510(k) Report

510(k) · Cleared

Recalled · 2009

Sterrad 50 Sterilizer

K981625510(k) TraditionalCleared Jan 11, 1999249-day reviewClass II21 CFR 880.6860Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Sterilization Products · MLR Sterilizer, Chemical

Cleared on Jan 11, 1999 after a 249-day review. The FDA summary names its predicate without a number. 19 later devices descend from it. It has been recalled 5 times.

Decided
Jan 11, 1999
Review
249 days
Class
Class II
Product code
MLR
Recalls
5
Predicates
0

19 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2012: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2024: 2 descendants cleared2016: 5 descendants cleared2018: 2 descendants cleared2017: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2019: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant cleared1995200020052010201520202025
  1. This deviceK981625 Sterrad 50 SterilizerAdvanced Sterilization Products · 1999
  2. Descendant · cites it directly · 18 of its ownK023290 Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products · 2003 · Recalled
  3. Descendant · 2 generations on · 17 of its ownK111377 Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · Recalled
  4. Descendant · 3 generations on · 9 of its ownK142454 STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · Recalled
  5. Descendant · 3 generations on · 5 of its ownK141163 Sterizone SterilizerTso3, Inc. · 2014
  6. Descendant · 3 generations on · 4 of its ownK160903 STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2003: 12012: 12014: 12015: 12016: 52017: 12018: 22019: 12021: 12022: 12024: 22025: 12026: 1
2003high 5 (2016) · low 0 (2004)2026

Most influential descendants

Showing 5 of 19 · See all
  • K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products · 2003 · cites it directly18descendants
  • K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · 2 generations on17descendants
  • K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · 3 generations on9descendants
  • K141163Sterizone SterilizerTso3, Inc. · 2014 · 3 generations on5descendants
  • K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K981625Sterrad 50 SterilizerAdvanced Sterilization Products1999—
+1K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products20039
+2K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products20121
+3K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products20152
+3K141163Sterizone SterilizerTso3, Inc.2014—
+3K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
201210
201410
201510
201650
201710
201820
201910
202110
202210
202420
202510
202610
FIG. 1 — Lineage of K981625, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—STERRAD® 100 Sterilization System
Named without a submission number
—
*—STERI-VAC Gas Sterilizer Model 5XL
Named without a submission number
—

What descends from it? (19)

See all 19

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products200318

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 9, 2009Not classifiedZ-1419-2009Component design/selectionASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs.
Jan 12, 2009Not classifiedZ-1570-2009OtherThe capacitor in the vacuum pump component within the Sterrad 50 System could overheat and potentially create a hazardous condition inside the system, resulting in smoke and/or fire.
Feb 25, 2008Not classifiedZ-1629-2008Device DesignOil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."
Nov 12, 2007Not classifiedZ-0844-2008OtherOutdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activities
Nov 12, 2007Not classifiedZ-0845-2008Employee errorUser Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K023290 (Advanced Sterilization Products, 2003) has gone furthest, with 18 descendants of its own.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The STERRAD® 50 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. (See list of recommended Materials in Section 3 of the Operator’s Manual.) The STERRAD® 50 Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged…

Read the full statement (158 words)

The STERRAD® 50 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. (See list of recommended Materials in Section 3 of the Operator’s Manual.)

The STERRAD® 50 Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Metal and nonmetal lumened instruments with inside diameters of 6 mm (% inch) or larger and lengths of 310 mm (12 inches) or shorter can be processed in the STERRAD® 50 Sterilizer. (See list of recommended Materials in Section 3 of the Operator’s Manual.)

Additionally, medical devices with only a single stainless steel lumen which has an inside diameter of 3 mm (1/8 inch) or larger and a length of 400 mm (15 34 inches) or shorter can be processed in the STERRAD® 50 Sterilizer.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K981625: Sterrad 50 Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K981625

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026