Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator

K151767510(k) TraditionalCleared Dec 11, 2015164-day reviewClass II21 CFR 862.1665Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics, Inc. · JGS Electrode, Ion Specific, Sodium

Cleared on Dec 11, 2015 as substantially equivalent to K990346 (Bayer Corp., 1999) and 1 other predicate after a 164-day review. Its family tree reaches back 1 generation to 1999. 3 later devices descend from it. It has been recalled 12 times.

Decided
Dec 11, 2015
Review
164 days
Class
Class II
Product code
JGS
Recalls
12
Predicates
2

Its longest line of descent runs back 1 generation to K990346 (1999). 2 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2023: 2 descendants cleared2017: 1 descendant clearedK990346 · Bayer Advia 1650 Chemistry System · Bayer Corp. · 1999 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants K990346 · Bayer Advia 1650 Chemistry System · Bayer Corp. · 1999 · recalled0
  1. Root · 1 generation backK990346 Bayer Advia 1650 Chemistry SystemBayer Corp. · 1999 · Recalled
  2. This deviceK151767 Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI)…Siemens Healthcare Diagnostics, Inc. · 2015
  3. Descendant · cites it directly · 0 of its ownK222438 A-LYTE® Integrated Multisensor (IMT Na K Cl)Siemens Healthcare Diagnostics, Inc. · 2023
  4. Descendant · cites it directly · 0 of its ownK222116 Atellica® CI Analyzer, Atellica® IMThyroid Stimulating Hormone…Siemens Healthcare Diagnostics, Inc. · 2023 · Recalled
  5. Descendant · cites it directly · 0 of its ownK161954 Atellica Solution, Atellica A-LYTE Integrated Multisensor…Siemens Healthcare Diagnostics · 2017 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K222438A-LYTE® Integrated Multisensor (IMT Na K Cl)Siemens Healthcare Diagnostics, Inc. · 2023 · cites it directly0descendants
  • K222116Atellica® CI Analyzer, Atellica® IMThyroid Stimulating Hormone 3-Ultra (TSH3-UL)…Siemens Healthcare Diagnostics, Inc. · 2023 · cites it directly0descendants
  • K161954Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH…Siemens Healthcare Diagnostics · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K990346Bayer Advia 1650 Chemistry SystemBayer Corp.199914
0 (this device)K151767Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT…Siemens Healthcare Diagnostics, Inc.2015—
+1K222438A-LYTE® Integrated Multisensor (IMT Na K Cl)Siemens Healthcare Diagnostics, Inc.2023—
+1K222116Atellica® CI Analyzer, Atellica® IMThyroid Stimulating Hormone 3-Ultra…Siemens Healthcare Diagnostics, Inc.20231
+1K161954Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica…Siemens Healthcare Diagnostics20178

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202320
FIG. 1 — Lineage of K151767, 1999–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990346Bayer Advia 1650 Chemistry System
Summary page 6
…Product Predicate Device number Class Regulation Code Trinidad CH Siemens ADVIA® K990346 I 862.2160 JJE System 1800 Chemistry* ADVIA® Sodium TD-LYTE…
Bayer Corp.199914
*K132664Advia Chemistry Albumin BCP Reagent (Albp), Advia Chemistry Albumin BCP Calibrator
Summary page 6
…ADVIA® Chemistry Standard B + Salt ISE Urine Standard Bridge Trinidad CH ADVIA® Chemistry Albumin BCP Albumin BCP Assay K132664 II 862.1035 CJW Reagent (Alb_P) (ALBP) Trinidad CH ADVIA® Chemistry Albumin BCP Albumin B…
Siemens Healthcare Diagnostics, Inc.2013—

Also cleared against K990346: 2 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K222438A-LYTE® Integrated Multisensor (IMT Na K Cl)Siemens Healthcare Diagnostics, Inc.20230
K222116Atellica® CI Analyzer, Atellica® IMThyroid Stimulating Hormone 3-Ultra (TSH3-UL)…Siemens Healthcare Diagnostics, Inc.20230
K161954Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH…Siemens Healthcare Diagnostics20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 22, 2025Class IIZ-2027-2025Under Investigation by firmContamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Dec 15, 2023Class IIZ-1003-2024Under Investigation by firmPotential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
Nov 18, 2021Class IIZ-0449-2022Software design(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On…
May 13, 2021Class IIZ-2026-2021Component design/selectionValves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject…
May 4, 2021Class IIZ-1870-2021Device DesignPhotometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
Dec 2, 2020Class IIZ-0861-2021Under Investigation by firmMultiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
Feb 27, 2020Class IIZ-1797-2020Under Investigation by firmEditing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or…
Oct 11, 2019Class IIZ-0862-2020Under Investigation by firmOn the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the…
Sep 23, 2019Class IIZ-0581-2020Software designPotential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.
Jul 30, 2019Class IIZ-0148-2020Software designAtellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
Sep 26, 2018Class IIZ-0782-2019Software designMultiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime
Jun 29, 2018Class IIZ-0194-2019Software designAtellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K222438 (Siemens Healthcare Diagnostics, Inc., 2023) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K990346.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—

What is it indicated for?

The Trinidad CH System is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. The system’s chemical and immunochemical assay applications utilize photometric and ion selective electrode technology for clinical use. The TD-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination of sodium, potassium and chloride (Na, K, Cl) in human serum…

Read the full statement (297 words)

The Trinidad CH System is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. The system’s chemical and immunochemical assay applications utilize photometric and ion selective electrode technology for clinical use.

The TD-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination of sodium, potassium and chloride (Na, K, Cl) in human serum, plasma and urine using the Trinidad CH System. Measurements of sodium obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The TD-LYTE IMT Standard A is intended for the calibration of Na, K, and Cl on the Trinidad CH System.

The TD-LYTE IMT Standard B + Salt Bridge is intended for the calibration of Na, K, and Cl on the Trinidad CH System.

The Albumin BCP Reagent (Alb_P) is intended for in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on the Trinidad CH System. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.

The Albumin BCP calibrator is for in vitro diagnostic use in the calibration of the Trinidad CH Albumin BCP Assay (AOE_P) on the Trinidad CH System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151767: Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151767

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026