510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Cios Alpha

K132094510(k) TraditionalCleared Mar 11, 2014246-day reviewClass II21 CFR 892.1650Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medi Cal Solutions, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Mar 11, 2014 as substantially equivalent to K051133 (Siemens Medical Solutions USA, Inc., 2005) after a 246-day review. Its family tree reaches back 6 generations to 1991. 19 later devices descend from it. It has been recalled 8 times.

Decided
Mar 11, 2014
Review
246 days
Class
Class II
Product code
OWB
Recalls
8
Predicates
1

Its longest line of descent runs back 6 generations to K912470 (1991). 17 ancestors in all, the oldest K860913 (1986). 19 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2001: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1993: 2 other ancestors1991: 3 other ancestors1986: 1 other ancestor1995: 1 other ancestor2016: 2 descendants cleared2018: 3 descendants cleared2025: 3 descendants cleared2020: 2 descendants cleared2019: 1 descendant cleared2021: 4 descendants cleared2026: 1 descendant cleared2024: 1 descendant cleared2022: 2 descendants clearedK912470 · Philips Diagnost 96 · Philips Medical Systems, Inc. · 1991 · recalledK960266 · Siregraph CF · Siemens Medical Solutions USA, Inc. · 1996 · recalledK970734 · Siregragh T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic X-Ray Systems · Siemens Medical Solutions USA, Inc. · 1997K992660 · Urf Digital - OT · Siemens Medical Solutions USA, Inc. · 1999 · recalledK040675 · Axiom Artis U · Siemens Medical Systems, Inc. · 2004 · recalledK051133 · Arcadis Avantic · Siemens Medical Solutions USA, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 4 other ancestors5 generations back: 1 other ancestor6 generations back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 7 descendants3 generations forward: 5 descendants4 generations forward: 2 descendants K051133 · Arcadis Avantic · Siemens Medical Solutions USA, Inc. · 2005K040675 · Axiom Artis U · Siemens Medical Systems, Inc. · 2004 · recalledK992660 · Urf Digital - OT · Siemens Medical Solutions USA, Inc. · 1999 · recalledK970734 · Siregragh T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic X-Ray Systems · Siemens Medical Solutions USA, Inc. · 1997K960266 · Siregraph CF · Siemens Medical Solutions USA, Inc. · 1996 · recalledK912470 · Philips Diagnost 96 · Philips Medical Systems, Inc. · 1991 · recalled−50
  1. Root · 6 generations backK912470 Philips Diagnost 96Philips Medical Systems, Inc. · 1991 · Recalled
  2. 5 generations backK960266 Siregraph CFSiemens Medical Solutions USA, Inc. · 1996 · Recalled
  3. 4 generations backK970734 Siregragh T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic…Siemens Medical Solutions USA, Inc. · 1997
  4. 3 generations backK992660 Urf Digital - OTSiemens Medical Solutions USA, Inc. · 1999 · Recalled
  5. 2 generations backK040675 Axiom Artis USiemens Medical Systems, Inc. · 2004 · Recalled
  6. 1 generation backK051133 Arcadis AvanticSiemens Medical Solutions USA, Inc. · 2005
  7. This deviceK132094 Cios AlphaSiemens Medi Cal Solutions, Inc. · 2014
  8. Descendant · cites it directly · 15 of its ownK153244 Cios FusionSiemens Medical Solutions USA, Inc. · 2016 · Recalled
  9. Descendant · cites it directly · 10 of its ownK181550 Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · Recalled
  10. Descendant · cites it directly · 10 of its ownK181560 Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · Recalled
  11. Descendant · 2 generations on · 9 of its ownK203504 Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · Recalled
  12. Descendant · 2 generations on · 5 of its ownK181767 Cios SelectSiemens Medical Solutions USA, Inc. · 2018

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2016: 22018: 32019: 12020: 22021: 42022: 22024: 12025: 32026: 1
2016high 4 (2021) · low 0 (2017)2026

Most influential descendants

Showing 5 of 19 · See all
  • K153244Cios FusionSiemens Medical Solutions USA, Inc. · 2016 · cites it directly15descendants
  • K181550Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · cites it directly10descendants
  • K181560Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · cites it directly10descendants
  • K203504Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · 2 generations on9descendants
  • K181767Cios SelectSiemens Medical Solutions USA, Inc. · 2018 · 2 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K912470Philips Diagnost 96Philips Medical Systems, Inc.19919
−5K960266Siregraph CFSiemens Medical Solutions USA, Inc.19963
−4K970734Siregragh T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic X-Ray SystemsSiemens Medical Solutions USA, Inc.1997—
−3K992660Urf Digital - OTSiemens Medical Solutions USA, Inc.19994
−2K040675Axiom Artis USiemens Medical Systems, Inc.20041
−1K051133Arcadis AvanticSiemens Medical Solutions USA, Inc.2005—
0 (this device)K132094Cios AlphaSiemens Medi Cal Solutions, Inc.2014—
+1K153244Cios FusionSiemens Medical Solutions USA, Inc.20161
+1K181550Cios SpinSiemens Medical Solutions USA, Inc.20184
+1K181560Cios AlphaSiemens Medical Systems USA, Inc.20187
+2K203504Cios FlowSiemens Medical Solutions USA, Inc.20201
+2K181767Cios SelectSiemens Medical Solutions USA, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199130
199320
199510
199610
199710
200110
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
201830
201910
202020
202140
202220
202410
202530
202610
FIG. 1 — Lineage of K132094, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051133Arcadis Avantic
Summary page 2
…. 7. Substantial Equivalence: The Cios Alpha is substantially equivalent to the commercially available Siemens ARCADIS Avantic, cleared 06/01/2005 with K051133. The Indication for use statement is similar to the predicate. The image intensifier has been replaced by a solid s…
Siemens Medical Solutions USA, Inc.2005—

Also cleared against K051133: 2 other devices →

What descends from it? (19)

See all 19

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153244Cios FusionSiemens Medical Solutions USA, Inc.201615
K181550Cios SpinSiemens Medical Solutions USA, Inc.201810
K181560Cios AlphaSiemens Medical Systems USA, Inc.201810
K251520Cios Alpha; Cios FlowSiemens Medical Solutions USA, Inc.20250
K153147Cios ConnectSiemens Medical Solutions USA, Inc.20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 14, 2020Class IIZ-2848-2020Device DesignHardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Nov 25, 2019Class IIZ-0782-2020Nonconforming Material/ComponentDuring a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely…
Jun 20, 2019Class IIZ-1953-2019Nonconforming Material/ComponentDAP chamber may be missing an insulating foil
Mar 18, 2019Class IIZ-1298-2019Device DesignContinuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a…
Feb 19, 2016Class IIZ-1278-2016Software in the Use EnvironmentSoftware issues on Cios Alpha mobile C-Arm system
Feb 16, 2016Class IIZ-1281-2016Nonconforming Material/ComponentSiemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when…
Aug 17, 2015Class IIZ-0118-2016Device Designpatient procedure interruption due to a potential system failure
May 5, 2015Class IIZ-1958-2015Software designunder certain circumstances the Cios Alpha system may freeze during a procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 4.6 years on median.
  • 12 of the 17 ancestors have a recall of any class.
  • Of its descendants, K153244 (Siemens Medical Solutions USA, Inc., 2016) has gone furthest, with 15 descendants of its own.
  • 2 other devices also cite its predicate K051133.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K132094: Cios Alpha.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K132094

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026