510(k) Report

510(k) · Cleared

Recalled · 2006

Modification to ABL800 Flex

K050869510(k) SpecialCleared Aug 11, 2005127-day reviewClass II21 CFR 862.1110Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical Aps · CIG Diazo Colorimetry, Bilirubin

Cleared on Aug 11, 2005 as substantially equivalent to K041874 (Radiometer Medical Aps, 2004) after a 127-day review. Its family tree reaches back 5 generations to 1986. 18 later devices descend from it. It has been recalled once.

Decided
Aug 11, 2005
Review
127 days
Class
Class II
Product code
CIG
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K861087 (1986). 16 ancestors in all. 18 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1998: 2 other ancestors1997: 1 other ancestor1996: 3 other ancestors1995: 1 other ancestor1992: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor2005: 1 descendant cleared2014: 1 descendant cleared2017: 1 descendant cleared2010: 1 descendant cleared2008: 1 descendant cleared2012: 3 descendants cleared2013: 3 descendants cleared2025: 1 descendant cleared2021: 1 descendant cleared2011: 1 descendant cleared2016: 1 descendant cleared2015: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant clearedK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986K961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 3 other ancestors5 generations back: 5 other ancestors 1 generation forward: 4 descendants2 generations forward: 8 descendants3 generations forward: 5 descendants4 generations forward: 1 descendant K041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986−50
  1. Root · 5 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 4 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 3 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 2 generations backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. 1 generation backK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. This deviceK050869 Modification to ABL800 FlexRadiometer Medical Aps · 2005
  7. Descendant · cites it directly · 12 of its ownK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled
  8. Descendant · 2 generations on · 6 of its ownK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  9. Descendant · 3 generations on · 5 of its ownK122729 ABL90 FlexRadiometer Medical Aps · 2012
  10. Descendant · cites it directly · 3 of its ownK132691 ABL90 FlexRadiometer Medical Aps · 2014
  11. Descendant · 4 generations on · 3 of its ownK131988 ABL90 Flex AnalyzerRadiometer Medical Aps · 2013

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2005: 12008: 12010: 12011: 12012: 32013: 32014: 12015: 12016: 12017: 12021: 12024: 12025: 12026: 1
2005high 3 (2012) · low 0 (2006)2026

Most influential descendants

Showing 5 of 18 · See all
  • K092686ABL90 Flex SeriesRadiometer Medical Aps · 2010 · cites it directly12descendants
  • K120197ABL90 FlexRadiometer Medical Aps · 2012 · 2 generations on6descendants
  • K122729ABL90 FlexRadiometer Medical Aps · 2012 · 3 generations on5descendants
  • K132691ABL90 FlexRadiometer Medical Aps · 2014 · cites it directly3descendants
  • K131988ABL90 Flex AnalyzerRadiometer Medical Aps · 2013 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−4K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−3K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−2K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
−1K041874ABL800 FlexRadiometer Medical Aps200422
0 (this device)K050869Modification to ABL800 FlexRadiometer Medical Aps2005—
+1K092686ABL90 Flex SeriesRadiometer Medical Aps201010
+2K120197ABL90 FlexRadiometer Medical Aps20123
+3K122729ABL90 FlexRadiometer Medical Aps2012—
+1K132691ABL90 FlexRadiometer Medical Aps2014—
+4K131988ABL90 Flex AnalyzerRadiometer Medical Aps2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
199210
199510
199630
199710
199820
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200810
201010
201110
201230
201330
201410
201510
201610
201710
202110
202410
202510
202610
FIG. 1 — Lineage of K050869, 1986–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CIG device

  1. #1 K251998 Atellica CH Diazo Total Bilirubin (D_TBil)score 11.8 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041874ABL800 Flex
Summary page 1
…Kos50 q 510(k) Summary As required by 807.92 For RADIANCE v2.5 modification to ABL800 FLEX (K041874) Prepared on March 30, 2005 Submitted by: Radiometer Medical ApS Akandevej 21…
Radiometer Medical Aps200422

Also cleared against K041874: 5 other devices →

What descends from it? (18)

See all 18

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092686ABL90 Flex SeriesRadiometer Medical Aps201012
K132691ABL90 FlexRadiometer Medical Aps20143
K051804ABL80 Flex AnalyzerSendx Medical, Inc.20052
K170882ABL90 Flex, ABL90 Flex PlusRadiometer Medical Aps20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 6, 2006Not classifiedZ-0115-2007OtherIncorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 9 of the 16 ancestors have a recall of any class.
  • Of its descendants, K092686 (Radiometer Medical Aps, 2010) has gone furthest, with 12 descendants of its own.
  • 5 other devices also cite its predicate K041874.

How does its category compare?

All CIG devices

CIG Diazo Colorimetry, Bilirubin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CIG

2in 2023

2007: 12008: 32009: 22010: 12011: 22013: 52015: 12017: 12019: 12022: 12023: 22026 (partial year): 1
2007high 5 (2013) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CIG

64in 2013

2008: 199 days median2013: 64 days median
2007high 199 (2008) · low 64 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CIG

1in 2019

2009: 32013: 32019: 1
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CIG

52in 2025

2007: 22008: 82009: 32010: 62011: 382012: 142013: 82014: 32015: 72016: 42017: 112018: 232019: 252020: 312021: 222022: 262023: 342024: 392025: 52
2007high 52 (2025) · low 2 (2007)2025

MAUDE reports naming product code CIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—02
2008319908
20092—33
20101—06
20112—038
20120—014
201356438
20140—03
20151—07
20160—04
20171—011
20180—023
20191—125
20200—031
20210—022
20221—026
20232—034
20240—039
20250—052
20261—0—

What is it indicated for?

Indications: The ABL800 FLEX with RADIANCE v2.5 modification is intended for in vitro testing of samples of whole blood for the parameters pH, P02.pCO 2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry paramenters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF). In addition, the ABL800 Flex with RADIANCE v2.5 modification is intended for in vitro testing of samples of expired air for the parameters p02 and pCO 2. The ABL800 FLEX with RADIANCE v2.5 modification includes an AutoCheck Module to perform automated analysis of quality control fluids.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050869: Modification to ABL800 Flex.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050869

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026