Recalled · Class II · 2024Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Healthcare Diagnostics · JGS Electrode, Ion Specific, Sodium
Cleared on Mar 9, 2017 as substantially equivalent to K151792 (Siemens Healthcare Diagnostics, Inc., 2015) and 2 other predicates after a 234-day review. Its family tree reaches back 5 generations to 1990. It has been recalled 8 times.
- Decided
- Mar 9, 2017
- Review
- 234 days
- Class
- Class II
- Product code
- JGS
- Recalls
- 8
- Predicates
- 3
Its longest line of descent runs back 5 generations to K902336 (1990). 17 ancestors in all, the oldest K833379 (1983). No later device cites it as a predicate yet.
- Root · 5 generations backK902336 Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp. · 1990
- 4 generations backK971418 Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur…Chiron Diagnostics Corp. · 1997 · Recalled
- 3 generations backK032525 Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003
- 2 generations backK041133 Modification to Bayer Advia Centaur Automated Chemiluminescence…Bayer Healthcare, LLC · 2004 · Recalled
- 1 generation backK151792 Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH…Siemens Healthcare Diagnostics, Inc. · 2015 · Recalled
- This deviceK161954 Atellica Solution, Atellica A-LYTE Integrated Multisensor…Siemens Healthcare Diagnostics · 2017
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K902336 | Ciba Corning Acs:180 Automated Chemiluminescence | Ciba Corning Diagnostics Corp. | 1990 | — |
| −4 | K971418 | Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur Analyzer | Chiron Diagnostics Corp. | 1997 | 4 |
| −3 | K032525 | Bayer Advia Centaur Automated Chemiluminescence Analyzer | Bayer Healthcare, LLC | 2003 | — |
| −2 | K041133 | Modification to Bayer Advia Centaur Automated Chemiluminescence Analyzer | Bayer Healthcare, LLC | 2004 | 8 |
| −1 | K151792 | Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH… | Siemens Healthcare Diagnostics, Inc. | 2015 | 22 |
| 0 (this device) | K161954 | Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica… | Siemens Healthcare Diagnostics | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1983 | 1 | 0 |
| 1989 | 1 | 0 |
| 1996 | 1 | 0 |
| 1999 | 2 | 0 |
| 2003 | 1 | 0 |
| 2004 | 1 | 0 |
| 2009 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 2 | 0 |
| 2016 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JGS device
- #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K151792 | Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH Calibrators…Summary page 5…xempt Chemistry For Clinical Use Chemistry (75) 1.1.6 Predicate Devices: Trinidad IM module with Direct Load cleared under k151792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH m… | 2015 |
| * | K151767 | Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A…Summary page 5…(75) 1.1.6 Predicate Devices: Trinidad IM module with Direct Load cleared under k151792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH module with Direct Load cleared under k160202 for Vancomycin Measurement === P… | 2015 |
| * | K160202 | Trinidad CH Vancomycin (Vanc), Trinidad CH Drug 3 Calibrator (DRUG3 CAL)Summary page 5…792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH module with Direct Load cleared under k160202 for Vancomycin Measurement 1.1.7 Device Description: 1.1.7.1 Name change and alignment with prior clearan… | 2016 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 21, 2024 | Class II | Z-0496-2025 | Under Investigation by firm | A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and… |
| Jan 12, 2020 | Class II | Z-1421-2020 | Under Investigation by firm | Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results. |
| Mar 7, 2019 | Class II | Z-1301-2019 | Under Investigation by firm | There were multiple issues identified in the system software which required an update. |
| Sep 26, 2018 | Class II | Z-0783-2019 | Software design | Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime |
| Jun 29, 2018 | Class II | Z-0189-2019 | Software design | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific… |
| Jun 29, 2018 | Class II | Z-0190-2019 | Software design | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific… |
| Jun 29, 2018 | Class II | Z-0191-2019 | Software design | Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific… |
| Feb 26, 2018 | Class II | Z-1587-2018 | Software design | When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 27 years and passes through 5 companies; a new generation was cleared every 6.1 years on median.
- 7 of the 17 ancestors have a recall of any class.
- 2 other devices also cite its predicate K151792.
How does its category compare?
All JGS devicesJGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · JGS
3in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · JGS
347in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JGS
4in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JGS
64in 2025
MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 7 | 33 |
| 2008 | 2 | — | 1 | 3 |
| 2009 | 2 | — | 0 | 57 |
| 2010 | 1 | — | 1 | 361 |
| 2011 | 3 | 191 | 1 | 912 |
| 2012 | 1 | — | 2 | 412 |
| 2013 | 2 | — | 5 | 510 |
| 2014 | 1 | — | 0 | 251 |
| 2015 | 1 | — | 1 | 158 |
| 2016 | 4 | 198 | 0 | 84 |
| 2017 | 1 | — | 0 | 292 |
| 2018 | 3 | 180 | 0 | 158 |
| 2019 | 3 | 58 | 5 | 105 |
| 2020 | 2 | — | 1 | 89 |
| 2021 | 0 | — | 1 | 91 |
| 2022 | 1 | — | 0 | 60 |
| 2023 | 3 | 347 | 4 | 98 |
| 2024 | 0 | — | 0 | 96 |
| 2025 | 0 | — | 0 | 64 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The Siemens Atellica™ Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use. The Atellica™ A-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination…
Read the full statement (346 words)
The Siemens Atellica™ Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.
The Atellica™ A-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination of sodium, potassium, and chloride (Na, K, Cl) in human serum, plasma, and urine using the Atellica™ CH system. Measurements of sodium obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
The Atellica™ CH Albumin BCP Reagent (Alb_P) is intended for in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on Atellica™ CH system. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.
The Atellica™ IM Thyroid Stimulating Hormone (TSH) assay is for in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in human serum, and plasma (EDTA and lithium heparin) using the Atellica™ IM system. Measurements of the thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
The Atellica™ CH Vancomycin (Vanc) assay is for in vitro diagnostic use in the quantitative measurement of vancomycin in human serum or plasma on the Atellica™ CH System. Vanc test results may be used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K161954 · printed Nov 6, 2026
Cite this page
510(k) Report. “K161954: Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161954
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026