510(k) Report

510(k) · Cleared

Cleared

LOWTEM Crystal 120

K231893510(k) TraditionalCleared Sep 18, 2024448-day reviewClass II21 CFR 880.6860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lowtem Co., Ltd. · MLR Sterilizer, Chemical

Cleared on Sep 18, 2024 as substantially equivalent to K062297 (STERIS Corporation, 2007) and 1 other predicate after a 448-day review. Its family tree reaches back 4 generations to 2003.

Decided
Sep 18, 2024
Review
448 days
Class
Class II
Product code
MLR
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K020875 (2003). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK020875 TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc. · 2003
  2. 3 generations backK051595 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2006
  3. 2 generations backK090636 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2009
  4. 1 generation backK102330 Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2011
  5. This deviceK231893 LOWTEM Crystal 120Lowtem Co., Ltd. · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K020875TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc.2003—
−3K051595TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2006—
−2K090636TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2009—
−1K102330Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2011—
0 (this device)K231893LOWTEM Crystal 120Lowtem Co., Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
200710
FIG. 1 — Lineage of K231893, 2003–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062297VHP 136 Low Temperature Sterilization System
Summary page 5
…sification: Class II Review Panel: General Hospital Predicate Devices:  Amsco® V-PROTM 1 Low Temperature Sterilization System – K062297  Amsco® V-PROTM MAX Low Temperature Sterilization System – K102330 3. Device Description The LOWTEM Crystal 120 low-tem…
STERIS Corporation20071
*K102330Amsco V-Pro Max Low Temperature Sterilization System
Summary page 5
…al Hospital Predicate Devices:  Amsco® V-PROTM 1 Low Temperature Sterilization System – K062297  Amsco® V-PROTM MAX Low Temperature Sterilization System – K102330 3. Device Description The LOWTEM Crystal 120 low-temperature vaporized hydrogen peroxide sterilizer is a self- contained st…
STERIS Corporation2011—

Also cleared against K062297: 7 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 3.4 years on median.
  • 2 of the 6 ancestors have a recall of any class.
  • 7 other devices also cite its predicate K062297.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The LOWTEM Crystal 120 Sterilizer System is indicated to sterilize pre-cleaned and dried reusable medical devices with nonabsorbable materials. The Crystal 120 sterilizer’s Standard cycle can sterilize: * Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors, * Medical devices with single stainless-steel lumen with: An inside diameter of 1mm or larger and a length of 125mm or shorter An inside diameter of 2mm or…

Read the full statement (212 words)

The LOWTEM Crystal 120 Sterilizer System is indicated to sterilize pre-cleaned and dried reusable medical devices with nonabsorbable materials. The Crystal 120 sterilizer’s Standard cycle can sterilize:

* Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors,

* Medical devices with single stainless-steel lumen with:

An inside diameter of 1mm or larger and a length of 125mm or shorter

An inside diameter of 2mm or larger and a length of 250mm or shorter

An inside diameter of 3mm or larger and a length of 400mm or shorter

※ The validation testing for this lumen size was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.

- Contra-indication: It is contraindicated to sterilize single usable medical devices and devices made up with absorbable materials such as cellulose or wood.

- Intended user: trained nurse level or higher level of healthcare providers

The validated maximum load conditions

Total weight:

The validation performances were conducted using a validation load consisting of two instrument trays containing a total weight of 8.9kg(19.6lbs).

Usage of sterilization pouch:

The medical devices to be sterilized are packaged in Tyvek sterilization pouch and sealed. The pouched medical devices are placed at the trays.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231893: LOWTEM Crystal 120.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231893

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026