510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus B2& Aquablus HF

K133829510(k) TraditionalCleared Apr 22, 2014126-day reviewClass II21 CFR 876.5665Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vivonic GmbH · FIP Subsystem, Water Purification

Cleared on Apr 22, 2014 as substantially equivalent to K974899 (Gambro Healthcare, 1998) and 1 other predicate after a 126-day review. Its family tree reaches back 4 generations to 1981. 3 later devices descend from it. It has been recalled 3 times.

Decided
Apr 22, 2014
Review
126 days
Class
Class II
Product code
FIP
Recalls
3
Predicates
2

Its longest line of descent runs back 4 generations to K811678 (1981). 8 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor2003: 1 other ancestor1994: 1 other ancestor1996: 1 other ancestor2024: 1 descendant cleared2022: 1 descendant cleared2025: 1 descendant clearedK811678 · Gambro Water Purification Monitor WRD · Gambro, Inc. · 1981K042797 · Wro 300 Water Purification System for Hemodialysis · Gambro Renal Products · 2005K093608 · Gambro Water Purification Unit, Models Wro 300 and Wro 300 H · Gambro Renal Products, Inc. · 2010K124059 · Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT) · Lauer Membran Wassertechnik GmbH · 201319801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K124059 · Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT) · Lauer Membran Wassertechnik GmbH · 2013K093608 · Gambro Water Purification Unit, Models Wro 300 and Wro 300 H · Gambro Renal Products, Inc. · 2010K042797 · Wro 300 Water Purification System for Hemodialysis · Gambro Renal Products · 2005K811678 · Gambro Water Purification Monitor WRD · Gambro, Inc. · 19810
  1. Root · 4 generations backK811678 Gambro Water Purification Monitor WRDGambro, Inc. · 1981
  2. 3 generations backK042797 Wro 300 Water Purification System for HemodialysisGambro Renal Products · 2005
  3. 2 generations backK093608 Gambro Water Purification Unit, Models Wro 300 and Wro 300 HGambro Renal Products, Inc. · 2010
  4. 1 generation backK124059 Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT)Lauer Membran Wassertechnik GmbH · 2013
  5. This deviceK133829 Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus&…Vivonic GmbH · 2014
  6. Descendant · cites it directly · 1 of its ownK213507 AquaAFresenius Medical Care Rental Therapies Group, LLC · 2022
  7. Descendant · 2 generations on · 0 of its ownK252181 AquaAFresenius Medical Care North America · 2025
  8. Descendant · cites it directly · 0 of its ownK232953 AquaBplus; AquaB LITEFresenius Medical Care Renal Therapies Group, LLC · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K213507AquaAFresenius Medical Care Rental Therapies Group, LLC · 2022 · cites it directly1descendant
  • K252181AquaAFresenius Medical Care North America · 2025 · 2 generations on0descendants
  • K232953AquaBplus; AquaB LITEFresenius Medical Care Renal Therapies Group, LLC · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K811678Gambro Water Purification Monitor WRDGambro, Inc.1981—
−3K042797Wro 300 Water Purification System for HemodialysisGambro Renal Products2005—
−2K093608Gambro Water Purification Unit, Models Wro 300 and Wro 300 HGambro Renal Products, Inc.2010—
−1K124059Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT)Lauer Membran Wassertechnik GmbH2013—
0 (this device)K133829Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus…Vivonic GmbH2014—
+1K213507AquaAFresenius Medical Care Rental Therapies Group, LLC2022—
+2K252181AquaAFresenius Medical Care North America2025—
+1K232953AquaBplus; AquaB LITEFresenius Medical Care Renal Therapies Group, LLC2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199610
199810
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202410
202510
FIG. 1 — Lineage of K133829, 1981–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FIP device

  1. #1 K251851 DIASAFEplusUS (F00013010)score 14.0 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K974899Gambro Central Water Treatment System CWP 100 - Wro H
Summary page 2
…02A40140210 CWFS400k DINEN IMO N01=800 -DIN EN 3S01345207 VIVONIC Claiming substantial equivalence to K974899 Gambro CWP 100 reverse osmosis Ki124059 Lauer Aquaboss EcoRO Die 1+11 (HT) Device Description The AquaB…
Gambro Healthcare19982
*K124059Aquaboss Ecoro Dia I (HT); Aquaboss Ecoro Dia II (HT)
Summary page 2
…N 3S01345207 VIVONIC Claiming substantial equivalence to K974899 Gambro CWP 100 reverse osmosis Ki124059 Lauer Aquaboss EcoRO Die 1+11 (HT) Device Description The AquaBplus is a microcontroller-controlled, fully automati…
Lauer Membran Wassertechnik GmbH2013—

Also cleared against K974899: 4 other devices →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213507AquaAFresenius Medical Care Rental Therapies Group, LLC20221
K232953AquaBplus; AquaB LITEFresenius Medical Care Renal Therapies Group, LLC20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 3, 2021Class IIZ-1364-2022Labeling Change ControlAquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and…
Mar 18, 2021Class IIZ-1536-2021Labeling Change ControlAquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and…
Sep 1, 2020Class IIZ-0098-2021Process controlIncorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 5 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 8 ancestors has a recall of any class.
  • Of its descendants, K213507 (Fresenius Medical Care Rental Therapies Group, LLC, 2022) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K974899.

How does its category compare?

All FIP devices

FIP Subsystem, Water Purification: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FIP

4in 2025

2007: 12008: 32009: 32010: 22011: 42012: 22013: 42014: 42015: 22016: 22017: 12018: 12019: 22020: 12021: 12022: 12023: 32024: 12025: 4
2007high 4 (2011) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FIP

88in 2025

2008: 202 days median2009: 167 days median2011: 116 days median2013: 125 days median2014: 126 days median2023: 69 days median2025: 88 days median
2007high 202 (2008) · low 69 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FIP

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FIP

77in 2025

2007: 12008: 52010: 122011: 42012: 42013: 412014: 342015: 122016: 32017: 62018: 32019: 132020: 122021: 352022: 572023: 1402024: 1742025: 77
2007high 174 (2024) · low 0 (2009)2025

MAUDE reports naming product code FIP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008320205
2009316700
20102—012
2011411604
20122—04
20134125041
20144126034
20152—012
20162—03
20171—06
20181—03
20192—013
20201—112
20211—035
20221—057
20233690140
20241—0174
2025488077
20260—0—

What is it indicated for?

The AquaBplus Water Purification Systems are reverse osmosis units intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies. These devices are intended to be a component in a complete water purification system, and are not complete water treatment systems. Each reverse osmosis unit must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well, to meet current AAMI/ANSI/ISO and Federal (U.S.) standards.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133829: Aquabplus,Aquabplus & Aquabplus B2,Aquabplus&Aquabplus HF,Aquabplus& Aquabplus B2& Aquablus HF.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133829

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026