V2 Injection Site

K955585510(k) TraditionalCleared May 15, 1996160-day reviewClass II21 CFR 880.5440Recalls 27

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FPA Set, Administration, Intravascular

Cleared on May 15, 1996 as substantially equivalent to K942988 (B. Braun of America, Inc., 1995) after a 160-day review. Its family tree reaches back 1 generation to 1995. 45 later devices descend from it. It has been recalled 27 times.

Decided
May 15, 1996
Review
160 days
Class
Class II
Product code
FPA
Recalls
27
Predicates
1

Its longest line of descent runs back 1 generation to K942988 (1995). 45 later devices descend from it, 7 citing it directly.

Descendants, by year cleared 2002: 1 descendant cleared1997: 1 descendant cleared2003: 4 descendants cleared2008: 1 descendant cleared2009: 1 descendant cleared1998: 1 descendant cleared2010: 2 descendants cleared2014: 3 descendants cleared1999: 1 descendant cleared2025: 3 descendants cleared2013: 2 descendants cleared2022: 2 descendants cleared2012: 1 descendant cleared2017: 5 descendants cleared2021: 6 descendants cleared2020: 2 descendants cleared2016: 1 descendant cleared2023: 3 descendants cleared2019: 2 descendants cleared2026: 2 descendants cleared2024: 1 descendant clearedK942988 · Filtered Extension Sets · B. Braun of America, Inc. · 1995 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 7 descendants2 generations forward: 5 descendants3 generations forward: 6 descendants4 generations forward: 7 descendants5 generations forward: 8 descendants6 generations forward: 8 descendants7 generations forward: 2 descendants8 generations forward: 2 descendants K942988 · Filtered Extension Sets · B. Braun of America, Inc. · 1995 · recalled0+5
  1. Root · 1 generation backK942988 Filtered Extension SetsB. Braun of America, Inc. · 1995 · Recalled
  2. This deviceK955585 V2 Injection SiteB.Braun Medical, Inc. · 1996
  3. Descendant · cites it directly · 35 of its ownK973167 Positive Pressure ConnectorIcu Medical, Inc. · 1997
  4. Descendant · cites it directly · 32 of its ownK031923 Ultrasite ValveB.Braun Medical, Inc. · 2003
  5. Descendant · cites it directly · 31 of its ownK083723 A6 Luer Access DeviceB.Braun Medical, Inc. · 2009 · Recalled
  6. Descendant · 2 generations on · 31 of its ownK991653 Invision-Plus Injection Port Connector SystemRymed Technologies, Inc. · 1999 · Recalled
  7. Descendant · 2 generations on · 29 of its ownK021692 CLC2000 Catheter Patency DeviceIcu Medical, Inc. · 2003

Showing 5 of 45 descendants. See all descendants

Descendants cleared per year

1997: 11998: 11999: 12002: 12003: 42008: 12009: 12010: 22012: 12013: 22014: 32016: 12017: 52019: 22020: 22021: 62022: 22023: 32024: 12025: 32026: 2
1997high 6 (2021) · low 0 (2000)2026

