510(k) Report

510(k) · Cleared

Cleared

Vis-U-All Low Temperature Sterilization Pouch/Tubing

K160908510(k) TraditionalCleared Jul 22, 2016112-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · FRG Wrap, Sterilization

Cleared on Jul 22, 2016 as substantially equivalent to K140487 (STERIS Corporation, 2014) after a 112-day review. Its family tree reaches back 4 generations to 1977. 6 later devices descend from it.

Decided
Jul 22, 2016
Review
112 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K771032 (1977). 7 ancestors in all. 6 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 2 descendants K140487 · Vis-U-All Low Temperature Sterilization Pouch/Tubing · STERIS Corporation · 2014K090371 · Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-Pro 1 Sterilization System · STERIS Corporation · 2009K070765 · Vis-U-All II Self Seal Pouch · STERIS Corporation · 2007K771032 · Pouches, Peel Open, Breathable · Surgicot, Inc. · 19770
  1. Root · 4 generations backK771032 Pouches, Peel Open, BreathableSurgicot, Inc. · 1977
  2. 3 generations backK070765 Vis-U-All II Self Seal PouchSTERIS Corporation · 2007
  3. 2 generations backK090371 Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-Pro 1…STERIS Corporation · 2009
  4. 1 generation backK140487 Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2014
  5. This deviceK160908 Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2016
  6. Descendant · cites it directly · 5 of its ownK172749 Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2018
  7. Descendant · 2 generations on · 3 of its ownK183401 Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2019
  8. Descendant · 3 generations on · 2 of its ownK222440 Vis-U-All Low Temperature Sterilization Pouches/Tubing, PRO-LITE…Steris · 2022
  9. Descendant · 4 generations on · 0 of its ownK231501 PRO-LITE Sterilization TraySTERIS Corporation · 2023
  10. Descendant · 4 generations on · 0 of its ownK231500 Vis-U-All Low Temperature Sterilization PouchesSTERIS Corporation · 2023

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2018: 12019: 22022: 12023: 2
2018high 2 (2019) · low 0 (2020)2026

Most influential descendants

Showing 5 of 6 · See all
  • K172749Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2018 · cites it directly5descendants
  • K183401Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation · 2019 · 2 generations on3descendants
  • K222440Vis-U-All Low Temperature Sterilization Pouches/Tubing, PRO-LITE Sterilization TraysSteris · 2022 · 3 generations on2descendants
  • K231501PRO-LITE Sterilization TraySTERIS Corporation · 2023 · 4 generations on0descendants
  • K231500Vis-U-All Low Temperature Sterilization PouchesSTERIS Corporation · 2023 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K771032Pouches, Peel Open, BreathableSurgicot, Inc.1977—
−3K070765Vis-U-All II Self Seal PouchSTERIS Corporation2007—
−2K090371Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-Pro 1 Sterilization…STERIS Corporation2009—
−1K140487Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation2014—
0 (this device)K160908Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation2016—
+1K172749Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation2018—
+2K183401Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation2019—
+3K222440Vis-U-All Low Temperature Sterilization Pouches/Tubing, PRO-LITE Sterilization…Steris2022—
+4K231501PRO-LITE Sterilization TraySTERIS Corporation2023—
+4K231500Vis-U-All Low Temperature Sterilization PouchesSTERIS Corporation2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
199510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201920
202210
202320
FIG. 1 — Lineage of K160908, 1977–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140487Vis-U-All Low Temperature Sterilization Pouch/Tubing
Summary page 6
…ion wrap (21 CFR 880.6850, Product Code FRG). 2. Predicate Device Vis-U-All Low Temperature Sterilization Pouch/Tubing (K140487) Reference Device - Vis-U-All Low Temperature Sterilization Pouch/Tubing (K092745) 3. Description of Device…
STERIS Corporation2014—
†K092745Vis-U-All Low Temperature Tyvek Sterilization Pouch for Ethylene Oxide, Models: Self and…STERIS Corporation20101

Also cleared against K140487: 1 other device →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172749Vis-U-All Low Temperature Sterilization Pouch/TubingSTERIS Corporation20185

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 39 years and passes through 2 companies; a new generation was cleared every 5.2 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • Of its descendants, K172749 (STERIS Corporation, 2018) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K140487.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose: • medical devices in a single or double pouch configuration • trays* containing medical devices in a single or double pouch configuration • small items requiring surface sterilization in a single pouch configuration within a tray* to be sterilized in the Lumen, Non Lumen and Flexible Cycles of the V-PRO Low…

Read the full statement (630 words)

The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

• medical devices in a single or double pouch configuration

• trays* containing medical devices in a single or double pouch configuration

• small items requiring surface sterilization in a single pouch configuration within a tray*

to be sterilized in the Lumen, Non Lumen and Flexible Cycles of the V-PRO Low Temperature Sterilization Systems. The pouches maintain the sterility of the enclosed devices until used.

When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

Vis-U-All Pouch/Tubing Claims with V-PRO Sterilization Systems

Intended Pouch Load when Medical Devices are:

• Directly Pouched or

• Placed Inside of a Tray* and the Tray* Pouched

V-PRO 60 Lumen Cycle

• Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

• Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

• Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:

o single or dual lumen devices

▪ ≥ 0.77 mm ID and ≤ 410 mm in length

o triple lumen devices

▪ ≥ 1.2 mm ID and ≤ 275 mm in length

▪ ≥ 1.8 mm ID and ≤ 310 mm in length

or

▪ ≥ 2.8 mm ID and ≤ 317 mm in length

V-PRO 60 Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO 60 Flexible Cycle

One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a single or dual lumen device with lumens that are > 1 mm ID and < 990 mm in length

V-PRO 1, 1 Plus & maX Lumen Cycle

• Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

• Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

• Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:

o single or dual lumen devices

▪ ≥ 0.77 mm ID and ≤ 527 mm in length

o triple lumen devices

▪ ≥ 1.2 mm ID and ≤ 275 mm in length

▪ ≥ 1.8 mm ID and ≤ 310 mm in length

or

▪ ≥ 2.8 mm ID and ≤ 317 mm in length

V-PRO 1 Plus & maX Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO maX Flexible Cycle

Load 1: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes with a light cord (if not integral to endoscope) and mat with no additional load.

The flexible endoscopes may contain either:

• a single lumen that is > 1 mm ID and < 1050 mm in length

• or two lumens with:

• one lumen that is > 1 mm ID and < 990 mm in length

• and the other lumen that is > 1 mm ID and < 850 mm in length

Load 2: Non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.

*Trays must be legally marketed for use in the V-PRO Low Temperature Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160908: Vis-U-All Low Temperature Sterilization Pouch/Tubing.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160908

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026