510(k) · Cleared
ClearedTEG 6s Hemostasis System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Haemonetics Corporation · JPA System, Multipurpose For In Vitro Coagulation Studies
Cleared on Apr 19, 2017 as substantially equivalent to K150041 (Coramed Technologies, LLC, 2015) after a 421-day review. Its family tree reaches back 2 generations to 2000.
- Decided
- Apr 19, 2017
- Review
- 421 days
- Class
- Class II
- Product code
- JPA
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 2 generations to K002177 (2000). No later device cites it as a predicate yet.
View FIG. 1 as a table
Line of descent and top descendants
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JPA device
- #1 K243858 TEG 6s Hemostasis System Citrated: K, KH, RT…score 12.0 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K150041 | CORA (Coagulation Resonance Analysis) SystemSummary page 4…II Predicate Device Trade Name: TEG 6s Hemostasis System 510(k) K140893 clearance Date Jan 3, 2015 TEG 6s Hemostasis System 510(k) K150041 clearance Date June 26, 2015 Device Characteristics Summary: The TEG6s analyzer is a four-channel in vitro diagnostic instrumen… | 2015 |
| † | K140893 | Cora (Coagulation Resonance Analysis) System with Platelet Mapping Assay | 2015 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 14.8 years on median.
- 1 of the 2 ancestors has a recall of any class.
- 2 other devices also cite its predicate K150041.
How does its category compare?
All JPA devicesJPA System, Multipurpose For In Vitro Coagulation Studies: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · JPA
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JPA
30in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JPA
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JPA
11in 2025
MAUDE reports naming product code JPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 1 | 1,744 |
| 2008 | 2 | — | 0 | 918 |
| 2009 | 6 | 91 | 0 | 1,363 |
| 2010 | 1 | — | 6 | 727 |
| 2011 | 1 | — | 4 | 161 |
| 2012 | 0 | — | 1 | 52 |
| 2013 | 1 | — | 0 | 43 |
| 2014 | 0 | — | 0 | 48 |
| 2015 | 2 | — | 0 | 45 |
| 2016 | 6 | 114 | 3 | 6 |
| 2017 | 4 | 117 | 0 | 15 |
| 2018 | 1 | — | 0 | 39 |
| 2019 | 3 | 175 | 0 | 27 |
| 2020 | 0 | — | 1 | 18 |
| 2021 | 1 | — | 0 | 7 |
| 2022 | 1 | — | 0 | 9 |
| 2023 | 3 | 506 | 0 | 11 |
| 2024 | 1 | — | 0 | 8 |
| 2025 | 3 | 30 | 0 | 11 |
| 2026 | 1 | — | 1 | — |
What is it indicated for?
The TEG 6s System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The TEG 6s System records the kinetic changes in a venous sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The TEG 6s System provides specific blood modifiers, in the form of…
Read the full statement (170 words)
The TEG 6s System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The TEG 6s System records the kinetic changes in a venous sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The TEG 6s System provides specific blood modifiers, in the form of reagents dried-in-place within TEG 6s cartridges.
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.
The indication for use for the TEG 6s System is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K160502 · printed Nov 6, 2026
Cite this page
510(k) Report. “K160502: TEG 6s Hemostasis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160502
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026