510(k) · Cleared
Recalled · Class II · 2015Firebird Spinal Fixation System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Orthofix, Inc. · NKB Thoracolumbosacral Pedicle Screw System
Cleared on Aug 7, 2013 as substantially equivalent to K081684 (Blackstone Medical, Inc., 2008) and 9 other predicates after a 125-day review. Its family tree reaches back 13 generations to 2004. 81 later devices descend from it. It has been recalled once.
- Decided
- Aug 7, 2013
- Review
- 125 days
- Class
- Class II
- Product code
- NKB
- Recalls
- 1
- Predicates
- 10
Its longest line of descent runs back 13 generations to K040583 (2004). 224 ancestors in all, the oldest K781443 (1978). 81 later devices descend from it, 4 citing it directly.
- Root · 13 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
- 9 generations between
- 3 generations backK112473 CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2011
- 2 generations backK113174 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2011 · Recalled
- 1 generation backK121630 Range Spinal SystemK2m, Inc. · 2012
- This deviceK130932 Firebird Spinal Fixation SystemOrthofix, Inc. · 2013
- Descendant · cites it directly · 78 of its ownK151488 Firebird Spinal Fixation System/Phoenix MIS Spinal Fixation SystemOrthofix, Inc. · 2015 · Recalled
- Descendant · cites it directly · 73 of its ownK153428 Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation SystemOrthofix, Inc. · 2016 · Recalled
- Descendant · 2 generations on · 61 of its ownK171082 Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation System…Orthofix, Inc. · 2017 · Recalled
- Descendant · 3 generations on · 37 of its ownK180179 Firebird® Spinal Fixation Systems: Firebird System; Firebird…Orthofix, Inc. · 2018
- Descendant · 3 generations on · 21 of its ownK173130 Reform® Midline Cortical Screw SystemPrecision Spine, Inc. · 2017
Show all 13 generations
- Root · 13 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
- 12 generations backK041118 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
- 11 generations backK042025 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2004 · Recalled
- 10 generations backK043151 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
- 9 generations backK050809 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2005
- 8 generations backK052187 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
- 7 generations backK053101 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2006
- 6 generations backK091442 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2009
- 5 generations backK091974 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2009 · Recalled
- 4 generations backK102555 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2010
- 3 generations backK112473 CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2011
- 2 generations backK113174 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2011 · Recalled
- 1 generation backK121630 Range Spinal SystemK2m, Inc. · 2012
- This deviceK130932 Firebird Spinal Fixation SystemOrthofix, Inc. · 2013
Showing 5 of 81 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 81 · See all- K151488Firebird Spinal Fixation System/Phoenix MIS Spinal Fixation SystemOrthofix, Inc. · 2015 · cites it directly78descendants
- K153428Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation SystemOrthofix, Inc. · 2016 · cites it directly73descendants
- K171082Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation System, JANUS Midline…Orthofix, Inc. · 2017 · 2 generations on61descendants
- K180179Firebird® Spinal Fixation Systems: Firebird System; Firebird Deformity System; Firebird®…Orthofix, Inc. · 2018 · 3 generations on37descendants
- K173130Reform® Midline Cortical Screw SystemPrecision Spine, Inc. · 2017 · 3 generations on21descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −13 | K040583 | Modification to: CD Horizon Spinal System | Medtronic Sofamor Danek | 2004 | — |
| −12 | K041118 | Modification to CD Horizon Spinal System | Medtronic Sofamor Danek | 2004 | — |
