Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573

K093902510(k) SpecialCleared May 27, 2010156-day reviewClass II21 CFR 876.5860Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care North America · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on May 27, 2010 as substantially equivalent to K080964 (Fresenius Medical Care North America, 2009) after a 156-day review. Its family tree reaches back 3 generations to 1994. 4 of the 40 devices in its family, counting itself, have a Class I recall. 36 later devices descend from it. It has been recalled 2 times.

Decided
May 27, 2010
Review
156 days
Class
Class II
Product code
KDI
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K921456 (1994). 36 later devices descend from it, 2 citing it directly. 4 of the 40 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2010: 1 descendant cleared2011: 2 descendants cleared2012: 3 descendants cleared2014: 2 descendants cleared2013: 2 descendants cleared2015: 2 descendants cleared, 1 with a Class I recall12020: 2 descendants cleared2016: 2 descendants cleared2021: 2 descendants cleared2018: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared2022: 3 descendants cleared, 1 with a Class I recall12024: 5 descendants cleared2023: 3 descendants cleared, 1 with a Class I recall12025: 4 descendants cleared2026: 1 descendant clearedK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalled1990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 4 descendants5 generations forward: 3 descendants, 1 with a Class I recall16 generations forward: 3 descendants7 generations forward: 5 descendants, 1 with a Class I recall18 generations forward: 4 descendants, 1 with a Class I recall19 generations forward: 7 descendants10 generations forward: 3 descendants, 1 with a Class I recall111 generations forward: 1 descendant K080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled0+5+10
  1. Root · 3 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 2 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 1 generation backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. This deviceK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010
  5. Descendant · cites it directly · 19 of its ownK111639 Fresenius 2008T Hemodialysis MachineFresenius Medical Care · 2011 · Recalled
  6. Descendant · 2 generations on · 18 of its ownK113427 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · Recalled
  7. Descendant · 3 generations on · 17 of its ownK120505 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012 · Recalled
  8. Descendant · 4 generations on · 16 of its ownK140866 Tablo ConsoleHome Dialysis Plus · 2014
  9. Descendant · cites it directly · 15 of its ownK101715 Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · Recalled

Showing 5 of 36 descendants. See all descendants

Descendants cleared per year

2010: 12011: 22012: 32013: 22014: 22015: 22016: 22018: 12019: 12020: 22021: 22022: 32023: 32024: 52025: 42026: 1
2010high 5 (2024) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 36 · See all
  • K111639Fresenius 2008T Hemodialysis MachineFresenius Medical Care · 2011 · cites it directly19descendants
  • K113427Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · 2 generations on18descendants
  • K120505Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012 · 3 generations on17descendants
  • K140866Tablo ConsoleHome Dialysis Plus · 2014 · 4 generations on16descendants
  • K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · cites it directly15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−2K994267Fresenius 2008KFresenius Medical Care20005
−1K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
0 (this device)K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America2010—
+1K111639Fresenius 2008T Hemodialysis MachineFresenius Medical Care20111
+2K113427Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America20114
+3K120505Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente20124
+4K140866Tablo ConsoleHome Dialysis Plus2014—
+1K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente20101

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201120
201230
201320
201420
201521
201620
201811
201910
202020
202120
202231
202331
202450
202540
202610
FIG. 1 — Lineage of K093902, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080964Fresenius Model 2008T Hemodialysis Machine
Summary page 2
…Does the new device have the same intended use? YES - The intended use for the modified 2008T Hemodialysis Machine is identical to that for the unmodified 2008T (K080964) as follows: Intended Use The Fresenius 2008T is indicated for acute and chronic hemodialysis 3. Does the device have…
Fresenius Medical Care North America20094

Also cleared against K080964: 2 other devices →

What descends from it? (36)

See all 36

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111639Fresenius 2008T Hemodialysis MachineFresenius Medical Care201119
K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente201015

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 12, 2013Class IIZ-0629-2014Under Investigation by firm2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
May 24, 2013Class IIZ-1693-2013Nonconforming Material/ComponentIncorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 5 generations on · ◆ in the 2015 descendants bar of FIG. 18 Class I recalls
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 7 generations on · ◆ in the 2018 descendants bar of FIG. 18 Class I recalls
  • K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · 8 generations on · ◆ in the 2022 descendants bar of FIG. 18 Class I recalls
  • K233335Tablo® Hemodialysis SystemOutset Medical, Inc. · 2023 · 10 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 5.9 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K111639 (Fresenius Medical Care, 2011) has gone furthest, with 19 descendants of its own.
  • 2 other devices also cite its predicate K080964.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

Fresenius 2008T is indicated for acute and chronic dialysis therapy

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093902: Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K093902

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026