510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc)

K161954510(k) TraditionalCleared Mar 9, 2017234-day reviewClass II21 CFR 862.1665Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics · JGS Electrode, Ion Specific, Sodium

Cleared on Mar 9, 2017 as substantially equivalent to K151792 (Siemens Healthcare Diagnostics, Inc., 2015) and 2 other predicates after a 234-day review. Its family tree reaches back 5 generations to 1990. It has been recalled 8 times.

Decided
Mar 9, 2017
Review
234 days
Class
Class II
Product code
JGS
Recalls
8
Predicates
3

Its longest line of descent runs back 5 generations to K902336 (1990). 17 ancestors in all, the oldest K833379 (1983). No later device cites it as a predicate yet.

  1. Root · 5 generations backK902336 Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp. · 1990
  2. 4 generations backK971418 Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur…Chiron Diagnostics Corp. · 1997 · Recalled
  3. 3 generations backK032525 Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003
  4. 2 generations backK041133 Modification to Bayer Advia Centaur Automated Chemiluminescence…Bayer Healthcare, LLC · 2004 · Recalled
  5. 1 generation backK151792 Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH…Siemens Healthcare Diagnostics, Inc. · 2015 · Recalled
  6. This deviceK161954 Atellica Solution, Atellica A-LYTE Integrated Multisensor…Siemens Healthcare Diagnostics · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K902336Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp.1990—
−4K971418Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur AnalyzerChiron Diagnostics Corp.19974
−3K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC2003—
−2K041133Modification to Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC20048
−1K151792Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH…Siemens Healthcare Diagnostics, Inc.201522
0 (this device)K161954Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica…Siemens Healthcare Diagnostics2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198910
199610
199920
200310
200410
200910
201310
201520
201610
FIG. 1 — Lineage of K161954, 1990–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K151792Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH Calibrators…
Summary page 5
…xempt Chemistry For Clinical Use Chemistry (75) 1.1.6 Predicate Devices: Trinidad IM module with Direct Load cleared under k151792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH m…
Siemens Healthcare Diagnostics, Inc.201522
*K151767Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A…
Summary page 5
…(75) 1.1.6 Predicate Devices: Trinidad IM module with Direct Load cleared under k151792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH module with Direct Load cleared under k160202 for Vancomycin Measurement === P…
Siemens Healthcare Diagnostics, Inc.201512
*K160202Trinidad CH Vancomycin (Vanc), Trinidad CH Drug 3 Calibrator (DRUG3 CAL)
Summary page 5
…792 for TSH Measurement Trinidad CH module with Direct Load cleared under k151767 for Na, K, Cl and Albumin Measurement Trinidad CH module with Direct Load cleared under k160202 for Vancomycin Measurement 1.1.7 Device Description: 1.1.7.1 Name change and alignment with prior clearan…
Siemens Healthcare Diagnostics, Inc.2016—

Also cleared against K151792: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 21, 2024Class IIZ-0496-2025Under Investigation by firmA manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and…
Jan 12, 2020Class IIZ-1421-2020Under Investigation by firmSome lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results.
Mar 7, 2019Class IIZ-1301-2019Under Investigation by firmThere were multiple issues identified in the system software which required an update.
Sep 26, 2018Class IIZ-0783-2019Software designMultiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime
Jun 29, 2018Class IIZ-0189-2019Software designAtellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific…
Jun 29, 2018Class IIZ-0190-2019Software designAtellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific…
Jun 29, 2018Class IIZ-0191-2019Software designAtellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific…
Feb 26, 2018Class IIZ-1587-2018Software designWhen more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 5 companies; a new generation was cleared every 6.1 years on median.
  • 7 of the 17 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K151792.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—

What is it indicated for?

The Siemens Atellica™ Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use. The Atellica™ A-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination…

Read the full statement (346 words)

The Siemens Atellica™ Solution is a multi-component system for in vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of various body fluids, using photometric, turbidimetric, chemiluminescent, and integrated ion selective electrode technology for clinical use.

The Atellica™ A-LYTE Integrated Multisensor (Na, K, Cl) is intended for in vitro diagnostic use in the quantitative determination of sodium, potassium, and chloride (Na, K, Cl) in human serum, plasma, and urine using the Atellica™ CH system. Measurements of sodium obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The Atellica™ CH Albumin BCP Reagent (Alb_P) is intended for in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on Atellica™ CH system. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.

The Atellica™ IM Thyroid Stimulating Hormone (TSH) assay is for in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in human serum, and plasma (EDTA and lithium heparin) using the Atellica™ IM system. Measurements of the thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

The Atellica™ CH Vancomycin (Vanc) assay is for in vitro diagnostic use in the quantitative measurement of vancomycin in human serum or plasma on the Atellica™ CH System. Vanc test results may be used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161954: Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica IM Thyroid Stimulating Hormone TSH, Atellica CH Vancomycin (Vanc).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161954

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.