510(k) · Cleared
Recalled · Class I · 2009Instatrak 3000
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Visualization Technology, Inc. · LLZ System, Image Processing, Radiological
Cleared on Dec 31, 1998 as substantially equivalent to K960330 (Visualization Technology, Inc., 1996) and 3 other predicates after an 84-day review. Its family tree reaches back 2 generations to 1994. 8 of the 330 devices in its family, counting itself, have a Class I recall. 325 later devices descend from it. It has been recalled 6 times, including a Class I recall.
- Decided
- Dec 31, 1998
- Review
- 84 days
- Class
- Class II
- Product code
- LLZ
- Recalls
- 6 (1 Class I)
- Predicates
- 4
Its longest line of descent runs back 2 generations to K911783 (1994). 4 ancestors in all. 325 later devices descend from it, 5 citing it directly. 8 of the 330 devices in this family, counting this one, have a Class I recall.
- Root · 2 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
- 1 generation backK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
- This deviceK983529 Instatrak 3000Visualization Technology, Inc. · 1998
- Descendant · cites it directly · 294 of its ownK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
- Descendant · cites it directly · 278 of its ownK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled
- Descendant · 2 generations on · 222 of its ownK121498 PercunavPhilips Health Care · 2012
- Descendant · 3 generations on · 211 of its ownK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- Descendant · 4 generations on · 100 of its ownK160807 EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound…Philips Ultrasound, Inc. · 2016 · Recalled
Showing 5 of 325 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 325 · See all- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · cites it directly294descendants
- K053610AbarisTraxtal Technologies, Inc. · 2006 · cites it directly278descendants
- K121498PercunavPhilips Health Care · 2012 · 2 generations on222descendants
- K132304Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · 3 generations on211descendants
- K160807EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound System, Affiniti 50…Philips Ultrasound, Inc. · 2016 · 4 generations on100descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K911783 | Allegro Viewing Wand | I.S.G. Technologies, Inc. | 1994 | — |
| −1 | K981998 | Pediatric Instatrak System | Visualization Technology, Inc. | 1998 | Class I |
| 0 (this device) | K983529 | Instatrak 3000 | Visualization Technology, Inc. | 1998 | — |
| +1 | K994270 | Instatrak 3000 System with Fluorotrak Module, Model IT3000 | Visualization Technology, Inc. | 2000 | Class I |
| +1 | K053610 | Abaris | Traxtal Technologies, Inc. | 2006 | 3 |
| +2 | K121498 | Percunav | Philips Health Care | 2012 | — |
| +3 | K132304 | Epiq Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2013 | 36 |
| +4 | K160807 | EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound System… | Philips Ultrasound, Inc. | 2016 | 33 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 2 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 1 |
| 2000 | 1 | 1 |
| 2002 | 1 | 1 |
| 2004 | 1 | 0 |
| 2006 | 3 | 0 |
| 2008 | 2 | 0 |
| 2009 | 2 | 0 |
| 2010 | 2 | 0 |
| 2011 | 2 | 0 |
| 2012 | 6 | 0 |
| 2013 | 4 | 0 |
| 2014 | 12 | 0 |
| 2015 | 18 | 0 |
| 2016 | 23 | 0 |
| 2017 | 34 | 0 |
| 2018 | 22 | 0 |
| 2019 | 17 | 0 |
| 2020 | 35 | 0 |
| 2021 | 26 | 0 |
| 2022 | 24 | 0 |
| 2023 | 19 | 0 |
| 2024 | 19 | 0 |
| 2025 | 28 | 2 |
| 2026 | 23 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LLZ device
- #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K960330 | InstatrakSummary page 1…000 Common/Usual Name: Interactive Image Guided Surgical System Classification Name: Computed Tomography X-Ray System 3. PREDICATE DEVICES InstaTrak System, subject of K960330, Pediatric InstaTrak System, subject of K981998, Stealth Station manufactured by Surgical Navigation Technologies, subject of K9… | 1996 |
| * | K981998 | Pediatric Instatrak SystemSummary page 1…Surgical System Classification Name: Computed Tomography X-Ray System 3. PREDICATE DEVICES InstaTrak System, subject of K960330, Pediatric InstaTrak System, subject of K981998, Stealth Station manufactured by Surgical Navigation Technologies, subject of K954276 Viewing Wand manufactured by ISG Technolog… | 1998 |
| * | K954276 | StealthstationSummary page 1…ICES InstaTrak System, subject of K960330, Pediatric InstaTrak System, subject of K981998, Stealth Station manufactured by Surgical Navigation Technologies, subject of K954276 Viewing Wand manufactured by ISG Technologies, Inc., subject of K911783. 4. DEVICE DESCRIPTION The InstaTrak System is an image… | 1996 |
