510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

GEM Premier 5000 (Measured Parameters: pH, pCO2 and pO2)

K160412510(k) TraditionalCleared Dec 14, 2016306-day reviewClass II21 CFR 862.1120Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Instrumentation Laboratory CO · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Dec 14, 2016 as substantially equivalent to K133407 (Instrumentation Laboratory CO, 2014) after a 306-day review. Its family tree reaches back 9 generations to 1980. It has been recalled once.

Decided
Dec 14, 2016
Review
306 days
Class
Class II
Product code
CHL
Recalls
1
Predicates
1

Its longest line of descent runs back 9 generations to K801848 (1980). 24 ancestors in all, the oldest K790608 (1979). No later device cites it as a predicate yet.

  1. Root · 9 generations backK801848 ICA1 Ionized Calcium AnalyzerThe London Co. · 1980
  2. 5 generations between
  3. 3 generations backK093623 Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter)…Instrumentation Laboratory CO · 2010 · Recalled
  4. 2 generations backK112995 Gem Premier 4000Instrumentation Laboratory CO · 2012
  5. 1 generation backK133407 Gem Premier 4000Instrumentation Laboratory CO · 2014
  6. This deviceK160412 GEM Premier 5000 (Measured Parameters: pH, pCO2 and pO2)Instrumentation Laboratory CO · 2016
Show all 9 generations
  1. Root · 9 generations backK801848 ICA1 Ionized Calcium AnalyzerThe London Co. · 1980
  2. 8 generations backK963800 IL SynthesisInstrumentation Laboratory CO · 1996
  3. 7 generations backK992834 Gem Premier 3000Instrumentation Laboratory CO · 1999 · Recalled
  4. 6 generations backK022158 Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP…Instrumentation Laboratory CO · 2002
  5. 5 generations backK052121 Gem Premier 3000Instrumentation Laboratory CO · 2005 · Recalled
  6. 4 generations backK061974 Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX…Instrumentation Laboratory CO · 2006 · Recalled
  7. 3 generations backK093623 Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter)…Instrumentation Laboratory CO · 2010 · Recalled
  8. 2 generations backK112995 Gem Premier 4000Instrumentation Laboratory CO · 2012
  9. 1 generation backK133407 Gem Premier 4000Instrumentation Laboratory CO · 2014
  10. This deviceK160412 GEM Premier 5000 (Measured Parameters: pH, pCO2 and pO2)Instrumentation Laboratory CO · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K801848ICA1 Ionized Calcium AnalyzerThe London Co.1980—
−8K963800IL SynthesisInstrumentation Laboratory CO1996—
−7K992834Gem Premier 3000Instrumentation Laboratory CO19995
−6K022158Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP (Calibration…Instrumentation Laboratory CO2002—
−5K052121Gem Premier 3000Instrumentation Laboratory CO20052
−4K061974Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX; Gem CVP 3…Instrumentation Laboratory CO20064
−3K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2…Instrumentation Laboratory CO20105
−2K112995Gem Premier 4000Instrumentation Laboratory CO2012—
−1K133407Gem Premier 4000Instrumentation Laboratory CO2014—
0 (this device)K160412GEM Premier 5000 (Measured Parameters: pH, pCO2 and pO2)Instrumentation Laboratory CO2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198410
198610
198910
199220
199410
199520
199620
199710
199810
199910
200110
FIG. 1 — Lineage of K160412, 1980–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133407Gem Premier 4000
Summary page 4
…Premier 5000 (Measured Parameters: pH, pCO2 and pO2) Predicate Device GEM Premier 4000 K133407 Regulatory Information Regulation…
Instrumentation Laboratory CO2014—

Also cleared against K133407: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 20, 2026Class IIZ-1535-2026Nonconforming Material/ComponentConfirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 36 years and passes through 2 companies; a new generation was cleared every 2.9 years on median.
  • 9 of the 24 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K133407.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The GEM Premier 5000 is a portable critical care system for use by health care professionals to rapidly analyze heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH and pO2 from venous, arterial and capillary heparinized whole blood, as well as quantitative measurements of pCO2 from venous and arterial heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient’s acid/base status.

pH, pCO2 and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160412: GEM Premier 5000 (Measured Parameters: pH, pCO2 and pO2).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160412

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026