510(k) Report

510(k) · Cleared

Cleared

CORA (Coagulation Resonance Analysis) System

K150041510(k) TraditionalCleared Jun 26, 2015168-day reviewClass II21 CFR 864.5425

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Coramed Technologies, LLC · JPA System, Multipurpose For In Vitro Coagulation Studies

Cleared on Jun 26, 2015 as substantially equivalent to K002177 (Haemoscope Corp., 2000) after a 168-day review. Its family tree reaches back 1 generation to 2000. 4 later devices descend from it.

Decided
Jun 26, 2015
Review
168 days
Class
Class II
Product code
JPA
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K002177 (2000). 4 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2017: 1 descendant cleared2024: 1 descendant cleared2025: 2 descendants clearedK002177 · Thrombelastograph Coagulation Analyzer Teg-5000 Series · Haemoscope Corp. · 2000 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 1 descendant K002177 · Thrombelastograph Coagulation Analyzer Teg-5000 Series · Haemoscope Corp. · 2000 · recalled0
  1. Root · 1 generation backK002177 Thrombelastograph Coagulation Analyzer Teg-5000 SeriesHaemoscope Corp. · 2000 · Recalled
  2. This deviceK150041 CORA (Coagulation Resonance Analysis) SystemCoramed Technologies, LLC · 2015
  3. Descendant · cites it directly · 1 of its ownK243858 TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics Corporation · 2025
  4. Descendant · 2 generations on · 0 of its ownK251024 TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics · 2025
  5. Descendant · cites it directly · 0 of its ownK232018 Citrated: K, KH, RTH, FFHHaemonetics Corporation · 2024
  6. Descendant · cites it directly · 0 of its ownK160502 TEG 6s Hemostasis SystemHaemonetics Corporation · 2017

Most influential descendants

Showing 4 of 4 · See all
  • K243858TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics Corporation · 2025 · cites it directly1descendant
  • K251024TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics · 2025 · 2 generations on0descendants
  • K232018Citrated: K, KH, RTH, FFHHaemonetics Corporation · 2024 · cites it directly0descendants
  • K160502TEG 6s Hemostasis SystemHaemonetics Corporation · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K002177Thrombelastograph Coagulation Analyzer Teg-5000 SeriesHaemoscope Corp.20007
0 (this device)K150041CORA (Coagulation Resonance Analysis) SystemCoramed Technologies, LLC2015—
+1K243858TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics Corporation2025—
+2K251024TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics2025—
+1K232018Citrated: K, KH, RTH, FFHHaemonetics Corporation2024—
+1K160502TEG 6s Hemostasis SystemHaemonetics Corporation2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202410
202520
FIG. 1 — Lineage of K150041, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JPA device

  1. #1 K243858 TEG 6s Hemostasis System Citrated: K, KH, RT…score 12.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002177Thrombelastograph Coagulation Analyzer Teg-5000 Series
Summary page 6
…n Name: 21CFR 864.5425 Multipurpose System for in vitro Coagulation Studies PREDICATE DEVICE o Thrombelastograph® Coagulation Analyzer (TEG®) – 5000 Series, K002177, Product Code JPA (System, Multipurpose, for In Vitro Coagulation Studies), Haemoscope Corporation ===…
Haemoscope Corp.20007

Also cleared against K002177: 3 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243858TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay CartridgeHaemonetics Corporation20251
K232018Citrated: K, KH, RTH, FFHHaemonetics Corporation20240
K160502TEG 6s Hemostasis SystemHaemonetics Corporation20170

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K243858 (Haemonetics Corporation, 2025) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K002177.

How does its category compare?

All JPA devices

JPA System, Multipurpose For In Vitro Coagulation Studies: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JPA

3in 2025

2007: 12008: 22009: 62010: 12011: 12013: 12015: 22016: 62017: 42018: 12019: 32021: 12022: 12023: 32024: 12025: 32026 (partial year): 1
2007high 6 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JPA

30in 2025

2009: 91 days median2016: 114 days median2017: 117 days median2019: 175 days median2023: 506 days median2025: 30 days median
2007high 506 (2023) · low 30 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JPA

1in 2020

2007: 12010: 62011: 42012: 12016: 32020: 12026 (partial year): 1
2007high 6 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JPA

11in 2025

2007: 1,7442008: 9182009: 1,3632010: 7272011: 1612012: 522013: 432014: 482015: 452016: 62017: 152018: 392019: 272020: 182021: 72022: 92023: 112024: 82025: 11
2007high 1,744 (2007) · low 6 (2016)2025

MAUDE reports naming product code JPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—11,744
20082—0918
200969101,363
20101—6727
20111—4161
20120—152
20131—043
20140—048
20152—045
2016611436
20174117015
20181—039
20193175027
20200—118
20211—07
20221—09
20233506011
20241—08
2025330011
20261—1—

What is it indicated for?

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The CORA System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The CORA System provides specific blood modifiers, in the form of reagents…

Read the full statement (162 words)

The CORA System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The CORA System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV.

The CORA System provides specific blood modifiers, in the form of reagents dried-in-place within CORA Cartridges.

Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

The indication for CORA System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K150041: CORA (Coagulation Resonance Analysis) System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150041

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026