510(k) · Cleared
Recalled · Class II · 2025A6 Luer Access Device
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
B.Braun Medical, Inc. · FPA Set, Administration, Intravascular
Cleared on Jun 5, 2009 as substantially equivalent to K031923 (B.Braun Medical, Inc., 2003) and 1 other predicate after a 172-day review. Its family tree reaches back 3 generations to 1995. 31 later devices descend from it. It has been recalled 66 times.
- Decided
- Jun 5, 2009
- Review
- 172 days
- Class
- Class II
- Product code
- FPA
- Recalls
- 66
- Predicates
- 2
Its longest line of descent runs back 3 generations to K942988 (1995). 31 later devices descend from it, 2 citing it directly.
- Root · 3 generations backK942988 Filtered Extension SetsB. Braun of America, Inc. · 1995 · Recalled
- 2 generations backK955585 V2 Injection SiteB.Braun Medical, Inc. · 1996 · Recalled
- 1 generation backK031923 Ultrasite ValveB.Braun Medical, Inc. · 2003
- This deviceK083723 A6 Luer Access DeviceB.Braun Medical, Inc. · 2009
- Descendant · cites it directly · 28 of its ownK100434 Clave NeutronIcu Medical, Inc. · 2010 · Recalled
- Descendant · 2 generations on · 26 of its ownK132413 MZ1000Care Fusion · 2013
- Descendant · 3 generations on · 13 of its ownK140831 Maxzero Extension Set with Needless ConnectorCare Fusion · 2014
- Descendant · 3 generations on · 6 of its ownK170336 BD Nexiva Closed IV Catheter System (Single Port) with MaxZero…Becton, Dickinson and Company · 2017
- Descendant · 4 generations on · 5 of its ownK183399 BD Nexiva Closed IV Catheter SystemBecton Dickinson Infusion Therapy Systems, Inc. · 2019 · Recalled
Showing 5 of 31 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 31 · See all- K100434Clave NeutronIcu Medical, Inc. · 2010 · cites it directly28descendants
- K132413MZ1000Care Fusion · 2013 · 2 generations on26descendants
- K140831Maxzero Extension Set with Needless ConnectorCare Fusion · 2014 · 3 generations on13descendants
- K170336BD Nexiva Closed IV Catheter System (Single Port) with MaxZero Needleless ConnectorBecton, Dickinson and Company · 2017 · 3 generations on6descendants
- K183399BD Nexiva Closed IV Catheter SystemBecton Dickinson Infusion Therapy Systems, Inc. · 2019 · 4 generations on5descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K942988 | Filtered Extension Sets | B. Braun of America, Inc. | 1995 | 32 |
| −2 | K955585 | V2 Injection Site | B.Braun Medical, Inc. | 1996 | 27 |
| −1 | K031923 | Ultrasite Valve | B.Braun Medical, Inc. | 2003 | — |
| 0 (this device) | K083723 | A6 Luer Access Device | B.Braun Medical, Inc. | 2009 | — |
| +1 | K100434 | Clave Neutron | Icu Medical, Inc. | 2010 | 3 |
| +2 | K132413 | MZ1000 | Care Fusion | 2013 | — |
| +3 | K140831 | Maxzero Extension Set with Needless Connector | Care Fusion | 2014 | — |
| +3 | K170336 | BD Nexiva Closed IV Catheter System (Single Port) with MaxZero Needleless… | Becton, Dickinson and Company | 2017 | — |
| +4 | K183399 | BD Nexiva Closed IV Catheter System | Becton Dickinson Infusion Therapy Systems, Inc. | 2019 | 3 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 3 | 0 |
| 2016 | 1 | 0 |
| 2017 | 4 | 0 |
| 2019 | 2 | 0 |
| 2020 | 2 | 0 |
| 2021 | 6 | 0 |
| 2022 | 2 | 0 |
| 2023 | 3 | 0 |
| 2024 | 1 | 0 |
| 2025 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FPA device
- #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K031923 | Ultrasite Valve | 2003 |
| * | K955585 | V2 Injection SiteSummary page 1…Braun Medical Inc. Ultrasite®Valve [for uise.with power injecotors. (K~03 1923), previously known as B. Braun Medica n.V2Ijction Site (K955585)] DESCRIPTION: B.~ Braun's A6 Luer Access Device is a poitv dipacement valve intended to provide needle-fre… | 1996 |
What descends from it? (31)
See all 312 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 29, 2025 | Class II | Z-0691-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0703-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0685-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0661-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0695-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0660-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0684-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0619-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0676-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0652-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0653-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0615-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0658-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0688-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0612-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0698-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0664-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0690-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0624-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0716-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0692-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0693-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0633-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0699-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0654-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0627-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0622-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0657-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0656-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0625-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0662-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0659-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0689-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0616-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0694-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0663-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0683-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0617-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0687-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0607-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0696-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0628-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0715-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0610-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0706-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0704-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0614-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0609-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0655-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| Oct 29, 2025 | Class II | Z-0626-2026 | Process control | Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
The 5 latest of 50 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 2 companies; a new generation was cleared every 5.8 years on median.
- 2 of the 3 ancestors have a recall of any class.
- Of its descendants, K100434 (Icu Medical, Inc., 2010) has gone furthest, with 28 descendants of its own.
How does its category compare?
All FPA devicesFPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · FPA
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FPA
225in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FPA
98in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · FPA
34,506in 2025
MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 75 | 17 | 1,499 |
| 2008 | 18 | 89 | 15 | 1,941 |
| 2009 | 10 | 166 | 166 | 2,512 |
| 2010 | 23 | 120 | 58 | 2,537 |
| 2011 | 12 | 87 | 28 | 4,407 |
| 2012 | 21 | 96 | 16 (4) | 4,930 |
| 2013 | 13 | 90 | 64 | 4,651 |
| 2014 | 22 | 89 | 24 | 4,695 |
| 2015 | 17 | 123 | 6 | 6,625 |
| 2016 | 12 | 195 | 10 | 8,470 |
| 2017 | 17 | 240 | 13 | 6,462 |
| 2018 | 4 | 177 | 4 | 6,574 |
| 2019 | 7 | 161 | 13 | 11,227 |
| 2020 | 10 | 327 | 107 | 9,563 |
| 2021 | 13 | 186 | 1 | 10,414 |
| 2022 | 14 | 213 | 9 (5) | 13,393 |
| 2023 | 14 | 340 | 12 (4) | 10,916 |
| 2024 | 7 | 244 | 13 (1) | 47,210 |
| 2025 | 11 | 225 | 98 (1) | 34,506 |
| 2026 | 5 | 266 | 10 (2) | — |
What is it indicated for?
The A6 Luer Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. The A6 Luer Access Device may be used with power injectors at a maximum pressure of 300 psi and a maximum flow rate of 10 ml/sec.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K083723 · printed Nov 5, 2026
Cite this page
510(k) Report. “K083723: A6 Luer Access Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K083723
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026