PointerShell Universal, PointerShell LS

K160479510(k) SpecialCleared Mar 23, 201630-day reviewClass II21 CFR 882.4560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fiagon GmbH · PGW Ear, Nose, And Throat Stereotaxic Instrument

Cleared on Mar 23, 2016 as substantially equivalent to K141456 (Fiagon GmbH, 2014) after a 30-day review. Its family tree reaches back 5 generations to 1994. 3 of the 17 devices in its family, counting itself, have a Class I recall.

Decided
Mar 23, 2016
Review
30 days
Class
Class II
Product code
PGW
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K911783 (1994). 16 ancestors in all, the oldest K871046 (1987). No later device cites it as a predicate yet. 3 of the 17 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2014: 1 other ancestor1996: 2 other ancestors, 1 with a Class I recall12013: 1 other ancestor2000: 1 other ancestor1998: 2 other ancestors, 2 with a Class I recall21995: 3 other ancestors1987: 1 other ancestorK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K141456 · The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer, Venterapointer, Guidewire, Pointershell · Fiagon GmbH · 20141985199019952000200520102015
Other ancestors, by generation 2 generations back: 2 other ancestors, 1 with a Class I recall13 generations back: 3 other ancestors, 1 with a Class I recall14 generations back: 4 other ancestors, 1 with a Class I recall15 generations back: 2 other ancestors K141456 · The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer, Venterapointer, Guidewire, Pointershell · Fiagon GmbH · 2014K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994−50
  1. Root · 5 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 4 generations backK960714 Isg Viewing WandI.S.G. Technologies, Inc. · 1996
  3. 3 generations backK964229 Regulus NavigatorCompass Intl., Inc. · 1997
  4. 2 generations backK133573 Fiagon Navigation SystemFiagon GmbH · 2014
  5. 1 generation backK141456 The Fiagon Navigation - Extended Instrument Set Ent…Fiagon GmbH · 2014
  6. This deviceK160479 PointerShell Universal, PointerShell LSFiagon GmbH · 2016

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−4K960714Isg Viewing WandI.S.G. Technologies, Inc.1996—
−3K964229Regulus NavigatorCompass Intl., Inc.1997—
−2K133573Fiagon Navigation SystemFiagon GmbH2014—
−1K141456The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer…Fiagon GmbH2014—
0 (this device)K160479PointerShell Universal, PointerShell LSFiagon GmbH2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199530
199621
199822
200010
201310
201410
FIG. 1 — Lineage of K160479, 1994–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PGW device

  1. #1 K261540 Cube 4D Navigation System, VirtuEye Proscore 12.9 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141456The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer…
Summary page 4
…Surgery. The instruments involved in this submission are a modified version of instruments that have been approved within Fiagon Navigation extended instrument Set ENT (K141456). 4. Predicate Device Information The device described in this submission is substantially equivalent to the following com…
Fiagon GmbH2014—

Also cleared against K141456: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K960330InstatrakVisualization Technology, Inc. · 1996 · 2 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 3 companies; a new generation was cleared every 1.5 years on median.
  • 6 of the 16 ancestors have a recall of any class; 3 have a Class I recall.
  • 5 other devices also cite its predicate K141456.

How does its category compare?

All PGW devices

PGW Ear, Nose, And Throat Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · PGW

1in 2025

2014: 42015: 12016: 92017: 42018: 22019: 12020: 82021: 12022: 22023: 22025: 12026 (partial year): 3
2007high 9 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PGW

100in 2020

2014: 116 days median2016: 115 days median2017: 125 days median2020: 100 days median2026 (partial year): 65 days median
2007high 125 (2017) · low 100 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PGW

1in 2025

2019: 32021: 12023: 12025: 1
2007high 3 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PGW

220in 2025

2014: 72015: 142016: 332017: 422018: 962019: 2472020: 1612021: 1902022: 1662023: 2002024: 2742025: 220
2007high 274 (2024) · low 0 (2007)2025

MAUDE reports naming product code PGW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
2014411607
20151—014
20169115033
20174125042
20182—096
20191—3247
202081000161
20211—1190
20222—0166
20232—1200
20240—0274
20251—1220
20263650—

What is it indicated for?

The PointerShell Universal and PointerShell LS are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. Both the instruments are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such…

Read the full statement (124 words)

The PointerShell Universal and PointerShell LS are intended as an aid for precisely locating

anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon

Navigation system using electromagnetic navigation.

Both the instruments are indicated for any medical condition in which the use of stereotactic surgery

may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery,

such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model

of the anatomy.

Example procedures include, but are not limited to:

ENT Procedures;

Transphenoidal access procedures.

Intranasal procedures.

Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid

explorations, Turbinate resections, and Frontal sinusotomies.

ENT related anterior skull base procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160479: PointerShell Universal, PointerShell LS.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160479

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026