510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

ABL90 Flex Plus

K160153510(k) SpecialCleared Nov 4, 2016287-day reviewClass II21 CFR 862.1120Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical Aps · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Nov 4, 2016 as substantially equivalent to K132691 (Radiometer Medical Aps, 2014) and 1 other predicate after a 287-day review. Its family tree reaches back 10 generations to 1986. 2 later devices descend from it. It has been recalled 5 times.

Decided
Nov 4, 2016
Review
287 days
Class
Class II
Product code
CHL
Recalls
5
Predicates
2

Its longest line of descent runs back 10 generations to K861087 (1986). 23 ancestors in all. 2 later devices descend from it, 2 citing it directly.

  1. Root · 10 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 6 generations between
  3. 3 generations backK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  4. 2 generations backK122729 ABL90 FlexRadiometer Medical Aps · 2012
  5. 1 generation backK131988 ABL90 Flex AnalyzerRadiometer Medical Aps · 2013
  6. This deviceK160153 ABL90 Flex PlusRadiometer Medical Aps · 2016
  7. Descendant · cites it directly · 0 of its ownK252207 ABL90 FLEX PLUS System, safeCLINITUBESRadiometer Medicals Aps · 2026
  8. Descendant · cites it directly · 0 of its ownK240998 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2024
Show all 10 generations
  1. Root · 10 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 9 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 8 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 7 generations backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. 6 generations backK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. 5 generations backK050869 Modification to ABL800 FlexRadiometer Medical Aps · 2005 · Recalled
  7. 4 generations backK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled
  8. 3 generations backK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  9. 2 generations backK122729 ABL90 FlexRadiometer Medical Aps · 2012
  10. 1 generation backK131988 ABL90 Flex AnalyzerRadiometer Medical Aps · 2013
  11. This deviceK160153 ABL90 Flex PlusRadiometer Medical Aps · 2016

Most influential descendants

Showing 2 of 2 · See all
  • K252207ABL90 FLEX PLUS System, safeCLINITUBESRadiometer Medicals Aps · 2026 · cites it directly0descendants
  • K240998ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−9K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−8K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−7K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
−6K041874ABL800 FlexRadiometer Medical Aps200422
−5K050869Modification to ABL800 FlexRadiometer Medical Aps20051
−4K092686ABL90 Flex SeriesRadiometer Medical Aps201010
−3K120197ABL90 FlexRadiometer Medical Aps20123
−2K122729ABL90 FlexRadiometer Medical Aps2012—
−1K131988ABL90 Flex AnalyzerRadiometer Medical Aps2013—
0 (this device)K160153ABL90 Flex PlusRadiometer Medical Aps2016—
+1K252207ABL90 FLEX PLUS System, safeCLINITUBESRadiometer Medicals Aps2026—
+1K240998ABL90 FLEX PLUS SystemRadiometer Medicals Aps2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
199210
199510
199630
199710
199820
199910
200010
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202610
FIG. 1 — Lineage of K160153, 1986–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.2 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132691ABL90 Flex
Summary page 9
…amount of fetal hemoglobin. FHbF is seldom used clinically. Page 4 of 12 4. Predicate device: ABL90 FLEX (K132691) Substantial Equivalence The ABL90 FLEX PLUS analyzer is substantially equivalent in Intended Use, fundamental scientific technol…
Radiometer Medical Aps2014—
*K131988ABL90 Flex Analyzer
Summary page 9
…hnology, features, and characteristics to the predicate devices: Predicate Device: 510(k) Number, Name, Device Manufacturer: K132691, ABL90 FLEX, Radiometer Medical ApS K131988, ABL90 FLEX, Radiometer Medical ApS Similarities Issue SE Devic…
Radiometer Medical Aps2013—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K252207ABL90 FLEX PLUS System, safeCLINITUBESRadiometer Medicals Aps20260
K240998ABL90 FLEX PLUS SystemRadiometer Medicals Aps20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 21, 2025Class IIZ-1659-2025Component design/selectionABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature…
Apr 14, 2021Class IIZ-1858-2021Device DesignDue to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Apr 14, 2021Class IIZ-1857-2021Device DesignDue to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Feb 3, 2020Class IIZ-1761-2020Software designThe firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement.…
Nov 21, 2019Class IIZ-2138-2020Nonconforming Material/ComponentThe analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 30 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 13 of the 23 ancestors have a recall of any class.
  • Of its descendants, K252207 (Radiometer Medicals Aps, 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood. The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care…

Read the full statement (385 words)

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH,

blood gases, electrolytes, glucose, lactate and oximetry in

heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood.

The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.

It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order.

Bilirubin measurements on the ABL90 FLEX PLUS analyzer are intended to aid in assessing the risk of kernicterus in neonates.

pH, pO2 and pCO2: pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-

base disturbances.

Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa+): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration,inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCl–): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Lactate (cLac): The lactate measurements measure the concentration of lactate in plasma. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.)

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus reduced hemoglobin.

FO2Hb: oxyhemoglobin as a fraction of total hemoglobin.

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoglobin as a fraction of total hemoglobin.

Fraction of Fetal Hemoglobin (FHbF): FHbF indicates the amount of fetal hemoglobin. FHbF is seldom used clinically.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K160153: ABL90 Flex Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160153

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026