510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Enflow IV Fluid Warmer Insulated Strap

K130867510(k) SpecialCleared Jun 14, 201378-day reviewClass II21 CFR 880.5725Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vital Signs, Inc., A GE Healthcare Company · LGZ Warmer, Thermal, Infusion Fluid

Cleared on Jun 14, 2013 as substantially equivalent to K121775 (Vital Signs, Inc., A GE Healthcare Company, 2012) after a 78-day review. Its family tree reaches back 6 generations to 1977. 3 of the 9 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times, including 2 Class I recalls.

Decided
Jun 14, 2013
Review
78 days
Class
Class II
Product code
LGZ
Recalls
2 (2 Class I)
Predicates
1

Its longest line of descent runs back 6 generations to K770232 (1977). 8 ancestors in all. No later device cites it as a predicate yet. 3 of the 9 devices in this family, counting this one, have a Class I recall.

  1. Root · 6 generations backK770232 Blood/Fluid Warmer Model DW-1000AGorman Rupp Industries · 1977
  2. 5 generations backK984640 Thermal AngelEstill Medical Technologies, Inc. · 1999 · Recalled
  3. 4 generations backK012031 Modification to Thermal AngelEstill Medical Technologies, Inc. · 2001 · Recalled
  4. 3 generations backK060537 Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model…Enginivity, LLC · 2006 · Recalled (Class I)
  5. 2 generations backK112902 Enflow IV Fluid WarmerVital Signs, Inc. · 2012
  6. 1 generation backK121775 Enflow IV Fluid Warmer, Enflow Disposable Cartridge, Enflow ControllerVital Signs, Inc., A GE Healthcare Company · 2012
  7. This deviceK130867 Enflow IV Fluid Warmer Insulated StrapVital Signs, Inc., A GE Healthcare Company · 2013

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K770232Blood/Fluid Warmer Model DW-1000AGorman Rupp Industries1977—
−5K984640Thermal AngelEstill Medical Technologies, Inc.19991
−4K012031Modification to Thermal AngelEstill Medical Technologies, Inc.20011
−3K060537Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model 200 and…Enginivity, LLC2006Class I
−2K112902Enflow IV Fluid WarmerVital Signs, Inc.2012—
−1K121775Enflow IV Fluid Warmer, Enflow Disposable Cartridge, Enflow ControllerVital Signs, Inc., A GE Healthcare Company2012—
0 (this device)K130867Enflow IV Fluid Warmer Insulated StrapVital Signs, Inc., A GE Healthcare Company2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199811
FIG. 1 — Lineage of K130867, 1977–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LGZ device

  1. #1 K260988 °M Warmer System with °M Stationscore 12.7 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K121775Enflow IV Fluid Warmer, Enflow Disposable Cartridge, Enflow Controller
Summary page 1
…Usual Name: | enFlow, Insulated strap, i-strap Classification Names: | Warmer, Thermal, Infusion fluid Product Code: | LGZ BSB - 21 CFR 864.9205 Predicate Devicels): | K121775, enFlow IV Fluid Warmer Strap 33 GE Healthcare , 510(k) Premarket Notification Submission Device Description:…
Vital Signs, Inc., A GE Healthcare Company2012—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 13, 2019Class IZ-1385-2019Device DesignTesting has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
Mar 13, 2019Class IZ-1386-2019Device DesignTesting has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060537Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model 200 and Eflow AC Power Supply, Model 120Enginivity, LLC · 2006 · 3 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K973741Bair Hugger Blood/Fluid WarmerAugustine Medical, Inc. · 1998 · 5 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 36 years and passes through 5 companies; a new generation was cleared every 4.9 years on median.
  • 5 of the 8 ancestors have a recall of any class; 2 have a Class I recall.

How does its category compare?

All LGZ devices

LGZ Warmer, Thermal, Infusion Fluid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LGZ

2in 2025

2007: 22008: 42009: 22010: 22011: 32012: 72013: 52015: 22016: 22017: 22018: 12019: 32022: 32023: 12024: 12025: 22026 (partial year): 2
2007high 7 (2012) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LGZ

546in 2022

2008: 99 days median2011: 77 days median2012: 162 days median2013: 143 days median2019: 209 days median2022: 546 days median
2007high 546 (2022) · low 77 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LGZ

1in 2016

2009: 12010: 12011: 82015: 12016: 12026 (partial year): 1
2007high 8 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LGZ

376in 2025

2007: 112008: 102009: 122010: 432011: 542012: 522013: 162014: 232015: 342016: 152017: 582018: 942019: 3402020: 1,2212021: 8852022: 1,5482023: 1,0332024: 8622025: 376
2007high 1,548 (2022) · low 10 (2008)2025

MAUDE reports naming product code LGZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—011
2008499010
20092—112
20102—143
2011377854
20127162052
20135143016
20140—023
20152—134
20162—115
20172—058
20181—094
201932090340
20200—01,221
20210—0885
2022354601,548
20231—01,033
20241—0862
20252—0376
20262—1—

What is it indicated for?

The enFlow IV Fluid Warmer is indicated for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130867: Enflow IV Fluid Warmer Insulated Strap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130867

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026