generations cleared close together, K140187 (2014), K141860 (2014) 19801990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 4 other ancestors5 generations back: 7 other ancestors6 generations back: 11 other ancestors7 generations back: 14 other ancestors8 generations back: 14 other ancestors9 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 2 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants8 generations forward: 2 descendants9 generations forward: 1 descendant10 generations forward: 1 descendant11 generations forward: 1 descendant K141860 · Raystation · Raysearch Laboratories AB · 2014 · recalledK140187 · Ray Station · Raysearch Laboratories AB · 2014 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalledK091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK073020 · Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK071873 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K030981 · Eclipse · Varian Medical Systems, Inc. · 2003K010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalled−10−50+5+10
  1. Root · 10 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 6 generations between
  3. 3 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
  4. 2 generations backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
  5. 1 generation backK141860 RaystationRaysearch Laboratories AB · 2014 · Recalled
  6. This deviceK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016
  7. Descendant · cites it directly · 15 of its ownK170355 RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · Recalled
  8. Descendant · 2 generations on · 13 of its ownK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled
  9. Descendant · 3 generations on · 12 of its ownK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
  10. Descendant · 4 generations on · 11 of its ownK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
  11. Descendant · 5 generations on · 6 of its ownK200569 RayStationRaySearch Laboratories AB (PUBL) · 2020 · Recalled
Show all 10 generations
  1. Root · 10 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 9 generations backK030981 EclipseVarian Medical Systems, Inc. · 2003
  3. 8 generations backK041403 Eclipse with AaaVarian Medical Systems · 2004
  4. 7 generations backK050296 Eclipse Treatment Planning SystemVarian Medical Systems · 2005
  5. 6 generations backK071873 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  6. 5 generations backK073020 Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  7. 4 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
  8. 3 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
  9. 2 generations backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
  10. 1 generation backK141860 RaystationRaysearch Laboratories AB · 2014 · Recalled
  11. This deviceK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2017: 32018: 12019: 12020: 12021: 32022: 22023: 22025: 22026: 1
2017high 3 (2017) · low 0 (2024)2026

Most influential descendants

Showing 5 of 16 · See all
  • K170355RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · cites it directly15descendants
  • K171536RayStationRaySearch Laboratories AB (PUBL) · 2017 · 2 generations on13descendants
  • K180379RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · 3 generations on12descendants
  • K190387RayStationRaySearch Laboratories AB (PUBL) · 2019 · 4 generations on11descendants
  • K200569RayStationRaySearch Laboratories AB (PUBL) · 2020 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K010975Eclipse 6.5Varian Medical Systems, Inc.20013
−9K030981EclipseVarian Medical Systems, Inc.2003—
−8K041403Eclipse with AaaVarian Medical Systems2004—
−7K050296Eclipse Treatment Planning SystemVarian Medical Systems2005—
−6K071873Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20075
−5K073020Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20076
−4K091492Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20092
−3K102216Xio RTP SystemComputerized Medical Systems, Inc.20107
−2K140187Ray StationRaysearch Laboratories AB20141
−1K141860RaystationRaysearch Laboratories AB20145
0 (this device)K160093RayStationRaySearch Laboratories AB (PUBL)2016—
+1K170355RayStation 6RaySearch Laboratories AB (PUBL)20171
+2K171536RayStationRaySearch Laboratories AB (PUBL)20176
+3K180379RayStation 7RaySearch Laboratories AB (PUBL)20187
+4K190387RayStationRaySearch Laboratories AB (PUBL)201932
+5K200569RayStationRaySearch Laboratories AB (PUBL)202014

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198710
198920
199020
199110
199240
199420
199650
199720
199940
200050
200150
200230
200320
200420
200510
200620
200710
200830
200920
201010
201110
201210
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201730
201810
201910
202010
202130
202220
202320
202520
202610
FIG. 1 — Lineage of K160093, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141860Raystation
Summary page 4
…assification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) 5.7 Predicate device RayStation 4.5 K141860 5.8 Device description RayStation 5 is a radiation therapy treatment planning system, i.e. a software program for plann…
Raysearch Laboratories AB20145

What descends from it? (16)

See all 16

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170355RayStation 6RaySearch Laboratories AB (PUBL)201715

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 16, 2025Class IIZ-1194-2025Under Investigation by firmInconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in…
Jan 18, 2017Class IIZ-1815-2017Device DesignAn error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board (graphics card) an…
Nov 9, 2016Class IIZ-0721-2017Software designSoftware anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient…
Oct 11, 2016Class IIZ-0720-2017Software designAn error may occur with the display of dose computed on images other than the planning CT (auxiliary CT) when using multiple patient cases in RayStation 5. If a CT image set with the same Frame of Reference as the displayed auxiliary CT…
Sep 1, 2016Class IIZ-0038-2017Under Investigation by firmRaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when using a region of…
Jul 1, 2016Class IIZ-2206-2016Under Investigation by firmAn issue was found with the evaluation of biological clinical goals in RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an adapted plan based on another…
May 30, 2016Class IIZ-0079-2017Software designFor a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 5 companies; a new generation was cleared every 1.5 years on median.
  • 20 of the 65 ancestors have a recall of any class.
  • Of its descendants, K170355 (RaySearch Laboratories AB (PUBL), 2017) has gone furthest, with 15 descendants of its own.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans

provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and

may be used to administer treatments after review and approval by the intended user.

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160093: RayStation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160093

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026