| −11 | K042025 | Modification to CD Horizon Spinal System | Medtronic Sofamor Danek, Inc. | 2004 | 4 |
| −10 | K043151 | Modification to: CD Horizon Spinal System | Medtronic Sofamor Danek | 2005 | — |
| −9 | K050809 | CD Horizon Spinal System | Medtronic Sofamor Danek USA, Inc. | 2005 | — |
| −8 | K052187 | Modification to: CD Horizon Spinal System | Medtronic Sofamor Danek | 2005 | — |
| −7 | K053101 | CD Horizon Spinal System | Medtronic Sofamor Danek | 2006 | — |
| −6 | K091442 | CD Horizon Spinal System | Medtronic Sofamor Danek | 2009 | — |
| −5 | K091974 | Modification to CD Horizon Spinal System | Medtronic Sofamor Danek USA, Inc. | 2009 | 1 |
| −4 | K102555 | CD Horizon Spinal System | Medtronic Sofamor Danek USA, Inc. | 2010 | — |
| −3 | K112473 | CD Horizon Spinal System | Medtronic Sofamor Danek, Inc. | 2011 | — |
| −2 | K113174 | CD Horizon Spinal System | Medtronic Sofamor Danek USA, Inc. | 2011 | 1 |
| −1 | K121630 | Range Spinal System | K2m, Inc. | 2012 | — |
| 0 (this device) | K130932 | Firebird Spinal Fixation System | Orthofix, Inc. | 2013 | — |
| +1 | K151488 | Firebird Spinal Fixation System/Phoenix MIS Spinal Fixation System | Orthofix, Inc. | 2015 | 1 |
| +1 | K153428 | Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation System | Orthofix, Inc. | 2016 | 1 |
| +2 | K171082 | Firebird Spinal Fixation System, Phoenix MIS Spinal Fixation System, JANUS… | Orthofix, Inc. | 2017 | 1 |
| +3 | K180179 | Firebird® Spinal Fixation Systems: Firebird System; Firebird Deformity System… | Orthofix, Inc. | 2018 | — |
| +3 | K173130 | Reform® Midline Cortical Screw System | Precision Spine, Inc. | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1978 | 1 | 0 |
| 1984 | 1 | 0 |
| 1990 | 2 | 0 |
| 1992 | 5 | 0 |
| 1993 | 2 | 0 |
| 1994 | 3 | 0 |
| 1995 | 7 | 0 |
| 1996 | 9 | 0 |
| 1997 | 6 | 0 |
| 1998 | 14 | 0 |
| 1999 | 5 | 0 |
| 2000 | 16 | 0 |
| 2001 | 19 | 0 |
| 2002 | 17 | 0 |
| 2003 | 16 | 0 |
| 2004 | 21 | 0 |
| 2005 | 17 | 0 |
| 2006 | 12 | 0 |
| 2007 | 5 | 0 |
| 2008 | 12 | 0 |
| 2009 | 4 | 0 |
| 2010 | 7 | 0 |
| 2011 | 6 | 0 |
| 2012 | 4 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2015 | 2 | 0 |
| 2016 | 2 | 0 |
| 2017 | 5 | 0 |
| 2018 | 3 | 0 |
| 2019 | 6 | 0 |
| 2020 | 4 | 0 |
| 2021 | 7 | 0 |
| 2022 | 12 | 0 |
| 2023 | 14 | 0 |
| 2024 | 7 | 0 |
| 2025 | 9 | 0 |
| 2026 | 10 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NKB device
- #1 K223494 CD Horizon Spinal Systemscore 40.4 · 30 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (10)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K081684 | Blackstone Pedicle Screw SystemSummary page 3…3 of 4 Predicate Devices: The Firebird Spinal Fixation System is substantially equivalent to the predicate devices: K081684, K082797, K100044, K093926 and Kl22901- Firebird Spinal Fixation System manufactured by Orthofix… | 2008 |
| * | K082797 | Blackstone Pedicle Screw SystemSummary page 3…redicate Devices: The Firebird Spinal Fixation System is substantially equivalent to the predicate devices: K081684, K082797, K100044, K093926 and Kl22901- Firebird Spinal Fixation System manufactured by Orthofix… | 2008 |
| * | K100044 | Firebird Spinal Fixation System, Firebird Side-Loading BodySummary page 3…Devices: The Firebird Spinal Fixation System is substantially equivalent to the predicate devices: K081684, K082797, K100044, K093926 and Kl22901- Firebird Spinal Fixation System manufactured by Orthofix… | 2010 |
| * | K093926 | Firebird Spinal Fixation SystemSummary page 3…The Firebird Spinal Fixation System is substantially equivalent to the predicate devices: K081684, K082797, K100044, K093926 and Kl22901- Firebird Spinal Fixation System manufactured by Orthofix Inc… | 2010 |
| * | K122901 | Firebird Spinal Fixation System, Phoenix Mis Spinal Fixation System | 2012 |
| * | K113666 | Xia 3 Spinal System | 2012 |
| * | K091445 | CD Horizon Spinal SystemSummary page 3…Inc. KI 13666-XIA Ill Spinal System manufactured by Stryker Spine K09 1445-CD HORIZON Spinal System manufactured by Medtronic Sofainor Danek USA K9… | 2010 |
