510(k) Report

510(k) · Cleared

Cleared

Anprolene AN75 Ethylene Oxide Gas Sterilizer

K170432510(k) TraditionalCleared Oct 31, 2017260-day reviewClass II21 CFR 880.6860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Andersen Sterilizers, Inc. · FLF Sterilizer, Ethylene-Oxide Gas

Cleared on Oct 31, 2017 as substantially equivalent to K150646 (Andersen Sterilizers, Inc., 2015) after a 260-day review. Its family tree reaches back 2 generations to 1981.

Decided
Oct 31, 2017
Review
260 days
Class
Class II
Product code
FLF
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K812867 (1981). No later device cites it as a predicate yet.

  1. Root · 2 generations backK812867 Steri-Vac #400C Gas Sterilizer3M Company · 1981 · Recalled
  2. 1 generation backK150646 EOGas 4 Ethylene Oxide Gas SterilizerAndersen Sterilizers, Inc. · 2015
  3. This deviceK170432 Anprolene AN75 Ethylene Oxide Gas SterilizerAndersen Sterilizers, Inc. · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K812867Steri-Vac #400C Gas Sterilizer3M Company19811
−1K150646EOGas 4 Ethylene Oxide Gas SterilizerAndersen Sterilizers, Inc.2015—
0 (this device)K170432Anprolene AN75 Ethylene Oxide Gas SterilizerAndersen Sterilizers, Inc.2017—
FIG. 1 — Lineage of K170432, 1981–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLF device

  1. #1 K192978 EOGas 4 Ethylene Oxide Gas Sterilizerscore 0.6 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150646EOGas 4 Ethylene Oxide Gas Sterilizer
Summary page 5
…69340-9). Primary Predicate Device - EOGas 4 Sterilizer Device Name EOGas 4® Ethylene Oxide Gas Sterilizer 510(k) number K150646 Manufacturer Andersen Sterilizers, Inc. For the subject device, the sterilization temperature, steriliza…
Andersen Sterilizers, Inc.2015—
*—B-2270/AN70 Anprolene Sterilizer
Named without a submission number
—

Also cleared against K150646: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 36 years and passes through 2 companies; a new generation was cleared every 34.0 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 1 other device also cites its predicate K150646.

How does its category compare?

All FLF devices

FLF Sterilizer, Ethylene-Oxide Gas: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · FLF

1in 2020

2009: 12015: 22017: 12020: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLF

Too few FLF 510(k)s in any year since 2007 to give a median.

Recalls per year · FLF

2in 2014

2009: 12012: 12013: 22014: 2
2007high 2 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLF

4in 2021

2007: 22008: 12009: 22010: 32011: 42012: 52013: 32014: 62015: 32016: 32017: 32019: 22021: 4
2007high 6 (2014) · low 0 (2018)2025

MAUDE reports naming product code FLF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20080—01
20091—12
20100—03
20110—04
20120—15
20130—23
20140—26
20152—03
20160—03
20171—03
20180—00
20190—02
20201—00
20210—04
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Anprolene AN75 Ethylene Oxide Gas Sterilizer is intended for use in hospitals and other human healthcare settings. It is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation. The 12 hour cycle at 20-29ºC in the Anprolene AN75 Ethylene Oxide Gas Sterilizer is for surface sterilization of medical devices, including mated surfaces. The critical process parameters for the cycle are…

Read the full statement (284 words)

The Anprolene AN75 Ethylene Oxide Gas Sterilizer is intended for use in hospitals and other human healthcare settings. It is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation.

The 12 hour cycle at 20-29ºC in the Anprolene AN75 Ethylene Oxide Gas Sterilizer is for surface sterilization of medical devices, including mated surfaces. The critical process parameters for the cycle are summarized in Table 1.

Table 1. Critical sterilization cycle parameters in the Anprolene AN75 Ethylene Oxide Gas Sterilizer

EO Amount: 17.6 g ± 5%; Temperature: 20-29°C (68-84ºF); Relative Humidity: 35-70%; EO Exposure Time: 12 hours; Total Cycle Time: 14 hours

Maximum loads of specific materials and devices that have been validated are listed in Table 2. All validated maximum loads were processed without additional devices in the sterilizer.

Table 2. Load and material types validated in the Anprolene AN75 Ethylene Oxide Gas Sterilizer

Device Type: Metal. Maximum Load: 24 lbs (11 kg). Device Examples: Delicate sharps and surgical instruments, including those with hinges and mated surfaces. Required Aeration: Metal instruments do not absorb EO; Follow pouch or wrap manufacturer’s instructions (Example: Sterisheet® wraps require ≥ 6 hours at 20-29°C).

Device Type: Plastic. Maximum Load: 3.5 lbs (1.6 kg). Device Examples: Reusable power cords, trocars. Required Aeration: Follow manufacturer’s instructions; 24 hours at 20-29°C.

Device Type: Fabric. Maximum Load: 3.0 lbs (1.4 kg). Device Examples: Reusable cloth gowns, towels. Required Aeration: Follow manufacturer’s instructions; 24 hours at 20-29°C.

Reusable medical devices must be aerated following the instructions of device manufacturers and packaging material manufacturers, then released for use after sterilization based on successful inactivation of an EZTest®- Gas Biological Indicator in the Anprolene SteriTest process challenge device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170432: Anprolene AN75 Ethylene Oxide Gas Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170432

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026