ABL800 Flex

K041874510(k) TraditionalCleared Oct 8, 200488-day reviewClass II21 CFR 862.1120Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical Aps · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Oct 8, 2004 as substantially equivalent to K002290 (Radiometer Medical A/S, 2000) and 1 other predicate after an 88-day review. Its family tree reaches back 4 generations to 1986. 29 later devices descend from it. It has been recalled 22 times.

Decided
Oct 8, 2004
Review
88 days
Class
Class II
Product code
CHL
Recalls
22
Predicates
2

Its longest line of descent runs back 4 generations to K861087 (1986). 15 ancestors in all. 29 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1998: 2 other ancestors1997: 1 other ancestor1996: 3 other ancestors1995: 1 other ancestor1992: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor2005: 3 descendants cleared2010: 1 descendant cleared2006: 1 descendant cleared2015: 3 descendants cleared2020: 1 descendant cleared2008: 2 descendants cleared2014: 1 descendant cleared2017: 1 descendant cleared2012: 3 descendants cleared2013: 3 descendants cleared2025: 1 descendant cleared2021: 2 descendants cleared2011: 2 descendants cleared2016: 2 descendants cleared2024: 1 descendant cleared2026: 1 descendant cleared2023: 1 descendant clearedK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986K961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 5 other ancestors 1 generation forward: 6 descendants2 generations forward: 12 descendants3 generations forward: 6 descendants4 generations forward: 5 descendants K992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 19860
  1. Root · 4 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 3 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 2 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 1 generation backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. This deviceK041874 ABL800 FlexRadiometer Medical Aps · 2004
  6. Descendant · cites it directly · 18 of its ownK050869 Modification to ABL800 FlexRadiometer Medical Aps · 2005 · Recalled
  7. Descendant · cites it directly · 12 of its ownK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled
  8. Descendant · 2 generations on · 6 of its ownK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  9. Descendant · 3 generations on · 5 of its ownK122729 ABL90 FlexRadiometer Medical Aps · 2012
  10. Descendant · cites it directly · 4 of its ownK043218 ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps · 2005 · Recalled

Showing 5 of 29 descendants. See all descendants

Descendants cleared per year

2005: 32006: 12008: 22010: 12011: 22012: 32013: 32014: 12015: 32016: 22017: 12020: 12021: 22023: 12024: 12025: 12026: 1
2005high 3 (2005) · low 0 (2007)2026

Most influential descendants

Showing 5 of 29 · See all
  • K050869Modification to ABL800 FlexRadiometer Medical Aps · 2005 · cites it directly18descendants
  • K092686ABL90 Flex SeriesRadiometer Medical Aps · 2010 · cites it directly12descendants
  • K120197ABL90 FlexRadiometer Medical Aps · 2012 · 2 generations on6descendants
  • K122729ABL90 FlexRadiometer Medical Aps · 2012 · 3 generations on5descendants
  • K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps · 2005 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−3K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−2K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−1K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
0 (this device)K041874ABL800 FlexRadiometer Medical Aps2004—
+1K050869Modification to ABL800 FlexRadiometer Medical Aps20051
+1K092686ABL90 Flex SeriesRadiometer Medical Aps201010
+2K120197ABL90 FlexRadiometer Medical Aps20123
+3K122729ABL90 FlexRadiometer Medical Aps2012—
+1K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps20053

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
199210
199510
199630
199710
199820
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200530
200610
200820
201010
201120
201230
201330
201410
201530
201620
201710
202010
202120
202310
202410
202510
202610
FIG. 1 — Lineage of K041874, 1986–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.2 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002290Abl-700 Series, Model Abl 7XX (Various Models)
Summary page 1
…bolite Analyzer Classification Name: Blood gases and blood pH test system Predicate Devices RADIOMETER ABL700 Series Upgrade (K002290) and ABL700 with AutoCheck Module (K992859). Device Description The ABL800 FLEX is an automatic analyzer for in vitro testing o…
Radiometer Medical A/S20009
*K992859ABL700 Series Analyzer with Autocheck Module
Summary page 1
…zer Classification Name: Blood gases and blood pH test system Predicate Devices RADIOMETER ABL700 Series Upgrade (K002290) and ABL700 with AutoCheck Module (K992859). Device Description The ABL800 FLEX is an automatic analyzer for in vitro testing of blood gases, electrolytes, metabolites an…
Radiometer America, Inc.20004

Also cleared against K002290: 2 other devices →

What descends from it? (29)

See all 29

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050869Modification to ABL800 FlexRadiometer Medical Aps200518
K092686ABL90 Flex SeriesRadiometer Medical Aps201012
K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps20054
K142898ABL800 FLEX with AQURE connectivityRadiometer Medical Aps20153
K200492i-STAT CG4+ Cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.20200
K051968ABL8X7 FlexRadiometer Medical Aps20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2026Class IIZ-0190-2027No Marketing ApplicationIn vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges…
May 31, 2024Class IIZ-2426-2024Nonconforming Material/ComponentIssue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased…
May 18, 2022Class IIZ-1321-2022Software designThere is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
May 26, 2020Class IIZ-2494-2020Software designPotential risk of patient mix-up on analyzers due to software issues.
May 26, 2020Class IIZ-2493-2020Software designPotential risk of patient mix-up on analyzers due to software issues.
Mar 10, 2020Class IIZ-1684-2020Software designAnalyzer's barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample…
Mar 6, 2020Class IIZ-1836-2020Component design/selectionThrough internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay…
Jul 3, 2018Class IIZ-0682-2019Software designThe analyzer software may cause a mis-match of patient demographics and test results when a certain sequence of events occur.
Oct 24, 2017Class IIZ-0231-2018Software in the Use EnvironmentDue to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.
Jan 26, 2017Class IIZ-1463-2017Device DesignReports that the sample type "cord blood" has been changed into the two sample types "cord blood arterial" and "cord blood venous.
Jul 5, 2013Class IIZ-1075-2014Under Investigation by firmRADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.
Jul 5, 2012Class IIZ-2163-2012Software designA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are…
Jul 5, 2012Class IIZ-2162-2012Software designA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are…
Apr 16, 2009Not classifiedZ-1554-2009Process designWhen removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.
Jan 18, 2008Not classifiedZ-1654-2008Software designReported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and…
Jun 22, 2007Not classifiedZ-0038-2008Process designTube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The…
Nov 13, 2006Not classifiedZ-1095-2007Software designSoftware anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas…
Oct 31, 2005Not classifiedZ-1325-06OtherABL800 Series Blood Gas Analyzer became inoperable when the 'restore default setup' button was touched.
Sep 29, 2005Not classifiedZ-1387-06OtherUnder certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and…
Sep 19, 2005Not classifiedZ-1324-06OtherReference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.
Jun 17, 2005Not classifiedZ-1318-06OtherSoftware defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.
Jun 17, 2005Not classifiedZ-1319-06OtherThe recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 4.7 years on median.
  • 8 of the 15 ancestors have a recall of any class.
  • Of its descendants, K050869 (Radiometer Medical Aps, 2005) has gone furthest, with 18 descendants of its own.
  • 2 other devices also cite its predicate K002290.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The ABL800 FLEX is intended for in vitro testing of samples of whole blood for the parameters pH, PO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF). In addition the ABL800 FLEX is intended for in vitro testing of samples of expired air for the parameters PO2 and PCO2. The ABL800 FLEX includes an AutoCheck Module to perform automated analysis of quality control fluids.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041874: ABL800 Flex.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K041874

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026