510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators

K153540510(k) TraditionalCleared Aug 4, 2016238-day reviewClass II21 CFR 880.6850Recalls 43

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Safe Secure Packaging Co., Ltd. · FRG Wrap, Sterilization

Cleared on Aug 4, 2016 as substantially equivalent to K112591 (Safe Secure Packaging Co., Ltd., 2012) after a 238-day review. Its family tree reaches back 5 generations to 1999. 4 later devices descend from it. It has been recalled 43 times.

Decided
Aug 4, 2016
Review
238 days
Class
Class II
Product code
FRG
Recalls
43
Predicates
1

Its longest line of descent runs back 5 generations to K990567 (1999). 4 later devices descend from it, 3 citing it directly.

  1. Root · 5 generations backK990567 Global Healthcare Self-Sealing Sterilization PouchGlobal Healthcare, Inc. · 1999
  2. 4 generations backK993764 Medipack See-Through Seal Sterilization PouchMedipack Medical Packaging Mfg. Co. · 1999
  3. 3 generations backK051242 Winner Self Seal Sterilization PouchWinner Medical (Usa), Inc. · 2005
  4. 2 generations backK062704 Sterilization Pouch with IndicatorsWinner Industries (Shenzhen) Co., Ltd. · 2007
  5. 1 generation backK112591 Safe Secure Sterilization Pounch with Steam and Ethylene Oxide…Safe Secure Packaging Co., Ltd. · 2012
  6. This deviceK153540 Safe Secure Sterilization Pouch with Steam and Ethylene Oxide…Safe Secure Packaging Co., Ltd. · 2016
  7. Descendant · cites it directly · 1 of its ownK180139 Safe Secure Sterilization Pouches and RollsSafe Secure Packing Co., Ltd. · 2019
  8. Descendant · cites it directly · 0 of its ownK250306 Chex-All Sterilization Pouches and TubesPropper Manufacturing Co., Inc. · 2025
  9. Descendant · cites it directly · 0 of its ownK243431 Medicom Sterilization Pouch and ReelMedicom Asia-Pacific Holdings Ltd. Taiwan Branch. · 2025
  10. Descendant · 2 generations on · 0 of its ownK242839 Safe Secure Sterilization Pouches and RollsShenzhen Safesecure Medical Infection Control Tech Co., Ltd. · 2024

Most influential descendants

Showing 4 of 4 · See all
  • K180139Safe Secure Sterilization Pouches and RollsSafe Secure Packing Co., Ltd. · 2019 · cites it directly1descendant
  • K250306Chex-All Sterilization Pouches and TubesPropper Manufacturing Co., Inc. · 2025 · cites it directly0descendants
  • K243431Medicom Sterilization Pouch and ReelMedicom Asia-Pacific Holdings Ltd. Taiwan Branch. · 2025 · cites it directly0descendants
  • K242839Safe Secure Sterilization Pouches and RollsShenzhen Safesecure Medical Infection Control Tech Co., Ltd. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K990567Global Healthcare Self-Sealing Sterilization PouchGlobal Healthcare, Inc.1999—
−4K993764Medipack See-Through Seal Sterilization PouchMedipack Medical Packaging Mfg. Co.1999—
−3K051242Winner Self Seal Sterilization PouchWinner Medical (Usa), Inc.2005—
−2K062704Sterilization Pouch with IndicatorsWinner Industries (Shenzhen) Co., Ltd.2007—
−1K112591Safe Secure Sterilization Pounch with Steam and Ethylene Oxide Process…Safe Secure Packaging Co., Ltd.2012—
0 (this device)K153540Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process IndicatorsSafe Secure Packaging Co., Ltd.2016—
+1K180139Safe Secure Sterilization Pouches and RollsSafe Secure Packing Co., Ltd.2019—
+1K250306Chex-All Sterilization Pouches and TubesPropper Manufacturing Co., Inc.2025—
+1K243431Medicom Sterilization Pouch and ReelMedicom Asia-Pacific Holdings Ltd. Taiwan Branch.2025—
+2K242839Safe Secure Sterilization Pouches and RollsShenzhen Safesecure Medical Infection Control Tech Co., Ltd.2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202410
202520
FIG. 1 — Lineage of K153540, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112591Safe Secure Sterilization Pounch with Steam and Ethylene Oxide Process Indicators
Summary page 6
…: II Classification Panel: General Hospital Model Numbers: ABHSP, ABSSP, ABHST, ABSST 5.5 Predicate Devices K112591 - Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators…
Safe Secure Packaging Co., Ltd.2012—

Also cleared against K112591: 1 other device →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180139Safe Secure Sterilization Pouches and RollsSafe Secure Packing Co., Ltd.20191
K250306Chex-All Sterilization Pouches and TubesPropper Manufacturing Co., Inc.20250
K243431Medicom Sterilization Pouch and ReelMedicom Asia-Pacific Holdings Ltd. Taiwan Branch.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 2, 2020Class IIZ-0787-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate (bleed) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained…
Dec 2, 2020Class IIZ-0796-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0798-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0799-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0786-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0791-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0792-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0793-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0795-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0797-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0790-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0794-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0788-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Dec 2, 2020Class IIZ-0789-2021Process controlPurple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s)…
Jun 15, 2017Class IIZ-0744-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0745-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0748-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0749-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0753-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0761-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0764-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0770-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0751-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0752-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0758-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0759-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0760-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0762-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0765-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0766-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0771-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0747-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0750-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0757-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0763-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0768-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0743-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0746-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0754-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0755-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0756-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0767-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…
Jun 15, 2017Class IIZ-0769-2018Component change controlOn June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 4.5 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K180139 (Safe Secure Packing Co., Ltd., 2019) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K112591.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using: • Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time • Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time • Pre-vacuum steam at 134°C (273°F) for 3 minutes; 20 minutes dry time • Pre-vacuum steam at 135°C (275°F)…

Read the full statement (279 words)

Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:

• Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time

• Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time

• Pre-vacuum steam at 134°C (273°F) for 3 minutes; 20 minutes dry time

• Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time

• Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60oC (140oF).

The device is not intended and has not been validated for sterilization of devices that contain lumens.

The external chemical indicators on the pouches/rolls are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed devices. The chemical indicators change from green to purple after exposure to steam and from yellow to brown after exposure to ethylene oxide.

The Tyvek version of the pouch is for EO sterilization only.

If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 5 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility (SAL=10 -6). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 6 months after steam sterilization and 24 months after EO sterilization.

The maximum validated pouch load is 2.64 pounds (1.2kg).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K153540: Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K153540

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026