510(k) Report

510(k) · Cleared

Cleared

BU-CAD

K210670510(k) TraditionalCleared Dec 21, 2021291-day reviewClass II21 CFR 892.2090

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Taihao Medical, Inc. · QDQ Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer

Cleared on Dec 21, 2021 as substantially equivalent to K181704 (Screenpoint Medical B.V., 2018) and 1 other predicate after a 291-day review. Its family tree reaches back 12 generations to 1991. 3 later devices descend from it.

Decided
Dec 21, 2021
Review
291 days
Class
Class II
Product code
QDQ
Recalls
None
Predicates
2

Its longest line of descent runs back 12 generations to K913770 (1991). 78 ancestors in all, the oldest K896095 (1990). 3 later devices descend from it, 3 citing it directly.

  1. Root · 12 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 8 generations between
  3. 3 generations backK081556 Cadstream, Version 5.0Confirma, Inc. · 2009
  4. 2 generations backK092954 Cadstream Version 5Confirma, Inc. · 2009 · Recalled
  5. 1 generation backK170195 QuantXQuantitative Insights, Inc. · 2017
  6. This deviceK210670 BU-CADTaihao Medical, Inc. · 2021
  7. Descendant · cites it directly · 0 of its ownK260086 CadAI-B DxBeamWorks, Inc. · 2026
  8. Descendant · cites it directly · 0 of its ownK260303 See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd. · 2026
  9. Descendant · cites it directly · 0 of its ownK240697 See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd. · 2024
Show all 12 generations
  1. Root · 12 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 11 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 10 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 9 generations backK023003 Imagechecker-CT WorkstationR2 Technology, Inc. · 2002
  5. 8 generations backK032748 VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc. · 2003 · Recalled
  6. 7 generations backK040305 VITREA2 Version 3.5Vital Images, Inc. · 2004
  7. 6 generations backK043333 VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc. · 2004 · Recalled
  8. 5 generations backK052963 Mediq Version 1.0Clario Medical Imaging, Inc. · 2005
  9. 4 generations backK080196 Z3D Contrast AcuityClario Medical Imaging, Inc. · 2008
  10. 3 generations backK081556 Cadstream, Version 5.0Confirma, Inc. · 2009
  11. 2 generations backK092954 Cadstream Version 5Confirma, Inc. · 2009 · Recalled
  12. 1 generation backK170195 QuantXQuantitative Insights, Inc. · 2017
  13. This deviceK210670 BU-CADTaihao Medical, Inc. · 2021

Most influential descendants

Showing 3 of 3 · See all
  • K260086CadAI-B DxBeamWorks, Inc. · 2026 · cites it directly0descendants
  • K260303See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd. · 2026 · cites it directly0descendants
  • K240697See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−11K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−10K020483Advantage Workstation 4.1GE Medical Systems20022
−9K023003Imagechecker-CT WorkstationR2 Technology, Inc.2002—
−8K032748VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc.20032
−7K040305VITREA2 Version 3.5Vital Images, Inc.2004—
−6K043333VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc.20041
−5K052963Mediq Version 1.0Clario Medical Imaging, Inc.2005—
−4K080196Z3D Contrast AcuityClario Medical Imaging, Inc.2008—
−3K081556Cadstream, Version 5.0Confirma, Inc.2009—
−2K092954Cadstream Version 5Confirma, Inc.20093
−1K170195QuantXQuantitative Insights, Inc.2017—
0 (this device)K210670BU-CADTaihao Medical, Inc.2021—
+1K260086CadAI-B DxBeamWorks, Inc.2026—
+1K260303See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd.2026—
+1K240697See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199130
199210
199450
199530
199650
199720
199850
199920
200040
2001110
200260
200370
200450
200520
200610
200810
201820

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202620
FIG. 1 — Lineage of K210670, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QDQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QDQ device

  1. #1 K241831 Transpara (2.1.0)score 15.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K181704Transpara
Summary page 4
…or Cancer Review Panel: Radiology Manufacturer: TaiHao Medical Inc. III. Predicate Device Predicate Device: TransparaTM (K181704) (primary), QuantX (K170195) Reference Device: Koios DS for Breast (K190442)…
Screenpoint Medical B.V.2018—
*K170195QuantX
Summary page 4
…Radiology Manufacturer: TaiHao Medical Inc. III. Predicate Device Predicate Device: TransparaTM (K181704) (primary), QuantX (K170195) Reference Device: Koios DS for Breast (K190442) 1…
Quantitative Insights, Inc.2017—
†K190442Koios DS for BreastKoios Medical, Inc.2019—

Also cleared against K181704: 5 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K260086CadAI-B DxBeamWorks, Inc.20260
K260303See-Mode Augmented Reporting Tool, Breast (SMART-B)See-Mode Technologies Pty, Ltd.20260
K240697See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd.20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 30 years and passes through 8 companies; a new generation was cleared every 12 months on median.
  • 18 of the 78 ancestors have a recall of any class.
  • Of its descendants, K260086 (BeamWorks, Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K181704.

How does its category compare?

All QDQ devices

QDQ Radiological Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · QDQ

10so far in 2026, the most since 2007

2018: 22019: 22020: 32021: 52022: 42023: 22024: 72025: 52026 (partial year): 10
2007high 10 (2026) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QDQ

216in 2025

2020: 173 days median2021: 169 days median2022: 131 days median2024: 153 days median2025: 216 days median2026 (partial year): 154 days median
2007high 216 (2025) · low 131 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QDQ

No recall names QDQ since 2007.

Adverse event reports per year · QDQ

No MAUDE report names QDQ since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20182—00
20192—00
2020317300
2021516900
2022413100
20232—00
2024715300
2025521600
2026101540—

What is it indicated for?

BU-CAD is a software application indicated to assist trained interpreting physicians in analyzing the breast ultrasound images of patients with soft tissue breast lesions suspicious for breast cancer who are being referred for further diagnostic ultrasound examination. Output of the device includes regions of interest (ROIs) and lesion contours placed on breast ultrasound images assisting physicians to identify suspicious soft tissue lesions from up to…

Read the full statement (221 words)

BU-CAD is a software application indicated to assist trained interpreting physicians in analyzing the breast ultrasound images of patients with soft tissue breast lesions suspicious for breast cancer who are being referred for further diagnostic ultrasound examination.

Output of the device includes regions of interest (ROIs) and lesion contours placed on breast ultrasound images assisting physicians to identify suspicious soft tissue lesions from up to two orthogonal views of a single lesion, and region-based analysis of lesion malignancy upon the physician's query. The region-based analysis indicates the score of lesion characteristics (SLC), and corresponding BI-RADS categories in user-selected ROIs or ROIs automatically identified by the software. In addition, BU-CAD also automatically classifies lesion shape, orientation, margin, echo pattern, and posterior features according to BI-RADS descriptors.

BU-CAD may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software includes tools that allow users to adjust, measure and document images, and output into a structured report (SR).

Patient management decisions should not be made solely on the basis of analysis by BU-CAD.

Limitations: BU-CAD is not to be used on sites of post-surgical excision, or images with Doppler, elastography, or other overlays present in them. BU-CAD is not intended for the primary interpretation of digital mammography images. BU-CAD is not intended for use on mobile devices.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K210670: BU-CAD.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K210670

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026