510(k) · Cleared
Recalled · Class I · 2013Instatrak
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Visualization Technology, Inc. · LLZ System, Image Processing, Radiological
Cleared on Apr 8, 1996 after a 75-day review. The FDA summary names its predicate without a number. 9 of the 329 devices in its family, counting itself, have a Class I recall. 328 later devices descend from it. It has been recalled 9 times, including a Class I recall.
- Decided
- Apr 8, 1996
- Review
- 75 days
- Class
- Class II
- Product code
- LLZ
- Recalls
- 9 (1 Class I)
- Predicates
- 0
328 later devices descend from it, 8 citing it directly. 9 of the 329 devices in this family, counting this one, have a Class I recall.
- This deviceK960330 InstatrakVisualization Technology, Inc. · 1996
- Descendant · cites it directly · 327 of its ownK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
- Descendant · cites it directly · 325 of its ownK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
- Descendant · 2 generations on · 294 of its ownK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
- Descendant · cites it directly · 280 of its ownK982994 Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
- Descendant · cites it directly · 278 of its ownK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled
Showing 5 of 328 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 328 · See all- K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · cites it directly327descendants
- K983529Instatrak 3000Visualization Technology, Inc. · 1998 · cites it directly325descendants
- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 2 generations on294descendants
- K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · cites it directly280descendants
- K053610AbarisTraxtal Technologies, Inc. · 2006 · cites it directly278descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K960330 | Instatrak | Visualization Technology, Inc. | 1996 | — |
| +1 | K981998 | Pediatric Instatrak System | Visualization Technology, Inc. | 1998 | Class I |
| +1 | K983529 | Instatrak 3000 | Visualization Technology, Inc. | 1998 | Class I |
| +2 | K994270 | Instatrak 3000 System with Fluorotrak Module, Model IT3000 | Visualization Technology, Inc. | 2000 | Class I |
| +1 | K982994 | Skull Base Instatrak System | Visualization Technology, Inc. | 1998 | Class I |
| +1 | K053610 | Abaris | Traxtal Technologies, Inc. | 2006 | 3 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 3 | 3 |
| 1999 | 1 | 1 |
| 2000 | 1 | 1 |
| 2002 | 1 | 1 |
| 2004 | 1 | 0 |
| 2006 | 3 | 0 |
| 2008 | 2 | 0 |
| 2009 | 2 | 0 |
| 2010 | 2 | 0 |
| 2011 | 2 | 0 |
| 2012 | 6 | 0 |
| 2013 | 4 | 0 |
| 2014 | 12 | 0 |
| 2015 | 18 | 0 |
| 2016 | 23 | 0 |
| 2017 | 34 | 0 |
| 2018 | 22 | 0 |
| 2019 | 17 | 0 |
| 2020 | 35 | 0 |
| 2021 | 26 | 0 |
| 2022 | 24 | 0 |
| 2023 | 19 | 0 |
| 2024 | 19 | 0 |
| 2025 | 28 | 2 |
| 2026 | 23 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LLZ device
- #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Viewing Wand Named without a submission number | |
| * | — | Barnes Suction Tube Named without a submission number |
What descends from it? (328)
See all 3288 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K981998 | Pediatric Instatrak System Class I | 1998 | 327 |
| K983529 | Instatrak 3000 Class I | 1998 | 325 |
| K982994 | Skull Base Instatrak System Class I | 1998 | 280 |
| K053610 | Abaris | 2006 | 278 |
| K133573 | Fiagon Navigation System | 2014 | 28 |
| K141456 | The Fiagon Navigation - Extended Instrument Set Ent, Registrationpointer… | 2014 | 13 |
| K060906 | Calypso 4D Localization System | 2006 | 11 |
| K990919 | Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended… Class I | 1999 | 0 |
Has anything in its family been recalled? (9 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 26, 2013 | Class II | Z-0524-2014 | Software Design Change | GE OEC Medical Systems Inc is recalling InstaTrak 3500 Plus, ENTrakTM Plus, InstaTrak 3500, ENTrakTM Computer-Assisted Navigation Systems (USA) due to several potential safety issues related to the use of these products. |
| Jun 22, 2009 | Class I | Z-2251-2012 | Labeling False and Misleading | GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment. |
| Oct 18, 2007 | Not classified | Z-0183-2008 | Device Design | Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field. |
| May 18, 2007 | Not classified | Z-1130-2007 | Other | Cleaning probes used to unclog disposable aspirators may shed small particles of nylon material which may be transferred into the nasal cavity. |
| Dec 13, 2004 | Not classified | Z-0687-05 | Other | Enhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised. |
| Dec 13, 2004 | Not classified | Z-0688-05 | Other | Enhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised. |
| Jul 12, 2004 | Not classified | Z-0813-05 | Other | Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit. |
| Jul 12, 2004 | Not classified | Z-0812-05 | Other | Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit. |
| Jul 12, 2004 | Not classified | Z-0814-05 | Other | Navigation inaccuracy may result if the headset registration method is used with the Axcess System Kit. |
The 5 latest of 9 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
- K983529Instatrak 3000Visualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
- K990919Straight Aspirator, 90 Degree Aspirator,45 Degree Aspirator,7 French Aspirator,Extended Straight…Visualization Technology, Inc. · 1999 · Cites it directly · ◆ in the 1999 descendants bar of FIG. 11 Class I recall
- K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 2 generations on · ◆ in the 2000 descendants bar of FIG. 12 Class I recalls
- K022069Oec Fluorotrak 9800 PlusGe Oec Medical Systems · 2002 · 3 generations on · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
- K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
- K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
Nearest 5 of 8; all are on the web page.
What does this lineage say?
- Of its descendants, K981998 (Visualization Technology, Inc., 1998) has gone furthest, with 327 descendants of its own.
How does its category compare?
All LLZ devicesLLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LLZ
58in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LLZ
173in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LLZ
11in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LLZ
116in 2025
MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 76 | 36 | 20 | 152 |
| 2008 | 94 | 52 | 32 | 154 |
| 2009 | 70 | 62 | 21 | 145 |
| 2010 | 51 | 98 | 17 | 145 |
| 2011 | 106 | 79 | 30 | 454 |
| 2012 | 82 | 87 | 12 | 266 |
| 2013 | 95 | 84 | 64 | 1,130 |
| 2014 | 78 | 89 | 6 | 15,631 |
| 2015 | 64 | 90 | 16 | 211 |
| 2016 | 82 | 90 | 12 | 290 |
| 2017 | 78 | 107 | 5 | 104 |
| 2018 | 104 | 105 | 7 | 72 |
| 2019 | 74 | 107 | 2 | 88 |
| 2020 | 81 | 115 | 6 | 82 |
| 2021 | 83 | 116 | 3 | 70 |
| 2022 | 65 | 137 | 4 | 28 |
| 2023 | 87 | 157 | 8 | 26 |
| 2024 | 62 | 131 | 6 | 42 |
| 2025 | 58 | 173 | 11 | 116 |
| 2026 | 45 | 110 | 14 | — |
What is it indicated for?
Intranasal or sinuses
Acute and chronic sinusitis, endoscopic dacryocystorhinostomy, optic nerve and orbital decompression, the removal of polyps, the biopsy and removal of tumors, and the repair of CSF leaks, pituitary disorder, and encephalocele
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K960330 · printed Nov 6, 2026
Cite this page
510(k) Report. “K960330: Instatrak.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K960330