510(k) Report

510(k) · Cleared

Cleared

Seviin Surface Replacement Shoulder

K130635510(k) TraditionalCleared Aug 26, 2013168-day reviewClass II21 CFR 888.3690

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ingen Orthopedics, LLC · HSD Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented

Cleared on Aug 26, 2013 as substantially equivalent to K010657 (Biomet Manufacturing, Inc., 2001) and 1 other predicate after a 168-day review. Its family tree reaches back 7 generations to 1980.

Decided
Aug 26, 2013
Review
168 days
Class
Class II
Product code
HSD
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K800494 (1980). 24 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2001: 2 other ancestors2006: 1 other ancestor2004: 2 other ancestors2003: 1 other ancestor2000: 4 other ancestors1999: 1 other ancestor1995: 1 other ancestor1996: 1 other ancestor1987: 1 other ancestor1980: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestorK800494 · New Jersey Hemiarthroplasty Sholder Pros · Depuy, Inc. · 1980K991585 · Bipolar Shoulder Prosthesis · Biomet, Inc. · 1999K002998 · Biomet BI-Polar Shoulder System · Biomet, Inc. · 2000 · recalledK010827 · Copeland MB/HA Resurfacing Humeral Heads · Biomet Orthopedics, Inc. · 2001 · recalled×2: 2 generations cleared close together, K031971 (2003), K033516 (2003) K062661 · Aequalis Resurfacing Head · Tornier · 200619801985199019952000200520102015
  1. Root · 7 generations backK800494 New Jersey Hemiarthroplasty Sholder ProsDepuy, Inc. · 1980
  2. 6 generations backK991585 Bipolar Shoulder ProsthesisBiomet, Inc. · 1999
  3. 5 generations backK002998 Biomet BI-Polar Shoulder SystemBiomet, Inc. · 2000 · Recalled
  4. 4 generations backK010827 Copeland MB/HA Resurfacing Humeral HeadsBiomet Orthopedics, Inc. · 2001 · Recalled
  5. 3 generations backK031971 Global Cap Resurfacing Replacement ShoulderDePuy Orthopaedics, Inc. · 2003
  6. 2 generations backK033516 Depuy Global Cap HA Resurfacing Shoulder Humeral HeadsDePuy Orthopaedics, Inc. · 2003
  7. 1 generation backK062661 Aequalis Resurfacing HeadTornier · 2006
  8. This deviceK130635 Seviin Surface Replacement ShoulderIngen Orthopedics, LLC · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K800494New Jersey Hemiarthroplasty Sholder ProsDepuy, Inc.1980—
−6K991585Bipolar Shoulder ProsthesisBiomet, Inc.1999—
−5K002998Biomet BI-Polar Shoulder SystemBiomet, Inc.20001
−4K010827Copeland MB/HA Resurfacing Humeral HeadsBiomet Orthopedics, Inc.20011
−3K031971Global Cap Resurfacing Replacement ShoulderDePuy Orthopaedics, Inc.2003—
−2K033516Depuy Global Cap HA Resurfacing Shoulder Humeral HeadsDePuy Orthopaedics, Inc.2003—
−1K062661Aequalis Resurfacing HeadTornier2006—
0 (this device)K130635Seviin Surface Replacement ShoulderIngen Orthopedics, LLC2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198710
199110
199310
199510
199610
199910
200040
200120
200310
200420
200610
FIG. 1 — Lineage of K130635, 1980–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HSD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HSD device

  1. #1 K191433 AGILON® XO Shoulder Replacement Systemscore 13.8 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010657Copeland MB Resurfacing Humeral Heads
Summary page 1
…d cemented prosthesis; Class 11 Product Codes: FISD, KWVS, KWTr Substantially Equivalent Devices: Copeland MB Resurfacing Humeral Heads, K010657, cleared Sept 14, 2001 Tornier Aequalis Resurfacing H-ead, 1(0…
Biomet Manufacturing, Inc.2001—
*K062661Aequalis Resurfacing Head
Summary page 1
…and MB Resurfacing Humeral Heads, K010657, cleared Sept 14, 2001 Tornier Aequalis Resurfacing H-ead, 1(062661, cleared December 22, 2006 Device Description: The SEVIIN Surface Replacement Shoulder…
Tornier2006—

Also cleared against K010657: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 3 of the 24 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K010657.

How does its category compare?

All HSD devices

HSD Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · HSD

1in 2024

2007: 42008: 32011: 22012: 22013: 42014: 12016: 32018: 12020: 22022: 12024: 1
2007high 4 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HSD

145in 2016

2007: 72 days median2008: 92 days median2013: 218 days median2016: 145 days median
2007high 218 (2013) · low 72 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HSD

1in 2020

2008: 22009: 22017: 32018: 12020: 1
2007high 3 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HSD

757in 2025

2007: 1282008: 2592009: 1562010: 2252011: 2412012: 2542013: 2882014: 4342015: 5742016: 7842017: 9852018: 6492019: 8302020: 6582021: 5732022: 5082023: 5862024: 6932025: 757
2007high 985 (2017) · low 128 (2007)2025

MAUDE reports naming product code HSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074720128
20083922259
20090—2156
20100—0225
20112—0241
20122—0254
201342180288
20141—0434
20150—0574
201631450784
20170—3985
20181—1649
20190—0830
20202—1658
20210—0573
20221—0508
20230—0586
20241—0693
20250—0757
20260—0—

What is it indicated for?

The SEVIIN Surface Replacement Shoulder is indicated for hemi or total shoulder replacement in patients where the humeral head and neck are of sufficient bone stock and there is an intact or reconstructable rotator cuff. Hemi or total shoulder replacement is indicated to relieve severe pain or significant disability caused by: * Rheumatoid arthritis * Non-inflammatory degenerative joint disease (osteoarthritis and avascular necrosis) * Correction of…

Read the full statement (98 words)

The SEVIIN Surface Replacement Shoulder is indicated for hemi or total shoulder replacement in patients where the humeral head and neck are of sufficient bone stock and there is an intact or reconstructable rotator cuff.

Hemi or total shoulder replacement is indicated to relieve severe pain or significant disability caused by:

* Rheumatoid arthritis

* Non-inflammatory degenerative joint disease (osteoarthritis and avascular necrosis)

* Correction of functional deformity

* Humeral head fractures

* Traumatic arthritis

The SEVIIN Resurfacing Humeral Heads are intended for cemented and uncemented applications. The glenoid components, when used, are intended for cemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130635: Seviin Surface Replacement Shoulder.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130635

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026