| * | K911783 | Allegro Viewing WandSummary page 1…subject of K981998, Stealth Station manufactured by Surgical Navigation Technologies, subject of K954276 Viewing Wand manufactured by ISG Technologies, Inc., subject of K911783. 4. DEVICE DESCRIPTION The InstaTrak System is an image guidance system indicated for use during sinus, skull base, cranial and… | 1994 |
What descends from it? (325)
See all 3255 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K994270 | Instatrak 3000 System with Fluorotrak Module, Model IT3000 Class I | 2000 | 294 |
| K053610 | Abaris | 2006 | 278 |
| K133573 | Fiagon Navigation System | 2014 | 28 |
| K060906 | Calypso 4D Localization System | 2006 | 11 |
| K990919 | Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended… Class I | 1999 | 0 |
Has anything in its family been recalled? (8 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 22, 2009 | Class I | Z-2251-2012 | Labeling False and Misleading | GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment. |
| Feb 6, 2006 | Not classified | Z-0709-06 | Other | Small retaining pin may detach and fall into the surgical field during stereotactic surgery. |
| Jan 26, 2005 | Not classified | Z-0704-2007 | Other | Tracking inaccuracy of 5-10 mm may result when the shunt guide accessory is used with version 5.1 software. |
| Dec 13, 2004 | Not classified | Z-0689-05 | Other | Enhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised. |
| Jul 12, 2004 | Not classified | Z-0814-05 | Other | Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit. |
| Jul 12, 2004 | Not classified | Z-0815-05 | Other | Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit. |
The 5 latest of 6 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K960330InstatrakVisualization Technology, Inc. · 1996 · Direct predicate · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Cites it directly · ◆ in the 2000 descendants bar of FIG. 12 Class I recalls
- K990919Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended Straight…Visualization Technology, Inc. · 1999 · Cites it directly · ◆ in the 1999 descendants bar of FIG. 11 Class I recall
- K022069Oec Fluorotrak 9800 PlusGe Oec Medical Systems · 2002 · 2 generations on · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
- K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
- K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
Nearest 5 of 7; all are on the web page.
What does this lineage say?
- The line of descent spans 4 years and passes through 2 companies; a new generation was cleared every 4.4 years on median.
- 3 of the 4 ancestors have a recall of any class; 2 have a Class I recall.
- Of its descendants, K994270 (Visualization Technology, Inc., 2000) has gone furthest, with 294 descendants of its own.
- 7 other devices also cite its predicate K960330.
How does its category compare?
All LLZ devicesLLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LLZ
58in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LLZ
173in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LLZ
11in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LLZ
116in 2025
MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 76 | 36 | 20 | 152 |
| 2008 | 94 | 52 | 32 | 154 |
| 2009 | 70 | 62 | 21 | 145 |
| 2010 | 51 | 98 | 17 | 145 |
| 2011 | 106 | 79 | 30 | 454 |
| 2012 | 82 | 87 | 12 | 266 |
| 2013 | 95 | 84 | 64 | 1,130 |
| 2014 | 78 | 89 | 6 | 15,631 |
| 2015 | 64 | 90 | 16 | 211 |
| 2016 | 82 | 90 | 12 | 290 |
| 2017 | 78 | 107 | 5 | 104 |
| 2018 | 104 | 105 | 7 | 72 |
| 2019 | 74 | 107 | 2 | 88 |
| 2020 | 81 | 115 | 6 | 82 |
| 2021 | 83 | 116 | 3 | 70 |
| 2022 | 65 | 137 | 4 | 28 |
| 2023 | 87 | 157 | 8 | 26 |
| 2024 | 62 | 131 | 6 | 42 |
| 2025 | 58 | 173 | 11 | 116 |
| 2026 | 45 | 110 | 14 | — |
What is it indicated for?
The InstaTrak 3000 System is intended as an aid to the surgeon for precisely locating anatomical structures anywhere on the human body during either open or percutaneous procedures. It is indicated for any medical condition that may benefit from the use of stereotactic surgery and which provides a reference to rigid anatomical structures such as sinus, skull, cranial, a long bone or vertebra, visible on medical images such as CT or MR.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K983529 · printed Nov 4, 2026
Cite this page
510(k) Report. “K983529: Instatrak 3000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K983529
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026