510(k) · Cleared
ClearedChex-All Sterilization Pouches and Tubes
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Propper Manufacturing Co., Inc. · FRG Wrap, Sterilization
Cleared on Oct 31, 2025 as substantially equivalent to K153540 (Safe Secure Packaging Co., Ltd., 2016) after a 270-day review. Its family tree reaches back 6 generations to 1999.
- Decided
- Oct 31, 2025
- Review
- 270 days
- Class
- Class II
- Product code
- FRG
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K990567 (1999). No later device cites it as a predicate yet.
- Root · 6 generations backK990567 Global Healthcare Self-Sealing Sterilization PouchGlobal Healthcare, Inc. · 1999
- 5 generations backK993764 Medipack See-Through Seal Sterilization PouchMedipack Medical Packaging Mfg. Co. · 1999
- 4 generations backK051242 Winner Self Seal Sterilization PouchWinner Medical (Usa), Inc. · 2005
- 3 generations backK062704 Sterilization Pouch with IndicatorsWinner Industries (Shenzhen) Co., Ltd. · 2007
- 2 generations backK112591 Safe Secure Sterilization Pounch with Steam and Ethylene Oxide…Safe Secure Packaging Co., Ltd. · 2012
- 1 generation backK153540 Safe Secure Sterilization Pouch with Steam and Ethylene Oxide…Safe Secure Packaging Co., Ltd. · 2016 · Recalled
- This deviceK250306 Chex-All Sterilization Pouches and TubesPropper Manufacturing Co., Inc. · 2025
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K990567 | Global Healthcare Self-Sealing Sterilization Pouch | Global Healthcare, Inc. | 1999 | — |
| −5 | K993764 | Medipack See-Through Seal Sterilization Pouch | Medipack Medical Packaging Mfg. Co. | 1999 | — |
| −4 | K051242 | Winner Self Seal Sterilization Pouch | Winner Medical (Usa), Inc. | 2005 | — |
| −3 | K062704 | Sterilization Pouch with Indicators | Winner Industries (Shenzhen) Co., Ltd. | 2007 | — |
| −2 | K112591 | Safe Secure Sterilization Pounch with Steam and Ethylene Oxide Process… | Safe Secure Packaging Co., Ltd. | 2012 | — |
| −1 | K153540 | Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators | Safe Secure Packaging Co., Ltd. | 2016 | 43 |
| 0 (this device) | K250306 | Chex-All Sterilization Pouches and Tubes | Propper Manufacturing Co., Inc. | 2025 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRG device
- #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K153540 | Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process IndicatorsSummary page 7…p Common Name: Sterilization pouch Product Code: FRG, JOJ Classification: Class II (21 C.F.R. 880.6850, 21 C.F.R. 880.2800) Predicate Device: K153540 Description of the Device The Chex-All® Sterilization pouches and tubes are made from a medical grade porous paper thermally se… | 2016 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 26 years and passes through 6 companies; a new generation was cleared every 4.7 years on median.
- 1 of the 6 ancestors has a recall of any class.
- 2 other devices also cite its predicate K153540.
How does its category compare?
All FRG devicesFRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · FRG
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRG
151in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRG
2in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FRG
18in 2025
MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 182 | 0 | 1 |
| 2008 | 2 | — | 0 | 2 |
| 2009 | 4 | 171 | 0 | 2 |
| 2010 | 4 | 213 | 0 | 0 |
| 2011 | 6 | 110 | 0 | 12 |
| 2012 | 10 | 133 | 0 | 4 |
| 2013 | 6 | 146 | 2 | 5 |
| 2014 | 8 | 177 | 0 | 13 |
| 2015 | 12 | 196 | 0 | 9 |
| 2016 | 9 | 131 | 0 | 20 |
| 2017 | 9 | 260 | 29 | 7 |
| 2018 | 5 | 150 | 0 | 22 |
| 2019 | 9 | 258 | 0 | 10 |
| 2020 | 5 | 122 | 14 | 21 |
| 2021 | 4 | 170 | 0 | 9 |
| 2022 | 9 | 217 | 2 | 4 |
| 2023 | 9 | 230 | 0 | 13 |
| 2024 | 11 | 182 | 0 | 21 |
| 2025 | 9 | 151 | 0 | 18 |
| 2026 | 4 | 247 | 0 | — |
What is it indicated for?
Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using: Hospital size sterilizer cycles • Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time • Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time • Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time Tabletop…
Read the full statement (399 words)
Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
Hospital size sterilizer cycles
• Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time
• Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
• Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
Tabletop sterilizer cycles:
• Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time
Ethylene Oxide sterilization cycles:
• Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following:
Metal lumen: 7 inch long, 3.2mm internal diameter
Plastic lumen: 5 inch long, 2.0 mm internal diameter.
Suitable Lumen cycles:
Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time).
Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time).
Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time).
The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide.
If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL≥10-6). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization.
The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13”x18”.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K250306 · printed Nov 6, 2026
Cite this page
510(k) Report. “K250306: Chex-All Sterilization Pouches and Tubes.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K250306
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026