510(k) Report

510(k) · Cleared

Cleared

Sterisheet Sterilization Wrap

K152291510(k) TraditionalCleared Dec 17, 2015126-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Andersen Sterilizers, Inc. · FRG Wrap, Sterilization

Cleared on Dec 17, 2015 as substantially equivalent to K931202 (Arjo Wiggins Medical, Inc., 1993) after a 126-day review. Its family tree reaches back 1 generation to 1993. 3 later devices descend from it.

Decided
Dec 17, 2015
Review
126 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K931202 (1993). 3 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2017: 1 descendant cleared2026: 1 descendant clearedK931202 · Sterilization Wrap-Sterisheet · Arjo Wiggins Medical, Inc. · 19931990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant K931202 · Sterilization Wrap-Sterisheet · Arjo Wiggins Medical, Inc. · 19930
  1. Root · 1 generation backK931202 Sterilization Wrap-SterisheetArjo Wiggins Medical, Inc. · 1993
  2. This deviceK152291 Sterisheet Sterilization WrapAndersen Sterilizers, Inc. · 2015
  3. Descendant · cites it directly · 1 of its ownK200335 Sterisheet Sterilization WrapAndersen Sterilizers, Inc. · 2020
  4. Descendant · 2 generations on · 0 of its ownK252823 Dukal Crepe Paper Sterilization WrapDukal, LLC · 2026
  5. Descendant · cites it directly · 0 of its ownK170437 Sterisheet Sterilization WrapAndersen Sterilizers, Inc. · 2017

Most influential descendants

Showing 3 of 3 · See all
  • K200335Sterisheet Sterilization WrapAndersen Sterilizers, Inc. · 2020 · cites it directly1descendant
  • K252823Dukal Crepe Paper Sterilization WrapDukal, LLC · 2026 · 2 generations on0descendants
  • K170437Sterisheet Sterilization WrapAndersen Sterilizers, Inc. · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K931202Sterilization Wrap-SterisheetArjo Wiggins Medical, Inc.1993—
0 (this device)K152291Sterisheet Sterilization WrapAndersen Sterilizers, Inc.2015—
+1K200335Sterisheet Sterilization WrapAndersen Sterilizers, Inc.2020—
+2K252823Dukal Crepe Paper Sterilization WrapDukal, LLC2026—
+1K170437Sterisheet Sterilization WrapAndersen Sterilizers, Inc.2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202010
202610
FIG. 1 — Lineage of K152291, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K931202Sterilization Wrap-Sterisheet
Summary page 5
…uct Code FRG 5.5 Predicate Device Device Name Sterisheet® Sterilization Wrap 510(k) number K931202 Manufacturer Arjowiggins Medical Inc. This 510(k) submission modifies the indications for use of…
Arjo Wiggins Medical, Inc.1993—

Also cleared against K931202: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200335Sterisheet Sterilization WrapAndersen Sterilizers, Inc.20201
K170437Sterisheet Sterilization WrapAndersen Sterilizers, Inc.20170

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K200335 (Andersen Sterilizers, Inc., 2020) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K931202.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The Sterisheet Sterilization Wraps are single use non woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized below in Table 1. Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer EO Amount…

Read the full statement (298 words)

The Sterisheet Sterilization Wraps are single use non woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Critical process parameters for the cycle are summarized below in Table 1.

Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer

EO Amount Temperature Relative Humidity EO Exposure Time Total Cycle Time

17.6 g ± 5% 50°C ± 3°C 35-90% 3 hours 3.5 hours

The product code for the Sterisheet Sterilization Wraps is listed in Table 2.

Table 2. Product code for the Sterisheet Sterilization Wraps

Product Designation Sterisheet S88 Blue

Product Code 0129

Maximum loads of specific materials and devices that have been validated are listed in Table 3. All validated maximum loads were processed without additional devices in the sterilizer. Sterility was maintained for at least 3 month after processing in an EOGas 4 sterilizer.

Table 3. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer

Device Type Maximum Load Device Examples Required Aeration

Metal 24 lbs (11 kg) Surgical instruments, Metal instruments do not absorb EO;

delicate sharps, including those Follow pouch or wrap manufacturer’s

with hinges and mated surfaces instructions (Example: Tyvek® pouches and

Sterisheet® wrap require ≥ 6 hours at 50°C).

Plastic 7.0 lbs (3.2 kg) Reusable power cords, trocars 24 hours at 50°C;

Fabric 6.1 lbs (2.8 kg) Reusable cloth gowns, towels Follow manufacturer’s instructions

Single-lumen One (1) ≥ 2.0mm ID Gastrovideoscopes, gastrointestinal 12 hours at 50°C;

Endoscope(s) ≤ 1100mm length; videoscopes Follow manufacturer’s instructions

No additional devices

Four (4) ≥ 1.2 mm ID Bronchoscopes, bronchovideoscopes,

≤ 700 mm length; cystoscopes, ureteroscopes,

No additional devices choledocoscopes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152291: Sterisheet Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K152291

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026