Siemens Medical Solution USA, Inc. · OWB Interventional Fluoroscopic X-Ray System
Cleared on Mar 9, 2017 as substantially equivalent to K123529 (Siemens Medical Solutions USA, Inc., 2013) and 2 other predicates after a 107-day review. Its family tree reaches back 9 generations to 1980. 23 later devices descend from it. It has been recalled 14 times.
- Decided
- Mar 9, 2017
- Review
- 107 days
- Class
- Class II
- Product code
- OWB
- Recalls
- 14
- Predicates
- 3
Its longest line of descent runs back 9 generations to K801623 (1980). 50 ancestors in all. 23 later devices descend from it, 4 citing it directly.
- Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 5 generations between
- 3 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
- 2 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
- 1 generation backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
- This deviceK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017
- Descendant · cites it directly · 10 of its ownK190768 ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · Recalled
- Descendant · cites it directly · 10 of its ownK181550 Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · Recalled
- Descendant · cites it directly · 10 of its ownK181560 Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · Recalled
- Descendant · 2 generations on · 9 of its ownK203504 Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · Recalled
- Descendant · cites it directly · 5 of its ownK201156 ARTIS pheno (VE2)Siemens Medical Solution USA, Inc. · 2020 · Recalled
Show all 9 generations
- Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 8 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
- 7 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
- 6 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
- 5 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
- 4 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
- 3 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
- 2 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
- 1 generation backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
- This deviceK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017
Showing 5 of 23 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 23 · See all- K190768ARTIS iconoSiemens Medical Solutions USA, Inc. · 2019 · cites it directly10descendants
- K181550Cios SpinSiemens Medical Solutions USA, Inc. · 2018 · cites it directly10descendants
- K181560Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · cites it directly10descendants
- K203504Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · 2 generations on9descendants
- K201156ARTIS pheno (VE2)Siemens Medical Solution USA, Inc. · 2020 · cites it directly5descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K801623 | Cavitron Ultrasonic Surg. Aspir. Cusa | Cavitron Corp. | 1980 | — |
| −8 | K962952 | Olympus Ultrasonic Surgical System (Usu) | Olympus Optical, Co. | 1996 | — |
| −7 | K972114 | Olympus Sonosurg System | The Olympus Optical Co. | 1998 | — |
| −6 | K981993 | Olympus Integrated Endosurgery System Endoalpha | The Olympus Optical Co. | 1998 | — |
| −5 | K994231 | Siemens Integrated Operating System (Sios) | Siemens Medical Corp. | 2000 | — |
| −4 | K052202 | Axiom Artis - Modular Angiographic System | Siemens Medical Solutions USA, Inc. | 2006 | 29 |
| −3 | K073290 | Artis Zee and Zeego | Siemens Medical Solutions USA, Inc. | 2008 | 39 |
| −2 | K090745 | Artis Zee / Zeegos with SW VC14 | Siemens Medical Solutions USA, Inc. | 2009 | 7 |
| −1 | K123529 | Artis Q and Q.Zen- Modular Angiographic System | Siemens Medical Solutions USA, Inc. | 2013 | 31 |
| 0 (this device) | K163286 | ARTIS pheno | Siemens Medical Solution USA, Inc. | 2017 | — |
| +1 | K190768 | ARTIS icono | Siemens Medical Solutions USA, Inc. | 2019 | 4 |
| +1 | K181550 | Cios Spin | Siemens Medical Solutions USA, Inc. | 2018 | 4 |
| +1 | K181560 | Cios Alpha | Siemens Medical Systems USA, Inc. | 2018 | 7 |
| +2 | K203504 | Cios Flow | Siemens Medical Solutions USA, Inc. | 2020 | 1 |
| +1 | K201156 | ARTIS pheno (VE2) | Siemens Medical Solution USA, Inc. | 2020 | 6 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1989 | 1 | 0 |
| 1991 | 3 | 0 |
| 1992 | 1 | 0 |
| 1993 | 2 | 0 |
| 1994 | 1 | 0 |
| 1995 | 10 | 0 |
| 1996 | 3 | 0 |
| 1997 | 2 | 0 |
| 1999 | 1 | 0 |
| 2000 | 3 | 0 |
| 2001 | 2 | 0 |
| 2002 | 3 | 0 |
| 2004 | 4 | 0 |
| 2013 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 3 | 0 |
| 2021 | 3 | 0 |
| 2022 | 4 | 0 |
| 2023 | 2 | 0 |
| 2024 | 2 | 0 |
| 2025 | 3 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWB device
