510(k) · Cleared
Recalled · Class II · 2020SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/ Scope Power
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solutions USA, Inc. · JAK System, X-Ray, Tomography, Computed
Cleared on Nov 24, 2015 as substantially equivalent to K133589 (Siemens Medical Solutions USA, Inc., 2014) and 10 other predicates after a 406-day review. Its family tree reaches back 8 generations to 2001. 27 later devices descend from it. It has been recalled 16 times.
- Decided
- Nov 24, 2015
- Review
- 406 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 16
- Predicates
- 11
Its longest line of descent runs back 8 generations to K013561 (2001). 29 ancestors in all, the oldest K940606 (1994). 27 later devices descend from it, 5 citing it directly.
- Root · 8 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
- 7 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
- 6 generations backK040372 Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC · 2004 · Recalled
- 5 generations backK052216 Somatom DefinitionSiemens Medical Systems, Inc. · 2005
- 4 generations backK082220 Somatom Flash DSSiemens Medical Solutions USA, Inc. · 2008
- 3 generations backK113342 Somatom Definition FlashSiemens Medical Systems, Inc. · 2011 · Recalled
- 2 generations backK121072 Somatom Definition FlashSiemens Medical Systems, Inc. · 2012 · Recalled
- 1 generation backK133589 Somatom ForceSiemens Medical Solutions USA, Inc. · 2014 · Recalled
- This deviceK142955 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge…Siemens Medical Solutions USA, Inc. · 2015
- Descendant · cites it directly · 24 of its ownK161196 SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · Recalled
- Descendant · cites it directly · 21 of its ownK162302 Somatom Confidence ® RT ProSiemens Medical Solutions USA, Inc. · 2016 · Recalled
- Descendant · cites it directly · 20 of its ownK173630 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM…Siemens Medical Solutions USA, Inc. · 2018 · Recalled
- Descendant · 2 generations on · 19 of its ownK190578 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM…Siemens Medical Solutions USA, Inc. · 2019 · Recalled
- Descendant · 2 generations on · 19 of its ownK163297 Scan&GOSiemens Medical Solutions USA, Inc. · 2017
Showing 5 of 27 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 27 · See all- K161196SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · cites it directly24descendants
- K162302Somatom Confidence ® RT ProSiemens Medical Solutions USA, Inc. · 2016 · cites it directly21descendants
- K173630SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc. · 2018 · cites it directly20descendants
- K190578SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc. · 2019 · 2 generations on19descendants
- K163297Scan&GOSiemens Medical Solutions USA, Inc. · 2017 · 2 generations on19descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −8 | K013561 | Lightspeed 4.0 CT Scanner System | GE Medical Systems | 2001 | 5 |
| −7 | K030420 | Lightspeed 5.0 Computed Tomography System | GE Medical Systems | 2003 | 4 |
| −6 | K040372 | Lightspeed 7.0 CT Scanner System | Ge Medical Systems, LLC | 2004 | 5 |
| −5 | K052216 | Somatom Definition | Siemens Medical Systems, Inc. | 2005 | — |
| −4 | K082220 | Somatom Flash DS | Siemens Medical Solutions USA, Inc. | 2008 | — |
| −3 | K113342 | Somatom Definition Flash | Siemens Medical Systems, Inc. | 2011 | 1 |
| −2 | K121072 | Somatom Definition Flash | Siemens Medical Systems, Inc. | 2012 | 5 |
| −1 | K133589 | Somatom Force | Siemens Medical Solutions USA, Inc. | 2014 | 9 |
| 0 (this device) | K142955 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM… | Siemens Medical Solutions USA, Inc. | 2015 | — |
| +1 | K161196 | SOMATOM Drive | Siemens Medical Solutions USA, Inc. | 2016 | 4 |
| +1 | K162302 | Somatom Confidence ® RT Pro | Siemens Medical Solutions USA, Inc. | 2016 | 1 |
| +1 | K173630 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition… | Siemens Medical Solutions USA, Inc. | 2018 | 4 |
| +2 | K190578 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition… | Siemens Medical Solutions USA, Inc. | 2019 | 13 |
| +2 | K163297 | Scan&GO | Siemens Medical Solutions USA, Inc. | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
| 1998 | 3 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2008 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 3 | 0 |
| 2014 | 4 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 2 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 3 | 0 |
| 2024 | 3 | 0 |
| 2025 | 4 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (11)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K133589 | Somatom ForceSummary page 18…Class II Product Code: 90JAK Legally Marketed Predicate Devices Trade Name: SOMATOM Definition Force 510(k)#: K133589 Clearance Date: April 17, 2014 Classification Name: System, X-Ray, Tomography, Computed Classification Panel… | 2014 |
| * | K121072 | Somatom Definition FlashSummary page 19…21 CFR § 892.1750 Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Definition Flash 510(k)#: K121072 Clearance Date: May 08, 2012 Classification Name: System, X-Ray, Tomography, Computed Classification Panel:… | 2012 |
| * | K120579 | Somatom Definition EdgeSummary page 19…21 CFR § 892.1750 Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Definition Edge 510(k)#: K120579 Clearance Date: May 23, 2012 Classification Name: System, X-Ray, Tomography, Computed Classification Panel:… | 2012 |
| * | K081022 | Somatom Definition, Model as/as+Summary page 18…1 CFR § 892.1750 Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Definition AS/AS+ 510(k)#: K081022 Clearance Date: June 02, 2008 Classification Name: System, X-Ray, Tomography, Computed Classification Panel:… | 2008 |
| * | K122471 | Somatom DefinitionSummary page 19…on: 21 CFR § 892.1750 Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Definition 510(k)#: K122471 Clearance Date: September 11, 2012 Classification Name: System, X-Ray, Tomography, Computed Classification P… | 2012 |
| * | K130901 | Somatom Definition as OpenSummary page 19…Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Definition AS Open 510(k)#: K130901 Clearance Date: January 02, 2014 Classification Name: System, X-Ray, Tomography, Computed Classification Pan… | 2014 |
