510(k) Report

510(k) · Cleared

Recalled · Class I · 2013

Instatrak

K960330510(k) TraditionalCleared Apr 8, 199675-day reviewClass II21 CFR 892.2050Recalls 9 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Visualization Technology, Inc. · LLZ System, Image Processing, Radiological

Cleared on Apr 8, 1996 after a 75-day review. The FDA summary names its predicate without a number. 9 of the 329 devices in its family, counting itself, have a Class I recall. 328 later devices descend from it. It has been recalled 9 times, including a Class I recall.

Decided
Apr 8, 1996
Review
75 days
Class
Class II
Product code
LLZ
Recalls
9 (1 Class I)
Predicates
0

328 later devices descend from it, 8 citing it directly. 9 of the 329 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2006: 3 descendants cleared1998: 3 descendants cleared, 3 with a Class I recall32014: 12 descendants cleared1999: 1 descendant cleared, 1 with a Class I recall12008: 2 descendants cleared2009: 2 descendants cleared2012: 6 descendants cleared2015: 18 descendants cleared2010: 2 descendants cleared2011: 2 descendants cleared2021: 26 descendants cleared2016: 23 descendants cleared2020: 35 descendants cleared2017: 34 descendants cleared2000: 1 descendant cleared, 1 with a Class I recall12013: 4 descendants cleared2018: 22 descendants cleared2022: 24 descendants cleared2019: 17 descendants cleared2026: 23 descendants cleared2002: 1 descendant cleared, 1 with a Class I recall12004: 1 descendant cleared2023: 19 descendants cleared2025: 28 descendants cleared, 2 with a Class I recall22024: 19 descendants cleared1995200020052010201520202025
  1. This deviceK960330 InstatrakVisualization Technology, Inc. · 1996
  2. Descendant · cites it directly · 327 of its ownK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  3. Descendant · cites it directly · 325 of its ownK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
  4. Descendant · 2 generations on · 294 of its ownK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
  5. Descendant · cites it directly · 280 of its ownK982994 Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  6. Descendant · cites it directly · 278 of its ownK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled

Showing 5 of 328 descendants. See all descendants

Descendants cleared per year

1998: 31999: 12000: 12002: 12004: 12006: 32008: 22009: 22010: 22011: 22012: 62013: 42014: 122015: 182016: 232017: 342018: 222019: 172020: 352021: 262022: 242023: 192024: 192025: 282026: 23
1998high 35 (2020) · low 0 (2001)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 328 · See all
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · cites it directly327descendants
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · cites it directly325descendants
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 2 generations on294descendants
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · cites it directly280descendants
  • K053610AbarisTraxtal Technologies, Inc. · 2006 · cites it directly278descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K960330InstatrakVisualization Technology, Inc.1996—
+1K981998Pediatric Instatrak SystemVisualization Technology, Inc.1998Class I
+1K983529Instatrak 3000Visualization Technology, Inc.1998Class I
+2K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc.2000Class I
+1K982994Skull Base Instatrak SystemVisualization Technology, Inc.1998Class I
+1K053610AbarisTraxtal Technologies, Inc.20063

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199833
199911
200011
200211
200410
200630
200820
200920
201020
201120
201260
201340
2014120
2015180
2016230
2017340
2018220
2019170
2020350
2021260
2022240
2023190
2024190
2025282
2026230
FIG. 1 — Lineage of K960330, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Viewing Wand
Named without a submission number
—
*—Barnes Suction Tube
Named without a submission number
—

What descends from it? (328)

See all 328

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K981998Pediatric Instatrak System Class IVisualization Technology, Inc.1998327
K983529Instatrak 3000 Class IVisualization Technology, Inc.1998325
K982994Skull Base Instatrak System Class IVisualization Technology, Inc.1998280
K053610AbarisTraxtal Technologies, Inc.2006278
K133573Fiagon Navigation SystemFiagon GmbH201428
K141456The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer…Fiagon GmbH201413
K060906Calypso 4D Localization SystemCalypso Medical Technologies, Inc.200611
K990919Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended… Class IVisualization Technology, Inc.19990

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 26, 2013Class IIZ-0524-2014Software Design ChangeGE OEC Medical Systems Inc is recalling InstaTrak 3500 Plus, ENTrakTM Plus, InstaTrak 3500, ENTrakTM Computer-Assisted Navigation Systems (USA) due to several potential safety issues related to the use of these products.
Jun 22, 2009Class IZ-2251-2012Labeling False and MisleadingGE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment.
Oct 18, 2007Not classifiedZ-0183-2008Device DesignCrack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.
May 18, 2007Not classifiedZ-1130-2007OtherCleaning probes used to unclog disposable aspirators may shed small particles of nylon material which may be transferred into the nasal cavity.
Dec 13, 2004Not classifiedZ-0687-05OtherEnhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised.
Dec 13, 2004Not classifiedZ-0688-05OtherEnhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised.
Jul 12, 2004Not classifiedZ-0813-05OtherNavigation inaccuracy may result if the headset registration method is used with the Axcess System Kit.
Jul 12, 2004Not classifiedZ-0812-05OtherNavigation inaccuracy may result if the headset registration method is used with the Axcess System Kit.
Jul 12, 2004Not classifiedZ-0814-05OtherNavigation inaccuracy may result if the headset registration method is used with the Axcess System Kit.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
  • K990919Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended Straight…Visualization Technology, Inc. · 1999 · Cites it directly · ◆ in the 1999 descendants bar of FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 2 generations on · ◆ in the 2000 descendants bar of FIG. 12 Class I recalls
  • K022069Oec Fluorotrak 9800 PlusGe Oec Medical Systems · 2002 · 3 generations on · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
  • K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K981998 (Visualization Technology, Inc., 1998) has gone furthest, with 327 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Intranasal or sinuses

Acute and chronic sinusitis, endoscopic dacryocystorhinostomy, optic nerve and orbital decompression, the removal of polyps, the biopsy and removal of tumors, and the repair of CSF leaks, pituitary disorder, and encephalocele

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K960330: Instatrak.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K960330

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026