510(k) Report

510(k) · Cleared

Cleared

Cardinal Health Sterilization Wrap

K151788510(k) TraditionalCleared Oct 30, 2015121-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardinal Health200, LLC · FRG Wrap, Sterilization

Cleared on Oct 30, 2015 as substantially equivalent to K132060 (Cardinal Health-Medical Products and Services, 2013) after a 121-day review. Its family tree reaches back 8 generations to 1988. 4 later devices descend from it.

Decided
Oct 30, 2015
Review
121 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K881471 (1988). 13 ancestors in all. 4 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 7 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 6 generations backK091685 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 5 generations backK112211 Dura Blue Sterilization WrapCardinal Health200, LLC · 2011
  5. 4 generations backK120658 Durablue Sterilization WrapCardinal Health200, LLC · 2012
  6. 3 generations backK122507 Durablue Sterilization Wrap, Model Numbers CH100, CH200, CH300…Cardinal Health200, LLC · 2012
  7. 2 generations backK123857 Durablue Sterilization WrapCardinal Health200, LLC · 2013
  8. 1 generation backK132060 Durablue Sterilization WrapCardinal Health-Medical Products and Services · 2013
  9. This deviceK151788 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2015
  10. Descendant · cites it directly · 3 of its ownK181174 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2018
  11. Descendant · 2 generations on · 1 of its ownK233777 PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc. · 2024
  12. Descendant · 3 generations on · 0 of its ownK242208 PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc. · 2024
  13. Descendant · 2 generations on · 0 of its ownK222151 Sterilization WrapKenpax International Limited · 2023

Most influential descendants

Showing 4 of 4 · See all
  • K181174Cardinal Health Sterilization WrapCardinal Health200, LLC · 2018 · cites it directly3descendants
  • K233777PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc. · 2024 · 2 generations on1descendant
  • K242208PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc. · 2024 · 3 generations on0descendants
  • K222151Sterilization WrapKenpax International Limited · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−7K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−6K091685Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC…Kimberly-Clark Corp.2009—
−5K112211Dura Blue Sterilization WrapCardinal Health200, LLC2011—
−4K120658Durablue Sterilization WrapCardinal Health200, LLC2012—
−3K122507Durablue Sterilization Wrap, Model Numbers CH100, CH200, CH300, CH400, CH500…Cardinal Health200, LLC2012—
−2K123857Durablue Sterilization WrapCardinal Health200, LLC2013—
−1K132060Durablue Sterilization WrapCardinal Health-Medical Products and Services2013—
0 (this device)K151788Cardinal Health Sterilization WrapCardinal Health200, LLC2015—
+1K181174Cardinal Health Sterilization WrapCardinal Health200, LLC2018—
+2K233777PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc.2024—
+3K242208PRIMED Sterilization Wrap (P100, P200, P300, P400, P500, P600)Primed Medical Products, Inc.2024—
+2K222151Sterilization WrapKenpax International Limited2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910
201110
201220
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202310
202420
FIG. 1 — Lineage of K151788, 1988–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132060Durablue Sterilization Wrap
Summary page 9
…850 Classification Name: Sterilization Wrap Common Name: Sterilization Wrap Product Code: FRG Predicate Device: K132060 - DuraBlue™ Sterilization Wrap for use with Pre- Vacuum Steam (4 min/270°F) & 100% Ethylene Oxide (…
Cardinal Health-Medical Products and Services2013—

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181174Cardinal Health Sterilization WrapCardinal Health200, LLC20183

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 3 companies; a new generation was cleared every 11 months on median.
  • 1 of the 13 ancestors has a recall of any class.
  • Of its descendants, K181174 (Cardinal Health200, LLC, 2018) has gone furthest, with 3 descendants of its own.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

Cardinal Health™ Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by using: • Gravity steam at 250°F/121°C for 30 minutes • Pre-vacuum steam at 270°F/132°C for 4 minutes • 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes • Advanced Sterilization Products (ASP) STERRAD® 100S System • Advanced Sterilization…

Read the full statement (375 words)

Cardinal Health™ Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by using:

• Gravity steam at 250°F/121°C for 30 minutes

• Pre-vacuum steam at 270°F/132°C for 4 minutes

• 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes

• Advanced Sterilization Products (ASP) STERRAD® 100S System

• Advanced Sterilization Products (ASP) STERRAD® NX System, Standard and Advanced Cycles

• Advanced Sterilization Products (ASP) STERRAD® 100NX, Standard, Flex, Express, and DUO cycles

• Lumen, Non Lumen, and Flexible Cycles in the STERIS V-PRO® 1, STERIS V-PRO® 1 Plus and STERIS V-PRO® maX Low Temperature Sterilization Systems

The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed devices until it is used.

For gravity steam sterilization, the wrap has been validated for dry times of 20 minutes for Models CH100 and CH200, and for 30 minutes for Models CH300, CH400, CH500 and CH600.

For pre-vacuum steam sterilization, the wrap has been validated for dry times of 20 minutes for Models CH100 and CH200, and for 30 minutes for Models CH300, CH400, CH500 and CH600. Models CH400, CH500 and CH600 have been validated for pre-vacuum steam sterilization of two lumens 3 mm in diameter or larger and 400 mm in length or less.

For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55 °C. Models CH400, CH500 and CH600 have been validated for EO sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less.

All models of Cardinal Health™ Sterilization Wrap have been validated for Advanced Sterilization Products (ASP) STERRAD® 100S sterilization of lumens 2.5 mm in diameter or larger and 250 mm in length or less.

All models of Cardinal Health™ Sterilization Wrap have been validated for use with the Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX cycles in Table 1.

All models of Cardinal Health™ Sterilization Wrap have been validated for use with the STERIS V-PRO® cycles in Table 2. Cardinal Health™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed STERIS V-PRO® sterilization cycles.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K151788: Cardinal Health Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151788

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026