510(k) Report

510(k) · Cleared

Cleared

SYSTEM 1E Liquid Chemical Sterilant Processing System

K161683510(k) SpecialCleared Aug 5, 201649-day reviewClass II21 CFR 880.6885

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MED Sterilant, Medical Devices

Cleared on Aug 5, 2016 after a 49-day review. The FDA summary names no predicate by number. 15 later devices descend from it.

Decided
Aug 5, 2016
Review
49 days
Class
Class II
Product code
MED
Recalls
None
Predicates
0

15 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2017: 1 descendant cleared2018: 2 descendants cleared2019: 4 descendants cleared2021: 2 descendants cleared2023: 3 descendants cleared2022: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant cleared201620182020202220242026
  1. This deviceK161683 SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2016
  2. Descendant · cites it directly · 14 of its ownK170956 SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2017
  3. Descendant · 2 generations on · 12 of its ownK173256 System 1 endo Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2018
  4. Descendant · 3 generations on · 10 of its ownK182827 SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model 6900STERIS Corporation · 2019
  5. Descendant · 2 generations on · 10 of its ownK180342 SYSTEM 1E Liquid Chemical Sterilant Processing SystemSteris Corporations · 2018
  6. Descendant · 4 generations on · 9 of its ownK191343 SYSTEM 1 endo Liquid Chemical Sterilant Processing System,Model P6900STERIS Corporation · 2019

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2017: 12018: 22019: 42021: 22022: 12023: 32024: 12026: 1
2017high 4 (2019) · low 0 (2020)2026

Most influential descendants

Showing 5 of 15 · See all
  • K170956SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2017 · cites it directly14descendants
  • K173256System 1 endo Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2018 · 2 generations on12descendants
  • K182827SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model 6900STERIS Corporation · 2019 · 3 generations on10descendants
  • K180342SYSTEM 1E Liquid Chemical Sterilant Processing SystemSteris Corporations · 2018 · 2 generations on10descendants
  • K191343SYSTEM 1 endo Liquid Chemical Sterilant Processing System,Model P6900STERIS Corporation · 2019 · 4 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K161683SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation2016—
+1K170956SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation2017—
+2K173256System 1 endo Liquid Chemical Sterilant Processing SystemSTERIS Corporation2018—
+3K182827SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model 6900STERIS Corporation2019—
+2K180342SYSTEM 1E Liquid Chemical Sterilant Processing SystemSteris Corporations2018—
+4K191343SYSTEM 1 endo Liquid Chemical Sterilant Processing System,Model P6900STERIS Corporation2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201820
201940
202120
202210
202330
202410
202610
FIG. 1 — Lineage of K161683, 2016–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (15)

See all 15

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170956SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation201714

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K170956 (STERIS Corporation, 2017) has gone furthest, with 14 descendants of its own.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46 to 55°C, and rinses the load with…

Read the full statement (194 words)

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.

The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46 to 55°C, and rinses the load with extensively treated* potable water. The processed load should be used immediately.

The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

* The extensive treatment of EPA potable water consists of:

1. Pre-filtration through two pre-filters:

• Pre-filter 1 is a gross depth filter that removes approximately 5 micron or larger particles/contaminants.

• Pre-filter 2 is a surface filter that removes particles/contaminants > 0.1 micron.

2. UV Irradiation:

• During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.

3. 0.1 micron filtration:

• The water prepared by pre-filtration and UV irradiation is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161683: SYSTEM 1E Liquid Chemical Sterilant Processing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K161683

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026