510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Centricity Pacs-IW with Universal Viewer

K123174510(k) SpecialCleared Nov 16, 201237-day reviewClass II21 CFR 892.2050Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · LLZ System, Image Processing, Radiological

Cleared on Nov 16, 2012 as substantially equivalent to K121387 (Ge Healthcare Integrated IT Solutions, 2012) after a 37-day review. Its family tree reaches back 6 generations to 1994. 15 later devices descend from it. It has been recalled 10 times.

Decided
Nov 16, 2012
Review
37 days
Class
Class II
Product code
LLZ
Recalls
10
Predicates
1

Its longest line of descent runs back 6 generations to K942120 (1994). 13 ancestors in all. 15 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor2003: 2 other ancestors2000: 1 other ancestor1996: 1 other ancestor2015: 2 descendants cleared2016: 3 descendants cleared2013: 1 descendant cleared2024: 2 descendants cleared2014: 1 descendant cleared2018: 1 descendant cleared2022: 1 descendant cleared2021: 1 descendant cleared2026: 2 descendants cleared2025: 1 descendant clearedK942120 · Advantage Windows Diagnostic X-Ray Workstation · GE Medical Systems · 1994K013630 · Isite Radiology · Stentor, Inc. · 2001K031311 · Integradweb by Dynamic Imaging, Inc. · Dynamic Imaging, Inc. · 2003K042313 · Integradweb MPR/Mip · Dynamic Imaging, Inc. · 2004K082318 · Centricity Pacs IW · Ge Healthcare Dynamic Imaging Solutions · 2008 · recalledK121387 · Centricity Pacs-IW · Ge Healthcare Integrated IT Solutions · 2012 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 2 other ancestors4 generations back: 2 other ancestors5 generations back: 3 other ancestors 1 generation forward: 6 descendants2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 4 descendants K121387 · Centricity Pacs-IW · Ge Healthcare Integrated IT Solutions · 2012 · recalledK082318 · Centricity Pacs IW · Ge Healthcare Dynamic Imaging Solutions · 2008 · recalledK042313 · Integradweb MPR/Mip · Dynamic Imaging, Inc. · 2004K031311 · Integradweb by Dynamic Imaging, Inc. · Dynamic Imaging, Inc. · 2003K013630 · Isite Radiology · Stentor, Inc. · 2001K942120 · Advantage Windows Diagnostic X-Ray Workstation · GE Medical Systems · 1994−50
  1. Root · 6 generations backK942120 Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems · 1994
  2. 5 generations backK013630 Isite RadiologyStentor, Inc. · 2001
  3. 4 generations backK031311 Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc. · 2003
  4. 3 generations backK042313 Integradweb MPR/MipDynamic Imaging, Inc. · 2004
  5. 2 generations backK082318 Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions · 2008 · Recalled
  6. 1 generation backK121387 Centricity Pacs-IWGe Healthcare Integrated IT Solutions · 2012 · Recalled
  7. This deviceK123174 Centricity Pacs-IW with Universal ViewerGE Healthcare · 2012
  8. Descendant · cites it directly · 6 of its ownK131977 Centricity Universal Viewer Zero Footprint ClientGE Healthcare · 2013 · Recalled
  9. Descendant · 2 generations on · 5 of its ownK141881 RapidresultsRamsoft, Inc. · 2014
  10. Descendant · 3 generations on · 3 of its ownK222476 OmegaAI Image ViewerRamsoft, Inc. · 2022
  11. Descendant · cites it directly · 3 of its ownK150420 Centricity Universal ViewerGE Healthcare · 2015 · Recalled
  12. Descendant · 2 generations on · 2 of its ownK182419 Centricity Universal ViewerGE Healthcare · 2018 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2013: 12014: 12015: 22016: 32018: 12021: 12022: 12024: 22025: 12026: 2
2013high 3 (2016) · low 0 (2017)2026

Most influential descendants

Showing 5 of 15 · See all
  • K131977Centricity Universal Viewer Zero Footprint ClientGE Healthcare · 2013 · cites it directly6descendants
  • K141881RapidresultsRamsoft, Inc. · 2014 · 2 generations on5descendants
  • K222476OmegaAI Image ViewerRamsoft, Inc. · 2022 · 3 generations on3descendants
  • K150420Centricity Universal ViewerGE Healthcare · 2015 · cites it directly3descendants
  • K182419Centricity Universal ViewerGE Healthcare · 2018 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K942120Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems1994—
−5K013630Isite RadiologyStentor, Inc.2001—
−4K031311Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc.2003—
−3K042313Integradweb MPR/MipDynamic Imaging, Inc.2004—
−2K082318Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions20087
−1K121387Centricity Pacs-IWGe Healthcare Integrated IT Solutions20121
0 (this device)K123174Centricity Pacs-IW with Universal ViewerGE Healthcare2012—
+1K131977Centricity Universal Viewer Zero Footprint ClientGE Healthcare20138
+2K141881RapidresultsRamsoft, Inc.2014—
+3K222476OmegaAI Image ViewerRamsoft, Inc.2022—
+1K150420Centricity Universal ViewerGE Healthcare201515
+2K182419Centricity Universal ViewerGE Healthcare20188

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200010
200320
200410
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201410
201520
201630
201810
202110
202210
202420
202510
202620
FIG. 1 — Lineage of K123174, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K121387Centricity Pacs-IW
Summary page 1
…tem Classification Names: 21 CFR 892.2050, System, Image Processing, Radiological Product Code: LLZ Predicate Device: K121387 - GE Healthcare Centricity PACS-IW Device Descrition: Centricity PACS-IW with Universal Viewer is an Internet…
Ge Healthcare Integrated IT Solutions20121

What descends from it? (15)

See all 15

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131977Centricity Universal Viewer Zero Footprint ClientGE Healthcare20136
K150420Centricity Universal ViewerGE Healthcare20153
K242810RadUnity (Version 1.0.0)Radunity Corp.20240
K153284Synaptive ImageDrive ProSynaptive Medical, Inc.20160
K152716EV Insite SystemPsp Corporation20160
K141561VM Pacs with VM Medical WorkstationVisual Medica20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 30, 2026Class IIZ-1571-2026Software designThere is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially…
Feb 19, 2025Class IIZ-1484-2025Software designThere is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be…
Sep 14, 2023Class IIZ-0139-2024Software designGE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal…
Mar 28, 2023Class IIZ-1500-2023OtherGE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
Oct 28, 2022Class IIZ-0721-2023Software Manufacturing/Software DeploymentThere is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
Aug 6, 2021Class IIZ-2502-2021Software designImage acquisition failures and synchronization failure with the Centricity Enterprise Archive
May 17, 2019Class IIZ-1885-2019Software designThere is a potential that one or more images or image series may be missing from exams without a warning displayed in the viewer.
Oct 11, 2018Class IIZ-0374-2019Software designSynchronization failure
Jan 25, 2018Class IIZ-2101-2018Software designPotential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.
Sep 28, 2015Class IIZ-0225-2016Software designInaccurate distance measurements with magnified projection X-ray images.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 6 companies; a new generation was cleared every 3.8 years on median.
  • 4 of the 13 ancestors have a recall of any class.
  • Of its descendants, K131977 (GE Healthcare, 2013) has gone furthest, with 6 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Centricity PACS-IW with Universal Viewer is a software only product that is part of a Picture Archiving and Communication System (PACS) medical device. Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed…

Read the full statement (129 words)

Centricity PACS-IW with Universal Viewer is a software only product that is part of a Picture Archiving and Communication System (PACS) medical device.

Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA.

Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123174: Centricity Pacs-IW with Universal Viewer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123174

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026