510(k) Report

510(k) · Cleared

Cleared

Sterilization Wrap

K222151510(k) TraditionalCleared Mar 23, 2023246-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Kenpax International Limited · FRG Wrap, Sterilization

Cleared on Mar 23, 2023 as substantially equivalent to K181174 (Cardinal Health200, LLC, 2018) after a 246-day review. Its family tree reaches back 10 generations to 1988.

Decided
Mar 23, 2023
Review
246 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K881471 (1988). 17 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 10 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 6 generations between
  3. 3 generations backK132060 Durablue Sterilization WrapCardinal Health-Medical Products and Services · 2013
  4. 2 generations backK151788 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2015
  5. 1 generation backK181174 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2018
  6. This deviceK222151 Sterilization WrapKenpax International Limited · 2023
Show all 10 generations
  1. Root · 10 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 9 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 8 generations backK091685 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 7 generations backK112211 Dura Blue Sterilization WrapCardinal Health200, LLC · 2011
  5. 6 generations backK120658 Durablue Sterilization WrapCardinal Health200, LLC · 2012
  6. 5 generations backK122507 Durablue Sterilization Wrap, Model Numbers CH100, CH200, CH300…Cardinal Health200, LLC · 2012
  7. 4 generations backK123857 Durablue Sterilization WrapCardinal Health200, LLC · 2013
  8. 3 generations backK132060 Durablue Sterilization WrapCardinal Health-Medical Products and Services · 2013
  9. 2 generations backK151788 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2015
  10. 1 generation backK181174 Cardinal Health Sterilization WrapCardinal Health200, LLC · 2018
  11. This deviceK222151 Sterilization WrapKenpax International Limited · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−9K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−8K091685Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC…Kimberly-Clark Corp.2009—
−7K112211Dura Blue Sterilization WrapCardinal Health200, LLC2011—
−6K120658Durablue Sterilization WrapCardinal Health200, LLC2012—
−5K122507Durablue Sterilization Wrap, Model Numbers CH100, CH200, CH300, CH400, CH500…Cardinal Health200, LLC2012—
−4K123857Durablue Sterilization WrapCardinal Health200, LLC2013—
−3K132060Durablue Sterilization WrapCardinal Health-Medical Products and Services2013—
−2K151788Cardinal Health Sterilization WrapCardinal Health200, LLC2015—
−1K181174Cardinal Health Sterilization WrapCardinal Health200, LLC2018—
0 (this device)K222151Sterilization WrapKenpax International Limited2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910
201110
201220
201310
201510
201710
FIG. 1 — Lineage of K222151, 1988–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K181174Cardinal Health Sterilization Wrap
Summary page 8
…Page 4 of 15 510(k) Summary 6. Identification of Predicate Device(s) Primary Predicate Device 510(k) Number: K181174 Product Name: Cardinal Health TM Sterilization Wrap Manufacturer: Cardinal Health 200 LLC Classification: 2 Produ…
Cardinal Health200, LLC2018—

Also cleared against K181174: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 4 companies; a new generation was cleared every 1.9 years on median.
  • 1 of the 17 ancestors has a recall of any class.
  • 1 other device also cites its predicate K181174.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used up to 365 days post sterilization. Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using: a) 100% ethylene oxide (EO) with a concentration of 725 - 735 mg/l at 131 °F/55°C and 40%-80% relative humidity for 60 minutes. b) Exhausting the EO…

Read the full statement (126 words)

The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used up to 365 days post sterilization.

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:

a) 100% ethylene oxide (EO) with a concentration of 725 - 735 mg/l at 131 °F/55°C and 40%-80% relative humidity for 60 minutes.

b) Exhausting the EO gas vacuum depth: 157 mBar to 160 mBar

c) Aeration time : 12 hours

d) Aeration temperature: 55°C

e) Aeration pressure : 841 mBar to 864 mBar

Suggestions for the packaging content are given in table, A single layer of wrapping needs to be packaged in two sheets.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K222151: Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K222151

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026