510(k) · Cleared
Recalled · Class II · 2016Syngo X-workplace SW VD10
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medi Cal Solutions, Inc. · LLZ System, Image Processing, Radiological
Cleared on Feb 24, 2015 as substantially equivalent to K042995 (Siemens Medical Solutions USA, Inc., 2004) after a 97-day review. Its family tree reaches back 3 generations to 2001. 39 later devices descend from it. It has been recalled 2 times.
- Decided
- Feb 24, 2015
- Review
- 97 days
- Class
- Class II
- Product code
- LLZ
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 3 generations to K010938 (2001). 39 later devices descend from it, 1 citing it directly.
- Root · 3 generations backK010938 Syngo Multimodality WorkstationSiemens Medical Solutions USA, Inc. · 2001 · Recalled
- 2 generations backK040970 LeonardoSiemens Medical Systems, Inc. · 2004 · Recalled
- 1 generation backK042995 X-LeonardoSiemens Medical Solutions USA, Inc. · 2004
- This deviceK143319 Syngo X-workplace SW VD10Siemens Medi Cal Solutions, Inc. · 2015
- Descendant · cites it directly · 38 of its ownK153346 syngo Application Software VD11Siemens Medical Solutions USA, Inc. · 2016
- Descendant · 2 generations on · 35 of its ownK162541 syngo Application SoftwareSiemens Medical Solutions USA, Inc. · 2016
- Descendant · 3 generations on · 23 of its ownK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · Recalled
- Descendant · 4 generations on · 10 of its ownK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled
- Descendant · 4 generations on · 10 of its ownK181550 Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · Recalled
Showing 5 of 39 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 39 · See all- K153346syngo Application Software VD11Siemens Medical Solutions USA, Inc. · 2016 · cites it directly38descendants
- K162541syngo Application SoftwareSiemens Medical Solutions USA, Inc. · 2016 · 2 generations on35descendants
- K163286ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · 3 generations on23descendants
- K190768ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · 4 generations on10descendants
- K181550Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · 4 generations on10descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K010938 | Syngo Multimodality Workstation | Siemens Medical Solutions USA, Inc. | 2001 | 5 |
| −2 | K040970 | Leonardo | Siemens Medical Systems, Inc. | 2004 | 1 |
| −1 | K042995 | X-Leonardo | Siemens Medical Solutions USA, Inc. | 2004 | — |
| 0 (this device) | K143319 | Syngo X-workplace SW VD10 | Siemens Medi Cal Solutions, Inc. | 2015 | — |
| +1 | K153346 | syngo Application Software VD11 | Siemens Medical Solutions USA, Inc. | 2016 | — |
| +2 | K162541 | syngo Application Software | Siemens Medical Solutions USA, Inc. | 2016 | — |
| +3 | K163286 | ARTIS pheno | Siemens Medical Solution USA, Inc. | 2017 | 14 |
| +4 | K190768 | ARTIS icono | Siemens Medical Solutions USA, Inc. | 2019 | 4 |
| +4 | K181550 | Cios Spin | Siemens Medical Solutions USA, Inc. | 2018 | 4 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 2 | 0 |
| 2017 | 3 | 0 |
| 2018 | 3 | 0 |
| 2019 | 2 | 0 |
| 2020 | 4 | 0 |
| 2021 | 4 | 0 |
| 2022 | 5 | 0 |
| 2023 | 3 | 0 |
| 2024 | 3 | 0 |
| 2025 | 6 | 0 |
| 2026 | 4 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LLZ device
- #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K042995 | X-LeonardoSummary page 5…510(k) Submission: syngo X Workplace 4. Legally Marketed Predicate Device Trade Name: X-LEONARDO Workstation 510(k) Clearance K042995 Clearance Date November 24, 2004 Classification Name: Picture Archiving and Communications Syst… | 2004 |
What descends from it? (39)
See all 391 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K153346 | syngo Application Software VD11 | 2016 | 38 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 24, 2016 | Class II | Z-1232-2016 | Software design | After importing, the segmentation results appear mirrored at the CARTO system and can't be used for the ablation procedure. |
| Nov 18, 2015 | Class II | Z-0597-2016 | Software design | Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
- 2 of the 3 ancestors have a recall of any class.
- Of its descendants, K153346 (Siemens Medical Solutions USA, Inc., 2016) has gone furthest, with 38 descendants of its own.
- 1 other device also cites its predicate K042995.
How does its category compare?
All LLZ devicesLLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · LLZ
58in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LLZ
173in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LLZ
11in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LLZ
116in 2025
MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 76 | 36 | 20 | 152 |
| 2008 | 94 | 52 | 32 | 154 |
| 2009 | 70 | 62 | 21 | 145 |
| 2010 | 51 | 98 | 17 | 145 |
| 2011 | 106 | 79 | 30 | 454 |
| 2012 | 82 | 87 | 12 | 266 |
| 2013 | 95 | 84 | 64 | 1,130 |
| 2014 | 78 | 89 | 6 | 15,631 |
| 2015 | 64 | 90 | 16 | 211 |
| 2016 | 82 | 90 | 12 | 290 |
| 2017 | 78 | 107 | 5 | 104 |
| 2018 | 104 | 105 | 7 | 72 |
| 2019 | 74 | 107 | 2 | 88 |
| 2020 | 81 | 115 | 6 | 82 |
| 2021 | 83 | 116 | 3 | 70 |
| 2022 | 65 | 137 | 4 | 28 |
| 2023 | 87 | 157 | 8 | 26 |
| 2024 | 62 | 131 | 6 | 42 |
| 2025 | 58 | 173 | 11 | 116 |
| 2026 | 45 | 110 | 14 | — |
What is it indicated for?
The syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures. The syngo X Workplace can be configured as stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured…
Read the full statement (200 words)
The syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures.
The syngo X Workplace can be configured as stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, the X Workplace can be combined with systems of the Artis family. In combination with an Artis system., it provides image guided solutions in the operating room, for image guided surgery, by Image Fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional oncology, interventional radiology. and interventional neuroradiology.
The syngo X Workplace can also be combined with fluoroscopy systems or Radiographic systems.
The syngo X Workplace can be configured with a variety of syngo or Windows-based software options, which are intended to assist the physician in diagnosis, treatment planning and treatment control. It includes commercially available post-processing techniques and OEM options.
Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K143319 · printed Nov 6, 2026
Cite this page
510(k) Report. “K143319: Syngo X-workplace SW VD10.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K143319