510(k) · Cleared
Recalled · Class II · 2019Cranial Drill Bits and Accessories
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ad-Tech Medical Instrument Corporation · HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Cleared on Sep 7, 2017 as substantially equivalent to K062842 (Walter Lorenz Surgical, Inc., 2006) and 2 other predicates after a 205-day review. Its family tree reaches back 4 generations to 1996. It has been recalled 2 times.
- Decided
- Sep 7, 2017
- Review
- 205 days
- Class
- Class II
- Product code
- HBE
- Recalls
- 2
- Predicates
- 3
Its longest line of descent runs back 4 generations to K961113 (1996). 9 ancestors in all, the oldest K832187 (1983). No later device cites it as a predicate yet.
- Root · 4 generations backK961113 CNS Interchangeable Bit Cranial DrillClinical Neuro Systems, LLC · 1996 · Recalled
- 3 generations backK122456 Mrii Crainial DrillMri Interventions, Inc. · 2013
- 2 generations backK151536 SmartTwist MRII Hand Drill, SmartTip MRII Drill bit Kit, 4.5mm, 6.0mmMri Interventions, Inc. · 2015
- 1 generation backK160129 MRII Cranial Drill and AccessoriesMri Interventions, Inc. · 2016
- This deviceK170442 Cranial Drill Bits and AccessoriesAd-Tech Medical Instrument Corporation · 2017
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K961113 | CNS Interchangeable Bit Cranial Drill | Clinical Neuro Systems, LLC | 1996 | 39 |
| −3 | K122456 | Mrii Crainial Drill | Mri Interventions, Inc. | 2013 | — |
| −2 | K151536 | SmartTwist MRII Hand Drill, SmartTip MRII Drill bit Kit, 4.5mm, 6.0mm | Mri Interventions, Inc. | 2015 | — |
| −1 | K160129 | MRII Cranial Drill and Accessories | Mri Interventions, Inc. | 2016 | — |
| 0 (this device) | K170442 | Cranial Drill Bits and Accessories | Ad-Tech Medical Instrument Corporation | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1983 | 1 | 0 |
| 1996 | 1 | 0 |
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HBE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HBE device
- #1 K251433 Surgify Halo (54.085.SHD.H1); Surgify Halo…score 13.1 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K062842 | Twist DrillSummary page 4…882.4300 Manual cranial drills, burrs, trephines, and their accessories, Product Code: HBG, Class II Predicate Devices: K062842 Twist Drills (Bits), Primary Predicate K122456 MRI Cranial Drill… | 2006 |
| * | K122456 | Mrii Crainial DrillSummary page 4…accessories, Product Code: HBG, Class II Predicate Devices: K062842 Twist Drills (Bits), Primary Predicate K122456 MRI Cranial Drill K160129 MRII Cranial Drill and Accessories 5.1 Product Description… | 2013 |
| * | K160129 | MRII Cranial Drill and AccessoriesSummary page 4…Predicate Devices: K062842 Twist Drills (Bits), Primary Predicate K122456 MRI Cranial Drill K160129 MRII Cranial Drill and Accessories 5.1 Product Description The device under review is a family of cranial drill bits an… | 2016 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 27, 2019 | Class II | Z-2544-2019 | Under Investigation by firm | The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery. |
| May 2, 2018 | Class II | Z-2539-2018 | Nonconforming Material/Component | The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
- 3 of the 9 ancestors have a recall of any class.
- 3 other devices also cite its predicate K062842.
How does its category compare?
All HBE devicesHBE Drills, Burrs, Trephines & Accessories (Simple, Powered): FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · HBE
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HBE
30in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HBE
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HBE
1,924in 2025
MAUDE reports naming product code HBE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 14 |
| 2008 | 2 | — | 47 | 14 |
| 2009 | 0 | — | 7 | 242 |
| 2010 | 0 | — | 23 | 492 |
| 2011 | 2 | — | 1 | 1,013 |
| 2012 | 1 | — | 26 | 1,146 |
| 2013 | 1 | — | 4 | 1,583 |
| 2014 | 2 | — | 0 | 559 |
| 2015 | 3 | 155 | 0 | 483 |
| 2016 | 2 | — | 7 | 848 |
| 2017 | 2 | — | 3 | 848 |
| 2018 | 2 | — | 0 | 787 |
| 2019 | 5 | 133 | 0 | 836 |
| 2020 | 0 | — | 0 | 1,025 |
| 2021 | 6 | 193 | 1 | 1,047 |
| 2022 | 2 | — | 0 | 900 |
| 2023 | 2 | — | 3 | 948 |
| 2024 | 0 | — | 1 | 1,153 |
| 2025 | 6 | 30 | 0 | 1,924 |
| 2026 | 1 | — | 19 | — |
What is it indicated for?
The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K170442 · printed Nov 6, 2026
Cite this page
510(k) Report. “K170442: Cranial Drill Bits and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170442
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026