| * | K994121 | Small Stature UssSummary page 3…Fixation System are substantially equivalent to the XIA Ill (KI 13666), CD Horizon Spinal System (K091445). USS Small Stature (K994121), Range Spinal System (K121630), and the TSR- Spinal System (KI 11492). The features of the subject… | 2000 |
| * | K121630 | Range Spinal SystemSummary page 3…tronic Sofainor Danek USA K9941 2l1-USS Small Stature manufactured by Synthes Spine K 121630- Range Spinal System manufactured by K2M, Inc. K I I I1492-TSRH Spinal System manufactured by… | 2012 |
| * | K111492 | Zimmer Patient Specific Instruments Planner Model 2.5, Zimmer Patient Specific Instruments | 2011 |
What descends from it? (81)
See all 814 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 18, 2015 | Class II | Z-1749-2015 | Unknown/Undetermined by firm | The Firebird Spinal Fixation System 6.5 mm x 55 mm Bone Screw, Self-Tapping is currently mislabeled at the distal tip of the bone screw showing a length measurement of 50 mm instead of 55 mm. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 9 years and passes through 5 companies; a new generation was cleared every 4 months on median.
- 25 of the 224 ancestors have a recall of any class.
- Of its descendants, K151488 (Orthofix, Inc., 2015) has gone furthest, with 78 descendants of its own.
- 6 other devices also cite its predicate K081684.
How does its category compare?
All NKB devicesNKB Thoracolumbosacral Pedicle Screw System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.
FDA decisions per year · NKB
24in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NKB
61in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NKB
1in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · NKB
1,171in 2025
MAUDE reports naming product code NKB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 38 | 87 | 4 | 34 |
| 2008 | 36 | 59 | 5 | 82 |
| 2009 | 44 | 71 | 33 | 171 |
| 2010 | 37 | 111 | 4 | 462 |
| 2011 | 48 | 84 | 3 | 577 |
| 2012 | 60 | 79 | 13 | 1,747 |
| 2013 | 35 | 73 | 0 | 1,962 |
| 2014 | 46 | 90 | 3 (1) | 1,671 |
| 2015 | 55 | 90 | 3 | 1,446 |
| 2016 | 50 | 84 | 1 | 1,585 |
| 2017 | 63 | 69 | 1 | 1,618 |
| 2018 | 56 | 61 | 3 | 1,684 |
| 2019 | 39 | 84 | 1 | 1,915 |
| 2020 | 48 | 68 | 1 | 2,076 |
| 2021 | 32 | 61 | 0 | 1,745 |
| 2022 | 34 | 88 | 5 | 1,339 |
| 2023 | 44 | 72 | 2 | 919 |
| 2024 | 39 | 57 | 3 | 1,614 |
| 2025 | 24 | 61 | 1 | 1,171 |
| 2026 | 30 | 91 | 3 | — |
What is it indicated for?
The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle, and non-pedicle fixation (TI-52/Iliumn). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allagraft. The device is indicated for all of the following indications: * degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and…
Read the full statement (238 words)
The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle, and non-pedicle fixation (TI-52/Iliumn). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allagraft. The device is indicated for all of the following indications:
* degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
* spondy lolisthesis,
* trauma (i.e., fracture or dislocation),
* spinal stenosis,
* deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
* tumnor,
* pseudoarthrosis, and
* failed previous fusion
When used far fixation to the ilium, the offset connectors of the Firebird Spinal Fixation System must be used in conjunction with pedicle screws placed at the SI or S2 spinal level.
The Phoenix MIS Fixation System when used with the Firebird Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The Firebird Spinal Fixation System components are used with certain components of the Orthofix Spinal Fixation System, including rods, rod connectors and cross-connectors.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Firebird Spinal Fixation System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Firebird Spinal Fixation System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K130932 · printed Nov 6, 2026
Cite this page
510(k) Report. “K130932: Firebird Spinal Fixation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130932
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026