510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Centricity Universal Viewer Zero Footprint Client

K131977510(k) TraditionalCleared Sep 17, 201381-day reviewClass II21 CFR 892.2050Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · LLZ System, Image Processing, Radiological

Cleared on Sep 17, 2013 as substantially equivalent to K123174 (GE Healthcare, 2012) after an 81-day review. Its family tree reaches back 7 generations to 1994. 6 later devices descend from it. It has been recalled 8 times.

Decided
Sep 17, 2013
Review
81 days
Class
Class II
Product code
LLZ
Recalls
8
Predicates
1

Its longest line of descent runs back 7 generations to K942120 (1994). 14 ancestors in all. 6 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor2003: 2 other ancestors2000: 1 other ancestor1996: 1 other ancestor2014: 1 descendant cleared2016: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedK942120 · Advantage Windows Diagnostic X-Ray Workstation · GE Medical Systems · 1994K013630 · Isite Radiology · Stentor, Inc. · 2001K031311 · Integradweb by Dynamic Imaging, Inc. · Dynamic Imaging, Inc. · 2003K042313 · Integradweb MPR/Mip · Dynamic Imaging, Inc. · 2004K082318 · Centricity Pacs IW · Ge Healthcare Dynamic Imaging Solutions · 2008 · recalledK121387 · Centricity Pacs-IW · Ge Healthcare Integrated IT Solutions · 2012 · recalledK123174 · Centricity Pacs-IW with Universal Viewer · GE Healthcare · 2012 · recalled1990200020102020
  1. Root · 7 generations backK942120 Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems · 1994
  2. 6 generations backK013630 Isite RadiologyStentor, Inc. · 2001
  3. 5 generations backK031311 Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc. · 2003
  4. 4 generations backK042313 Integradweb MPR/MipDynamic Imaging, Inc. · 2004
  5. 3 generations backK082318 Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions · 2008 · Recalled
  6. 2 generations backK121387 Centricity Pacs-IWGe Healthcare Integrated IT Solutions · 2012 · Recalled
  7. 1 generation backK123174 Centricity Pacs-IW with Universal ViewerGE Healthcare · 2012 · Recalled
  8. This deviceK131977 Centricity Universal Viewer Zero Footprint ClientGE Healthcare · 2013
  9. Descendant · cites it directly · 5 of its ownK141881 RapidresultsRamsoft, Inc. · 2014
  10. Descendant · 2 generations on · 3 of its ownK222476 OmegaAI Image ViewerRamsoft, Inc. · 2022
  11. Descendant · 3 generations on · 0 of its ownK254237 Cci Pacs Viewer (Pacs-US-001)CliniComp, Intl. · 2026
  12. Descendant · 3 generations on · 0 of its ownK241546 aiCockpit AI ViewerFovia, Inc. · 2025
  13. Descendant · 3 generations on · 0 of its ownK242362 Lightning ViewerNexsys Electronics Inc. Dba Medweb · 2024

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2014: 12016: 12022: 12024: 12025: 12026: 1
2014high 1 (2014) · low 0 (2015)2026

Most influential descendants

Showing 5 of 6 · See all
  • K141881RapidresultsRamsoft, Inc. · 2014 · cites it directly5descendants
  • K222476OmegaAI Image ViewerRamsoft, Inc. · 2022 · 2 generations on3descendants
  • K254237Cci Pacs Viewer (Pacs-US-001)CliniComp, Intl. · 2026 · 3 generations on0descendants
  • K241546aiCockpit AI ViewerFovia, Inc. · 2025 · 3 generations on0descendants
  • K242362Lightning ViewerNexsys Electronics Inc. Dba Medweb · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K942120Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems1994—
−6K013630Isite RadiologyStentor, Inc.2001—
−5K031311Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc.2003—
−4K042313Integradweb MPR/MipDynamic Imaging, Inc.2004—
−3K082318Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions20087
−2K121387Centricity Pacs-IWGe Healthcare Integrated IT Solutions20121
−1K123174Centricity Pacs-IW with Universal ViewerGE Healthcare201210
0 (this device)K131977Centricity Universal Viewer Zero Footprint ClientGE Healthcare2013—
+1K141881RapidresultsRamsoft, Inc.2014—
+2K222476OmegaAI Image ViewerRamsoft, Inc.2022—
+3K254237Cci Pacs Viewer (Pacs-US-001)CliniComp, Intl.2026—
+3K241546aiCockpit AI ViewerFovia, Inc.2025—
+3K242362Lightning ViewerNexsys Electronics Inc. Dba Medweb2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200010
200320
200410
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201610
202210
202410
202510
202610
FIG. 1 — Lineage of K131977, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K123174Centricity Pacs-IW with Universal Viewer
Summary page 1
…Classification Names: 21 CFR 892.2050, System, Image Processing, Radiological Product Code: LLZ Predicate Device: K1 23174 - GE Healthcare Centricity PACS-IW with Universal Viewer Page 5 -…
GE Healthcare201210

Also cleared against K123174: 5 other devices →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141881RapidresultsRamsoft, Inc.20145

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 16, 2026Class IIZ-1475-2026Device DesignUnder certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient…
Oct 3, 2024Class IIZ-0318-2025Software Manufacturing/Software DeploymentGE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate…
Aug 29, 2024Class IIZ-0039-2025Process controlGE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM…
Jul 23, 2024Class IIZ-3160-2024Software Design ChangeFootprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and…
Jun 8, 2022Class IIZ-1333-2022Software designPotential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
Apr 28, 2022Class IIZ-1229-2022Software designPotential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on…
Oct 15, 2021Class IIZ-0329-2022Software designThere is a potential to display incomplete patient imaging study.
Nov 9, 2018Class IIZ-0641-2019Software designIssue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2. When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • 5 of the 14 ancestors have a recall of any class.
  • Of its descendants, K141881 (Ramsoft, Inc., 2014) has gone furthest, with 5 descendants of its own.
  • 5 other devices also cite its predicate K123174.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Centricity Universal Viewer Zero Footprint client is a device that displays medical images, data from various imaging sources, and other healthcare information sources. Medical images and data can be viewed, communicated, processed and displayed within a computer network or on a workstation. The device may be used to provide images for diagnostic purposes by trained professionals. Typical users of this system are authorized individuals and trained…

Read the full statement (137 words)

Centricity Universal Viewer Zero Footprint client is a device that displays medical images, data from various imaging sources, and other healthcare information sources. Medical images and data can be viewed, communicated, processed and displayed within a computer network or on a workstation. The device may be used to provide images for diagnostic purposes by trained professionals.

Typical users of this system are authorized individuals and trained healthcare professionals who view medical images and data.

Mammographic images may only be interpreted using a monitor compliant with requirements of local regulations and must meet other technical specifications reviewed and accepted by the local regulatory agencies.

Contraindications:

Centricity Universal Viewer Zero Footprint client is contraindicated for the use of lossy compressed mammographic images. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131977: Centricity Universal Viewer Zero Footprint Client.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K131977

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026