510(k) Report

510(k) · Cleared

Cleared

2008T Hemodialysis Machine with Optional Crit-Line Monitor (Clic)

K131908510(k) TraditionalCleared Oct 11, 2013107-day reviewClass II21 CFR 876.5860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care Renal Therapies Group, LLC · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Oct 11, 2013 as substantially equivalent to K121341 (Fresenius Medical Care North America, 2012) and 1 other predicate after a 107-day review. Its family tree reaches back 7 generations to 1994. 3 of the 23 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it.

Decided
Oct 11, 2013
Review
107 days
Class
Class II
Product code
KDI
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K921456 (1994). 12 ancestors in all, the oldest K873155 (1987). 10 later devices descend from it, 1 citing it directly. 3 of the 23 devices in this family, counting this one, have a Class I recall.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor4 generations back: 2 other ancestors5 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 1 descendant3 generations forward: 1 descendant, 1 with a Class I recall14 generations forward: 2 descendants, 1 with a Class I recall15 generations forward: 4 descendants6 generations forward: 1 descendant K121341 · Fresenius 2008T Hemodialysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK120017 · Fresenius 2008 Hemodiaysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK101715 · Fresenius 2008T Hemodialysis Machine Model 190573 · Fresenius Medical Care North America, Design Cente · 2010 · recalledK093902 · Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 · Fresenius Medical Care North America · 2010 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled−50+5
  1. Root · 7 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 6 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 5 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. 4 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
  5. 3 generations backK101715 Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · Recalled
  6. 2 generations backK120017 Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  7. 1 generation backK121341 Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  8. This deviceK131908 2008T Hemodialysis Machine with Optional Crit-Line Monitor (Clic)Fresenius Medical Care Renal Therapies Group, LLC · 2013
  9. Descendant · cites it directly · 9 of its ownK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
  10. Descendant · 2 generations on · 8 of its ownK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  11. Descendant · 3 generations on · 7 of its ownK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  12. Descendant · 4 generations on · 4 of its ownK231125 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2023 · Recalled
  13. Descendant · 4 generations on · 3 of its ownK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Recalled (Class I)

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12018: 12022: 12023: 12024: 12025: 4
2015high 4 (2025) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · cites it directly9descendants
  • K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · 2 generations on8descendants
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 3 generations on7descendants
  • K2311252008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2023 · 4 generations on4descendants
  • K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−6K994267Fresenius 2008KFresenius Medical Care20005
−5K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
−4K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America20102
−3K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente20101
−2K120017Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America20123
−1K121341Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America20125
0 (this device)K1319082008T Hemodialysis Machine with Optional Crit-Line Monitor (Clic)Fresenius Medical Care Renal Therapies Group, LLC2013—
+1K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC2015Class I
+2K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20162
+3K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC2018Class I
+4K2311252008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20234
+4K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC2022Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199410
199710
200210
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201511
201610
201811
202211
202310
202410
202540
FIG. 1 — Lineage of K131908, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K121341Fresenius 2008T Hemodialysis Machine with Bibag System
Summary page 3
…ne V MEDICAL CARE Traditional 510(k) Notification Modifications to the previously cleared 2008T Hemodialysis Machine (K121341) include: "Crit-Line Clip Monitor (CLIC) (Optional): Hardware, software and labeling modifications to integrate…
Fresenius Medical Care North America20125
*K121599Crit-Line Clip (Clic)
Summary page 3
…21341) include: "Crit-Line Clip Monitor (CLIC) (Optional): Hardware, software and labeling modifications to integrate the Crit-Line Clip Monitor (CLiC; K121599) and expansion of the Indications for Use for use of the CLiC with the 2008T Hemodialysis Machine. When integ…
Fresenius Medical Care-North America20121

Also cleared against K121341: 2 other devices →

What descends from it? (10)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine Class IFresenius Medical Care Renal Therapies Group, LLC20159

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 18 Class I recalls
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 3 generations on · ◆ in the 2018 descendants bar of FIG. 18 Class I recalls
  • K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · 4 generations on · ◆ in the 2022 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 19 years and passes through 5 companies; a new generation was cleared every 1.1 years on median.
  • 8 of the 12 ancestors have a recall of any class.
  • Of its descendants, K150708 (Fresenius Medical Care Renal Therapies Group, LLC, 2015) has gone furthest, with 9 descendants of its own.
  • 2 other devices also cite its predicate K121341.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy. The bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is Intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription. The Crit-Line Clip…

Read the full statement (183 words)

The 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.

The bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is Intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.

The Crit-Line Clip Monitor is used with the 2008T Hemodialysis Machine to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131908: 2008T Hemodialysis Machine with Optional Crit-Line Monitor (Clic).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K131908

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026