510(k) Report

510(k) · Cleared

Cleared

QuantX

K170195510(k) TraditionalCleared May 17, 2017114-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Quantitative Insights, Inc. · LLZ System, Image Processing, Radiological

Cleared on May 17, 2017 as substantially equivalent to K081556 (Confirma, Inc., 2009) and 1 other predicate after a 114-day review. Its family tree reaches back 11 generations to 1991. 4 later devices descend from it.

Decided
May 17, 2017
Review
114 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
2

Its longest line of descent runs back 11 generations to K913770 (1991). 75 ancestors in all, the oldest K896095 (1990). 4 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2006: 1 other ancestor2005: 2 other ancestors2004: 5 other ancestors2003: 7 other ancestors1996: 5 other ancestors2002: 6 other ancestors2001: 11 other ancestors2000: 4 other ancestors1999: 2 other ancestors1998: 5 other ancestors1991: 3 other ancestors1997: 2 other ancestors1995: 3 other ancestors1994: 5 other ancestors1992: 1 other ancestor1990: 1 other ancestor2021: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants clearedK913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK023003 · Imagechecker-CT Workstation · R2 Technology, Inc. · 2002×2: 2 generations cleared close together, K032748 (2003), K040305 (2004) K043333 · VITREA2, Version 3.7 Medical Image Processing System · Vital Images, Inc. · 2004 · recalledK052963 · Mediq Version 1.0 · Clario Medical Imaging, Inc. · 2005K080196 · Z3D Contrast Acuity · Clario Medical Imaging, Inc. · 2008K081556 · Cadstream, Version 5.0 · Confirma, Inc. · 2009K092954 · Cadstream Version 5 · Confirma, Inc. · 2009 · recalled1990200020102020
  1. Root · 11 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 7 generations between
  3. 3 generations backK080196 Z3D Contrast AcuityClario Medical Imaging, Inc. · 2008
  4. 2 generations backK081556 Cadstream, Version 5.0Confirma, Inc. · 2009
  5. 1 generation backK092954 Cadstream Version 5Confirma, Inc. · 2009 · Recalled
  6. This deviceK170195 QuantXQuantitative Insights, Inc. · 2017
  7. Descendant · cites it directly · 3 of its ownK210670 BU-CADTaihao Medical, Inc. · 2021
  8. Descendant · 2 generations on · 0 of its ownK260086 CadAI-B DxBeamWorks, Inc. · 2026
  9. Descendant · 2 generations on · 0 of its ownK260303 See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd. · 2026
  10. Descendant · 2 generations on · 0 of its ownK240697 See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd. · 2024
Show all 11 generations
  1. Root · 11 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 10 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 9 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 8 generations backK023003 Imagechecker-CT WorkstationR2 Technology, Inc. · 2002
  5. 7 generations backK032748 VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc. · 2003 · Recalled
  6. 6 generations backK040305 VITREA2 Version 3.5Vital Images, Inc. · 2004
  7. 5 generations backK043333 VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc. · 2004 · Recalled
  8. 4 generations backK052963 Mediq Version 1.0Clario Medical Imaging, Inc. · 2005
  9. 3 generations backK080196 Z3D Contrast AcuityClario Medical Imaging, Inc. · 2008
  10. 2 generations backK081556 Cadstream, Version 5.0Confirma, Inc. · 2009
  11. 1 generation backK092954 Cadstream Version 5Confirma, Inc. · 2009 · Recalled
  12. This deviceK170195 QuantXQuantitative Insights, Inc. · 2017

Most influential descendants

Showing 4 of 4 · See all
  • K210670BU-CADTaihao Medical, Inc. · 2021 · cites it directly3descendants
  • K260086CadAI-B DxBeamWorks, Inc. · 2026 · 2 generations on0descendants
  • K260303See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd. · 2026 · 2 generations on0descendants
  • K240697See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−10K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−9K020483Advantage Workstation 4.1GE Medical Systems20022
−8K023003Imagechecker-CT WorkstationR2 Technology, Inc.2002—
−7K032748VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc.20032
−6K040305VITREA2 Version 3.5Vital Images, Inc.2004—
−5K043333VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc.20041
−4K052963Mediq Version 1.0Clario Medical Imaging, Inc.2005—
−3K080196Z3D Contrast AcuityClario Medical Imaging, Inc.2008—
−2K081556Cadstream, Version 5.0Confirma, Inc.2009—
−1K092954Cadstream Version 5Confirma, Inc.20093
0 (this device)K170195QuantXQuantitative Insights, Inc.2017—
+1K210670BU-CADTaihao Medical, Inc.2021—
+2K260086CadAI-B DxBeamWorks, Inc.2026—
+2K260303See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd.2026—
+2K240697See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199130
199210
199450
199530
199650
199720
199850
199920
200040
2001110
200260
200370
200450
200520
200610
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202410
202620
FIG. 1 — Lineage of K170195, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081556Cadstream, Version 5.0
Summary page 4
…ame: Picture archiving and communications system Product Code: LLZ A. LEGALLY MARKETED PREDICATE DEVICE Legally marketed predicate device is: CADStream Version 5 - K081556, K092954 B. DEVICE DESCRIPTION QuantX is a software program that analyzes patient breast images, and is designed to aid rad…
Confirma, Inc.2009—
*K092954Cadstream Version 5
Summary page 4
…ure archiving and communications system Product Code: LLZ A. LEGALLY MARKETED PREDICATE DEVICE Legally marketed predicate device is: CADStream Version 5 - K081556, K092954 B. DEVICE DESCRIPTION QuantX is a software program that analyzes patient breast images, and is designed to aid radiologists…
Confirma, Inc.20093

Also cleared against K081556: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210670BU-CADTaihao Medical, Inc.20213

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 7 companies; a new generation was cleared every 11 months on median.
  • 18 of the 75 ancestors have a recall of any class.
  • Of its descendants, K210670 (Taihao Medical, Inc., 2021) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K081556.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on…

Read the full statement (151 words)

QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor.

When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis.

QuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report.

Limitations: QuantX is not intended for primary interpretation of digital mammography images.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K170195: QuantX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K170195

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.