510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Codman Certas Programmable Valve System, Codman Certas Therapy Management System

K112156510(k) TraditionalCleared Oct 17, 201182-day reviewClass II21 CFR 882.5550Recalls 16

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medos International SARL · JXG Shunt, Central Nervous System And Components

Cleared on Oct 17, 2011 as substantially equivalent to K974739 (Johnson & Johnson Professionals, Inc., 1998) and 3 other predicates after an 82-day review. Its family tree reaches back 7 generations to 1995. 3 later devices descend from it. It has been recalled 16 times.

Decided
Oct 17, 2011
Review
82 days
Class
Class II
Product code
JXG
Recalls
16
Predicates
4

Its longest line of descent runs back 7 generations to K944222 (1995). 14 ancestors in all, the oldest K861377 (1986). 3 later devices descend from it, 2 citing it directly.

  1. Root · 7 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 6 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
  3. 5 generations backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
  4. 4 generations backK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
  5. 3 generations backK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
  6. 2 generations backK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
  7. 1 generation backK060681 Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery · 2006 · Recalled
  8. This deviceK112156 Codman Certas Programmable Valve System, Codman Certas Therapy…Medos International SARL · 2011
  9. Descendant · cites it directly · 1 of its ownK143111 Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL · 2015
  10. Descendant · 2 generations on · 0 of its ownK181902 Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp. · 2018
  11. Descendant · cites it directly · 0 of its ownK113526 Codman Certas Programmable Valve SystemMedos International SARL · 2012 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K143111Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL · 2015 · cites it directly1descendant
  • K181902Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp. · 2018 · 2 generations on0descendants
  • K113526Codman Certas Programmable Valve SystemMedos International SARL · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K944222Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc.1995—
−6K973774Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc.1997—
−5K980778Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc.1998—
−4K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical2002—
−3K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without…Medtronic Neurosurgery20042
−2K042465PS Medical Strata II ValveMedtronic Neurosurgery20044
−1K060681Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery20062
0 (this device)K112156Codman Certas Programmable Valve System, Codman Certas Therapy Management SystemMedos International SARL2011—
+1K143111Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL2015—
+2K181902Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp.2018—
+1K113526Codman Certas Programmable Valve SystemMedos International SARL201216

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199010
199510
199810
200010
200410
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201510
201810
FIG. 1 — Lineage of K112156, 1995–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K974739Hakim Programmable Valve System
Summary page 1
…ion: Class 11(JXG) Central Nervous System Fluid Shunt and Components Legally Marketed Predicate Devices: K<974739 - Codman Hakim Programmable Valve K091 346- Codman Hakim Shunt System…
Johnson & Johnson Professionals, Inc.1998—
*K060681Medtronic PS Medical Strata Type Value
Summary page 1
…K<992173 - Codman SIPHONGUARD CSF Control Device K003322 - Codman Bactiseal Catheters K<060681 - Medtronic PS Medical® Strata® Type® Valve K040943 - Medtronic PS Strata® Valves and Handtoo…
Medtronic Neurosurgery20062
*K040943Medtronic PS Medical Strata Valve
Summary page 1
…K003322 - Codman Bactiseal Catheters K<060681 - Medtronic PS Medical® Strata® Type® Valve K040943 - Medtronic PS Strata® Valves and Handtools K<090342 - Sophysa Polaris Device Description:…
Medtronic Neurosurgery2004—
*K090342Polaris Pressure Adjustable Valve, Models SPV, Spva, SPVB, SPV-140, SPV-300, SPV-400
Summary page 1
…K<060681 - Medtronic PS Medical® Strata® Type® Valve K040943 - Medtronic PS Strata® Valves and Handtools K<090342 - Sophysa Polaris Device Description: The Certas Programmable Valve System is designed for shunting…
Sophysa SA2009—

Also cleared against K974739: 9 other devices →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143111Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL20151
K113526Codman Certas Programmable Valve SystemMedos International SARL20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 3, 2014Class IIZ-2184-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2189-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2191-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2190-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2188-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2185-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2187-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Jul 3, 2014Class IIZ-2186-2014Device DesignTo clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
Dec 13, 2013Class IIZ-1468-2014Process controlEnsuring all customers have the tools to assist in verifying the setting of the valve on x-ray images and the location and orientation of implanted CODMAN CERTAS Programmable Valves.
May 16, 2013Class IIZ-0858-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0863-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0860-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0861-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0862-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0859-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly
May 16, 2013Class IIZ-0864-2014Device DesignCodman Certas Programmable Valves used for hydrocephalus may not operate properly

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 1.9 years on median.
  • 5 of the 14 ancestors have a recall of any class.
  • Of its descendants, K143111 (Medos International SARL, 2015) has gone furthest, with 1 descendant of its own.
  • 9 other devices also cite its predicate K974739.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The CODMAN Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

The CODMAN Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112156: Codman Certas Programmable Valve System, Codman Certas Therapy Management System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112156

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.