Most influential descendants

Showing 5 of 45 · See all
  • K973167Positive Pressure ConnectorIcu Medical, Inc. · 1997 · cites it directly35descendants
  • K031923Ultrasite ValveB.Braun Medical, Inc. · 2003 · cites it directly32descendants
  • K083723A6 Luer Access DeviceB.Braun Medical, Inc. · 2009 · cites it directly31descendants
  • K991653Invision-Plus Injection Port Connector SystemRymed Technologies, Inc. · 1999 · 2 generations on31descendants
  • K021692CLC2000 Catheter Patency DeviceIcu Medical, Inc. · 2003 · 2 generations on29descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K942988Filtered Extension SetsB. Braun of America, Inc.199532
0 (this device)K955585V2 Injection SiteB.Braun Medical, Inc.1996—
+1K973167Positive Pressure ConnectorIcu Medical, Inc.1997—
+1K031923Ultrasite ValveB.Braun Medical, Inc.2003—
+1K083723A6 Luer Access DeviceB.Braun Medical, Inc.200966
+2K991653Invision-Plus Injection Port Connector SystemRymed Technologies, Inc.19994
+2K021692CLC2000 Catheter Patency DeviceIcu Medical, Inc.2003—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
199910
200210
200340
200810
200910
201020
201210
201320
201430
201610
201750
201920
202020
202160
202220
202330
202410
202530
202620
FIG. 1 — Lineage of K955585, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K942988Filtered Extension Sets
Summary page 1
…Set CLASSIFICATION NAME: General Hospital Class II, 80 FPA, Intravascular Administration Set 21 CFR 880.5860 SUBSTANTIAL EQUIVALENCE' TO: 510 (k) number Name Applicant K942988 Filtered Extension B. Braun of America Sets DEVICE DESCRIPTION: In accordance with section 510(k) for the Federal Food, Drug, a…
B. Braun of America, Inc.199532

What descends from it? (45)

See all 45

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K973167Positive Pressure ConnectorIcu Medical, Inc.199735
K031923Ultrasite ValveB.Braun Medical, Inc.200332
K083723A6 Luer Access DeviceB.Braun Medical, Inc.200931
K024124System 100-UF 500 Circuit with in-Line Needleless Access PortChf Solutions, Inc.20031
K073396Antimicrobial Ultrasite Value, Ultrasite AGB.Braun Medical, Inc.20080
K021488Surecan Safety Huber Needle Infusion SetsB.Braun Medical, Inc.20020
K974145V3 ValveB.Braun Medical, Inc.19980

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2025Class IIZ-0714-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Sep 15, 2022Class IIZ-0090-2023Process designThe extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and…
Jan 8, 2014Class IIZ-0987-2014Process controlComplaints were received reporting leakage on certain lots of SAFSITE Injection Sites and IV Sets manufactured with SAFSITE Injection Sites.The SAFSITE Injection Site is intended to provide needle free access on IV and Extension Sets.
Aug 5, 2013Class IIZ-2166-2013Process controlReports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve…
Nov 24, 2009Not classifiedZ-0685-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0684-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0686-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0706-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0708-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0678-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Nov 24, 2009Not classifiedZ-0707-2010Nonconforming Material/ComponentSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the…
Mar 27, 2009Not classifiedZ-1428-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1427-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1430-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1429-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1431-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1433-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Mar 27, 2009Not classifiedZ-1432-2009Mixed-up of materials/componentsDevice defect is occlusion/no flow.
Sep 17, 2008Not classifiedZ-0199-2009Employee errorProduct was assembled incorrectly and connected to the wrong part.
Jun 30, 2008Not classifiedZ-2410-2008Mixed-up of materials/componentsincorrect burette was packaged with the product
Jun 30, 2008Not classifiedZ-2407-2008Mixed-up of materials/componentsincorrect burette was packaged with the product
Jun 30, 2008Not classifiedZ-2411-2008Mixed-up of materials/componentsincorrect burette was packaged with the product
Jun 30, 2008Not classifiedZ-2413-2008Mixed-up of materials/componentsincorrect burette was packaged with the product
Jun 30, 2008Not classifiedZ-2412-2008Mixed-up of materials/componentsincorrect burette was packaged with the product
Apr 18, 2007Not classifiedZ-0818-2007OtherSecondary infusion backs up into primary container.
Apr 18, 2007Not classifiedZ-0819-2007OtherSecondary infusion backs up into primary container.
Apr 14, 2003Not classifiedZ-0933-03Otherpossible leakage

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K973167 (Icu Medical, Inc., 1997) has gone furthest, with 35 descendants of its own.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The V2 Injection Site is designed to provide needle-free access on I.V. and Extension Sets as an alternative to latex injection sites or other needle-free access systems.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K955585: V2 Injection Site.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K955585

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026