| * | K133424 | Somatom Emotion 6 CT Systems, Somatom Emotion 16Summary page 19…FR § 892.1750 Device Class: II Product Code: 90 JAK Trade Name: SOMATOM Emotion 16 CT System 510(k)#: K133424 Clearance Date: January 17, 2014 Classification Name: Computed Tomography X-ray System Classification Panel:… | 2014 |
| * | K133590 | Somatom PerspectiveSummary page 20…e 20 === Product Code: 90JAK Trade Name: SOMATOM Perspective (16/32) 510(k)#: K133590 Clearance Date: April 30, 2014 Classification Name: Computed Tomography X-ray Syst… | 2014 |
| * | K113287 | Somatom PerspectiveSummary page 19…892.1750 Device Class: Class II Product Code: 90JAK Trade Name: SOMATOM Perspective (128/64) 510(k)#: K113287 Clearance Date: May 23, 2012 Classification Name: Computed Tomography X-ray System Classification Panel:… | 2012 |
| * | K140912 | Somatom Scope/Somatom Scope PowerSummary page 20…II Product Code: 90JAK Trade Name: SOMATOM Scope/SOMATOM Scope Power 510(k)#: K140912 Clearance Date: September 15, 2014 Classification Name: Computed Tomography X-ray… | 2014 |
| * | K982349 | Somatom Plus 4WITH Volume Zoom CT ScannersSummary page 20…Trade Name: SOMATOM Plus 4WITH Volume Zoom CT Scanners 510(k)#: K982349 Clearance Date: September 30, 1998 Classification Name: Computed Tomography X-ray… | 1998 |
What descends from it? (27)
See all 275 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K161196 | SOMATOM Drive | 2016 | 24 |
| K162302 | Somatom Confidence ® RT Pro | 2016 | 21 |
| K173630 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM… | 2018 | 20 |
| K163296 | SOMATOM go.Up, SOMATOM go.Now | 2017 | 19 |
| K173023 | Symbia T16, Symbia Intevo 16 and Symbia Intevo Bold | 2017 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 24, 2020 | Class II | Z-0771-2021 | Software design | System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan |
| Jun 22, 2020 | Class II | Z-3025-2020 | Software design | Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan |
| Oct 7, 2019 | Class II | Z-0406-2020 | Software design | Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes |
| Feb 15, 2018 | Class II | Z-1446-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1449-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1455-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1456-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1450-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1445-2018 | Software design | To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1457-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1458-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1447-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Nov 8, 2017 | Class II | Z-1745-2018 | Device Design | There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM… |
| Nov 8, 2017 | Class II | Z-1747-2018 | Device Design | There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM… |
| Nov 8, 2017 | Class II | Z-1746-2018 | Device Design | There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM… |
| Nov 8, 2017 | Class II | Z-1748-2018 | Device Design | There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM… |
The 5 latest of 16 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 1.6 years on median.
- 22 of the 29 ancestors have a recall of any class.
- Of its descendants, K161196 (Siemens Medical Solutions USA, Inc., 2016) has gone furthest, with 24 descendants of its own.
- 2 other devices also cite its predicate K133589.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the…
Read the full statement (1,842 words)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Definition AS/ AS+ (Project P46) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Definition AS Open systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Definition Flash system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
In addition the SOMATOM Definition Flash is able to produce additional image planes and analysis results by executing optional post processing features, which operate on DICOM images.
The images and results delivered by the system can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Definition Edge (Project P46F) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The SOMATOM P45 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The SOMATOM Emotion 16 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Perspective 64/128 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
In addition the SOMATOM Perspective is able to produce additional image planes and analysis results by executing optional post-processing features, which operate on DICOM images. For cardiac imaging, which is an option on the system, a new reconstruction algorithm (iTRIM - Iterative Temporal Resolution Improvement Method) is used. iTRIM improves the temporal resolution of cardiac CT images compared to conventional cardiac CT image reconstruction. Actual results obtained with iTRIM can vary depending on the particular clinical situation.
The images and results delivered by the SOMATOM Perspective can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Siemens SOMATOM Perspective 16/32 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
In addition the SOMATOM Perspective is able to produce additional image planes and analysis results by executing optional post-processing features, which operate on DICOM images. For cardiac imaging, which is an option on the system, a new reconstruction algorithm (iTRIM - Iterative Temporal Resolution Improvement Method) is used. iTRIM improves the temporal resolution of cardiac CT images compared to conventional cardiac CT image reconstruction. Actual results obtained with iTRIM can vary depending on the particular clinical situation.
The images and results delivered by the SOMATOM Perspective can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The SOMATOM Scope and the SOMATOM Scope Power are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles.
The images and results delivered by the SOMATOM Scope and the SOMATOM Scope Power can be used by a trained physician as an aid in diagnosis.
(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Somatom Plus 4 with Volume Zoom package is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K142955 · printed Nov 6, 2026
Cite this page
510(k) Report. “K142955: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/ Scope Power.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142955