510(k) Report

510(k) · Cleared

Cleared

PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator

K160595510(k) TraditionalCleared Nov 18, 2016262-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pms Medikal Ambalaj Sanayi VE Ticaret Ltd Sti · FRG Wrap, Sterilization

Cleared on Nov 18, 2016 as substantially equivalent to K140487 (STERIS Corporation, 2014) after a 262-day review. Its family tree reaches back 4 generations to 1977.

Decided
Nov 18, 2016
Review
262 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K771032 (1977). 7 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK771032 Pouches, Peel Open, BreathableSurgicot, Inc. · 1977
  2. 3 generations backK070765 Vis-U-All II Self Seal PouchSTERIS Corporation · 2007
  3. 2 generations backK090371 Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-Pro 1…STERIS Corporation · 2009
  4. 1 generation backK140487 Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2014
  5. This deviceK160595 PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with…Pms Medikal Ambalaj Sanayi VE Ticaret Ltd Sti · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K771032Pouches, Peel Open, BreathableSurgicot, Inc.1977—
−3K070765Vis-U-All II Self Seal PouchSTERIS Corporation2007—
−2K090371Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-Pro 1 Sterilization…STERIS Corporation2009—
−1K140487Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation2014—
0 (this device)K160595PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical…Pms Medikal Ambalaj Sanayi VE Ticaret Ltd Sti2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
199510
200710
FIG. 1 — Lineage of K160595, 1977–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140487Vis-U-All Low Temperature Sterilization Pouch/Tubing
Summary page 6
…icator, Physical/Chemical Sterilization Process Product Code: FRG Product Classification: Class II, 21 CFR 880.6850, General Hospital (FRG) 3. Predicate Device K140487, STERIS Vis-U-All Low Temperature Sterilization Pouch/Tubing 4.Indications for use PMS Steripack Tyvek Sterilization Pouches (T…
STERIS Corporation2014—

Also cleared against K140487: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 39 years and passes through 3 companies; a new generation was cleared every 5.2 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • 1 other device also cites its predicate K140487.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method. PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a…

Read the full statement (296 words)

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator; are intended to provide healthcare workers with an effective method to enclose medical devices intended for sterilization in vaporized hydrogen peroxide (VHP) sterilization method.

PMS Steripack Tyvek Sterilization Pouches (TP) and Rolls (TY) with Chemical Indicator is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical devices and also maintain sterility of the enclosed devices until used.

The recommended (and validated) sterilization cycle parameters are:

The Steris V-PRO 1 PLUS LUMEN CYCLE is a single preprogramed standard sterilization cycle with NO cycle parameter variables allowed by the end user.

Chemical process indicator on the exterior of the Tyvek pouches and rolls indicates that the pouch has been exposed to V-PRO 1 Plus, Lumen Cycle sterilization process by changing color from red to blue or (lighter).

Devices with stainless steel lumens with the following internal diameter (ID) and maximum length can be processed in V-PRO 1 Plus, Lumen Cycle.

Table 1. Lumen Claims

Pouch Size: 50 x 200mm; Max. lumen length: 50 mm; Min. internal diameter: 1.5 mm; Max # Lumens: 1

Pouch Size: 150 x 300mm; Max. lumen length: 3 mm; Min. internal diameter: 200 mm; Max # Lumens: 1

Pouch Size: 350 x 600mm; Max. lumen length: 3 mm; Min. internal diameter: 400 mm; Max # Lumens: 1

Worst Case Load:

The worst case load for the 50 mm x 200 mm pouch is 0.058 lbs. of instruments.

The worst case load for the 150 mm x 300 mm pouch is 0.630 lbs. of instruments.

The worst case load for the 350 mm x 600 mm pouch is 3.054 lbs. of instruments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160595: PMSSteripack Tyvek Sterilization Pouch (TP) and Roll (TY) with Chemical Indicator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160595

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026