510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Caresite Luer Access Device

K140311510(k) TraditionalCleared May 7, 201489-day reviewClass II21 CFR 880.5440Recalls 21

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FPA Set, Administration, Intravascular

Cleared on May 7, 2014 as substantially equivalent to K083723 (B.Braun Medical, Inc., 2009) after an 89-day review. Its family tree reaches back 4 generations to 1995. 1 later device descends from it. It has been recalled 21 times.

Decided
May 7, 2014
Review
89 days
Class
Class II
Product code
FPA
Recalls
21
Predicates
1

Its longest line of descent runs back 4 generations to K942988 (1995). 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK942988 Filtered Extension SetsB. Braun of America, Inc. · 1995 · Recalled
  2. 3 generations backK955585 V2 Injection SiteB.Braun Medical, Inc. · 1996 · Recalled
  3. 2 generations backK031923 Ultrasite ValveB.Braun Medical, Inc. · 2003
  4. 1 generation backK083723 A6 Luer Access DeviceB.Braun Medical, Inc. · 2009 · Recalled
  5. This deviceK140311 Caresite Luer Access DeviceB.Braun Medical, Inc. · 2014
  6. Descendant · cites it directly · 0 of its ownK212842 Microvolume Luer Access DeviceB.Braun Medical, Inc. · 2022

Most influential descendants

Showing 1 of 1 · See all
  • K212842Microvolume Luer Access DeviceB.Braun Medical, Inc. · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K942988Filtered Extension SetsB. Braun of America, Inc.199532
−3K955585V2 Injection SiteB.Braun Medical, Inc.199627
−2K031923Ultrasite ValveB.Braun Medical, Inc.2003—
−1K083723A6 Luer Access DeviceB.Braun Medical, Inc.200966
0 (this device)K140311Caresite Luer Access DeviceB.Braun Medical, Inc.2014—
+1K212842Microvolume Luer Access DeviceB.Braun Medical, Inc.2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K140311, 1995–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K083723A6 Luer Access Device
Summary page 1
…Intravascular Administration Set Classification Product Code: FPA PREDICATE DEVICE: 5 10(k) Number: K083723 Device Name: A6 Luer Access Device Classification Product Cod…
B.Braun Medical, Inc.200966

Also cleared against K083723: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K212842Microvolume Luer Access DeviceB.Braun Medical, Inc.20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2025Class IIZ-0621-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0618-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0620-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0665-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0666-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0667-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0671-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0672-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0668-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0682-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0697-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0674-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0686-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Jun 17, 2024Class IIZ-2404-2024Process controlPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction…
Aug 10, 2021Class IIZ-2553-2021Under Investigation by firmLeakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.
Aug 10, 2021Class IIZ-2555-2021Under Investigation by firmLeakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.
Aug 10, 2021Class IIZ-2557-2021Under Investigation by firmLeakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.
Aug 10, 2021Class IIZ-2554-2021Under Investigation by firmLeakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.
Nov 20, 2019Class IIZ-0935-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing
Nov 20, 2019Class IIZ-0939-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing
Nov 20, 2019Class IIZ-0937-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 2 companies; a new generation was cleared every 5.8 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K212842 (B.Braun Medical, Inc., 2022) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K083723.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The Caresite Luer Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. The Caresite Luer Access Device may be used with power injectors at a maximum pressure of 400 psi and a maximum flow rate of 15 mL/sec.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140311: Caresite Luer Access Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140311

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.