510(k) · Cleared
Recalled · Class I · 2017StrataMR Valves and Shunts
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medtronic Neurosurgery · JXG Shunt, Central Nervous System And Components
Cleared on Apr 7, 2016 as substantially equivalent to K060681 (Medtronic Neurosurgery, 2006) after a 199-day review. Its family tree reaches back 7 generations to 1995. 1 of the 16 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled once, including a Class I recall.
- Decided
- Apr 7, 2016
- Review
- 199 days
- Class
- Class II
- Product code
- JXG
- Recalls
- 1 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 7 generations to K944222 (1995). 13 ancestors in all, the oldest K861377 (1986). 2 later devices descend from it, 2 citing it directly. 1 of the 16 devices in this family, counting this one, has a Class I recall.
- Root · 7 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
- 6 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
- 5 generations backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
- 4 generations backK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
- 3 generations backK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
- 2 generations backK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
- 1 generation backK060681 Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery · 2006 · Recalled
- This deviceK152700 StrataMR Valves and ShuntsMedtronic Neurosurgery · 2016
- Descendant · cites it directly · 0 of its ownK212641 StrataMR II Valves and ShuntsMedtronic Neurosurgery · 2021
- Descendant · cites it directly · 0 of its ownK181622 StrataMR guider toolMedtronic Neurosurgery · 2018
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K944222 | Codman-Medos Valve Shunt System | Johnson & Johnson Professionals, Inc. | 1995 | — |
| −6 | K973774 | Codman Hakim Micro Precision Valve | Johnson & Johnson Professionals, Inc. | 1997 | — |
| −5 | K980778 | Hakim Micro Programmable Valve System | Johnson & Johnson Professionals, Inc. | 1998 | — |
| −4 | K012052 | Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855 | Medtronic PS Medical | 2002 | — |
| −3 | K033850 | Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without… | Medtronic Neurosurgery | 2004 | 2 |
| −2 | K042465 | PS Medical Strata II Valve | Medtronic Neurosurgery | 2004 | 4 |
| −1 | K060681 | Medtronic PS Medical Strata Type Value | Medtronic Neurosurgery | 2006 | 2 |
| 0 (this device) | K152700 | StrataMR Valves and Shunts | Medtronic Neurosurgery | 2016 | — |
| +1 | K212641 | StrataMR II Valves and Shunts | Medtronic Neurosurgery | 2021 | — |
| +1 | K181622 | StrataMR guider tool | Medtronic Neurosurgery | 2018 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1986 | 1 | 0 |
| 1990 | 1 | 0 |
| 1995 | 1 | 0 |
| 1998 | 1 | 0 |
| 2000 | 1 | 0 |
| 2004 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2018 | 1 | 0 |
| 2021 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JXG device
- #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K060681 | Medtronic PS Medical Strata Type ValueSummary page 4…and Components (21 CFR §882.5550, Product Code JXG) Predicate Device: Medtronic PS Medical Strata Type Valve (K060681) Device Description: The Medtronic StrataMR™ valves are implantable adjustable valves for the management of hydrocephalus. The v… | 2006 |
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 22, 2017 | Class I | Z-1702-2017 | Device Design | Incorrect valve rotor orientation due to misalignment and/or off-center positioning of the StrataMR Locator Tool: When the StrataMR valve is set in the incorrect orientation, higher valve opening pressures result leading to a… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 1.9 years on median.
- 5 of the 13 ancestors have a recall of any class.
- Of its descendants, K212641 (Medtronic Neurosurgery, 2021) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K060681.
How does its category compare?
All JXG devicesJXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · JXG
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JXG
255in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JXG
3in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JXG
1,198in 2025
MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 184 | 0 | 649 |
| 2008 | 6 | 60 | 1 | 630 |
| 2009 | 5 | 106 | 30 | 629 |
| 2010 | 3 | 30 | 0 | 812 |
| 2011 | 7 | 136 | 5 | 787 |
| 2012 | 3 | 78 | 1 | 977 |
| 2013 | 3 | 108 | 14 | 1,004 |
| 2014 | 0 | — | 6 | 1,251 |
| 2015 | 6 | 276 | 0 | 984 |
| 2016 | 6 | 143 | 0 | 1,091 |
| 2017 | 7 | 90 | 1 | 1,335 |
| 2018 | 6 | 102 | 0 | 1,484 |
| 2019 | 4 | 101 | 0 | 1,295 |
| 2020 | 2 | — | 8 | 1,206 |
| 2021 | 1 | — | 0 | 1,467 |
| 2022 | 4 | 101 | 0 | 1,329 |
| 2023 | 5 | 134 | 2 | 1,221 |
| 2024 | 3 | 90 | 3 | 1,214 |
| 2025 | 8 | 255 | 0 | 1,198 |
| 2026 | 2 | — | 1 | — |
What is it indicated for?
The Medtronic StrataMR™ valves and shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasively adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K152700 · printed Nov 6, 2026
Cite this page
510(k) Report. “K152700: StrataMR Valves and Shunts.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152700
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026