510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

FilmArray Blood Culture Identification (BCID) panel for use with FilmArray Torch

K160457510(k) SpecialCleared Mar 15, 201625-day reviewClass II21 CFR 866.3365Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biofire Diagnostics, LLC · PEN Gram-Negative Bacteria And Associated Resistance Markers

Cleared on Mar 15, 2016 as substantially equivalent to K143171 (Biofire Diagnostics, LLC, 2015) after a 25-day review. Its family tree reaches back 4 generations to 2012. It has been recalled 5 times.

Decided
Mar 15, 2016
Review
25 days
Class
Class II
Product code
PEN
Recalls
5
Predicates
1

Its longest line of descent runs back 4 generations to DEN120014 (2012). No later device cites it as a predicate yet.

  1. Root · 4 generations backDEN120014 Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP)Nanosphere, Inc. · 2012
  2. 3 generations backK122514 Verigene Gram-Positive Blood Culture (GC-GP) Nucleic Acid TestNanosphere, Inc. · 2012 · Recalled
  3. 2 generations backK130914 Filmarray Blood Culture Identification (Bcid) PanelBiofire Diagnostics, Inc. · 2013 · Recalled
  4. 1 generation backK143171 FilmArray Blood Culture Identification (BCID) Panel for use with the…Biofire Diagnostics, LLC · 2015 · Recalled
  5. This deviceK160457 FilmArray Blood Culture Identification (BCID) panel for use with…Biofire Diagnostics, LLC · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4DEN120014Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP)Nanosphere, Inc.2012—
−3K122514Verigene Gram-Positive Blood Culture (GC-GP) Nucleic Acid TestNanosphere, Inc.20128
−2K130914Filmarray Blood Culture Identification (Bcid) PanelBiofire Diagnostics, Inc.20136
−1K143171FilmArray Blood Culture Identification (BCID) Panel for use with the FilmArray…Biofire Diagnostics, LLC20155
0 (this device)K160457FilmArray Blood Culture Identification (BCID) panel for use with FilmArray TorchBiofire Diagnostics, LLC2016—
FIG. 1 — Lineage of K160457, 2012–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PEN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PEN device

  1. #1 K262168 QIAstat-Dx BCID GN Plus AMR Panelscore 12.1 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K143171FilmArray Blood Culture Identification (BCID) Panel for use with the FilmArray 2.0
Summary page 5
…ation (BCID) Panel Classification Name: Multiplex devices that use DNA hybridization to detect bacteria and their resistance markers. (21 CFR 866.3365) Predicate Device: K143171 – FilmArray Blood Culture Identification (BCID) Panel Intended Use: The FilmArray Blood Culture Identification (BCID) Panel is a…
Biofire Diagnostics, LLC20155

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 22, 2020Class IIZ-2296-2020Under Investigation by firmThere is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive…
Jan 22, 2020Class IIZ-2295-2020Under Investigation by firmThere is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive…
Mar 7, 2019Class IIZ-1133-2019Material/Component ContaminationIncreased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.
Sep 5, 2018Class IIZ-0296-2019Material/Component ContaminationIncreased risk of false positive results for Proteus when the product is used with a specific blood culture bottles.
Mar 14, 2018Class IIZ-1513-2018Material/Component ContaminationThere is an increased risk of false positive Proteus results when the product is used with specific types of BD BACTEC blood culture bottles.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 3 of the 4 ancestors have a recall of any class.

How does its category compare?

All PEN devices

PEN Gram-Negative Bacteria And Associated Resistance Markers: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · PEN

1in 2025

2013: 12014: 12015: 12016: 12018: 12019: 22020: 12022: 12024: 12025: 12026 (partial year): 1
2007high 2 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PEN

Too few PEN 510(k)s in any year since 2007 to give a median.

Recalls per year · PEN

1in 2023

2020: 12021: 12023: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PEN

13in 2025

2018: 12019: 52020: 42021: 152022: 52023: 12024: 82025: 13
2007high 15 (2021) · low 0 (2007)2025

MAUDE reports naming product code PEN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20131—00
20141—00
20151—00
20161—00
20170—00
20181—01
20192—05
20201—14
20210—115
20221—05
20230—11
20241—08
20251—013
20261—0—

What is it indicated for?

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous…

Read the full statement (349 words)

The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrate the presence of organisms as determined by Gram stain.

The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, Staphylococci (including specific differentiation of Staphylococcus aureus), Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis.

The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaKPC) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.

FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.

Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160457: FilmArray Blood Culture Identification (BCID) panel for use with FilmArray Torch.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160457

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026