- #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.2 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K123529 | Artis Q and Q.Zen- Modular Angiographic SystemSummary page 4…Device Trade Name: Artis Q and Artis Q.zen - Modular Angiographic System 510(k) Clearance K123529 Clearance Date February 12, 2013 Classification Name: Interventional… | 2013 |
| * | K122644 | Artis Zee/Zeego with CSX-10 Detector SW VC21Summary page 5…ndary Predicate Device Trade Name: Artis zee/zeego with CSX-10 Detector SW VC21 510(k) Clearance K122644 Clearance Date May 16, 2013 Classification Name: Interventional fluoroscopic X-Ray… | 2013 |
| * | K162541 | syngo Application SoftwareSummary page 5…le issues. Legally Marketed Secondary Predicate Device Trade Name: syngo Application Software 510(k) Clearance K162541 Clearance Date November 16, 2016 Classification Name: System, Image processing, Rad… | 2016 |
What descends from it? (23)
See all 234 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 2, 2023 | Class II | Z-1498-2023 | Software design | If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach… |
| Feb 2, 2023 | Class II | Z-1311-2023 | Software design | In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment. |
| Dec 16, 2022 | Class II | Z-0956-2023 | Device Design | Mechanical connection between the tabletop and table base may be lost. |
| Sep 20, 2022 | Class II | Z-0119-2023 | Software design | Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be… |
| Feb 25, 2022 | Class II | Z-1339-2022 | Software design | After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even… |
| May 27, 2021 | Class II | Z-2029-2021 | Software design | Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9") result in canceling… |
| Apr 22, 2020 | Class II | Z-2108-2020 | Software design | If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only… |
| Jan 24, 2020 | Class II | Z-1340-2020 | Device Design | Potential exists for one or more axes of the robotic stand to become blocked by their safety brakes; In the event the axis is blocked by its safety brake, the robotic stand can no longer perform the intended movements |
| Dec 20, 2019 | Class II | Z-0807-2020 | Device Design | The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred. |
| Jun 6, 2019 | Class II | Z-1905-2019 | Device Design | Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work… |
| Jul 29, 2018 | Class II | Z-3190-2018 | Device Design | Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g.… |
| Jun 22, 2018 | Class II | Z-2748-2018 | Nonconforming Material/Component | Potential hardware issue which may cause thermal effects and possible damage to system parts. |
| May 30, 2018 | Class II | Z-2524-2018 | Device Design | During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm… |
| May 1, 2018 | Class II | Z-2283-2018 | Device Design | If the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 ). Under certain… |
The 5 latest of 14 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 37 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
- 13 of the 50 ancestors have a recall of any class.
- Of its descendants, K190768 (Siemens Medical Solutions USA, Inc., 2019) has gone furthest, with 10 descendants of its own.
- 4 other devices also cite its predicate K123529.
How does its category compare?
All OWB devicesOWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · OWB
22in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWB
148in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWB
42so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWB
14,008in 2025
MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 52 | 0 | 1 |
| 2008 | 4 | 135 | 1 | 0 |
| 2009 | 9 | 99 | 0 | 0 |
| 2010 | 7 | 66 | 0 | 0 |
| 2011 | 6 | 79 | 0 | 1 |
| 2012 | 6 | 37 | 0 | 1 |
| 2013 | 11 | 119 | 10 | 27 |
| 2014 | 15 | 158 | 9 | 103 |
| 2015 | 11 | 66 | 35 | 2,134 |
| 2016 | 16 | 103 | 13 | 4,074 |
| 2017 | 20 | 76 | 13 | 1,604 |
| 2018 | 23 | 54 | 6 | 1,449 |
| 2019 | 9 | 174 | 7 | 1,558 |
| 2020 | 16 | 53 | 8 | 1,581 |
| 2021 | 24 | 60 | 31 | 1,981 |
| 2022 | 21 | 90 | 24 | 2,932 |
| 2023 | 12 | 152 | 22 | 7,604 |
| 2024 | 13 | 126 | 11 | 7,942 |
| 2025 | 22 | 148 | 30 | 14,008 |
| 2026 | 15 | 221 | 42 | — |
What is it indicated for?
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table…
Read the full statement (185 words)
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image-guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.
ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.
The ARTIS family include also the software option DynaCT with following indications for use:
DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K163286 · printed Nov 6, 2026
Cite this page
510(k) Report. “K163286: ARTIS pheno.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K163286