510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Accutorr 7/VS-900/VS-900c Vital Signs Monitor

K182821510(k) SpecialCleared Mar 27, 2019174-day reviewClass II21 CFR 870.2300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Cleared on Mar 27, 2019 as substantially equivalent to K170712 (Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd., 2017) and 1 other predicate after a 174-day review. Its family tree reaches back 11 generations to 1994. It has been recalled once.

Decided
Mar 27, 2019
Review
174 days
Class
Class II
Product code
MWI
Recalls
1
Predicates
2

Its longest line of descent runs back 11 generations to K933143 (1994). 48 ancestors in all, the oldest K842616 (1984). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2001: 1 other ancestor2006: 2 other ancestors2000: 2 other ancestors1997: 4 other ancestors1996: 4 other ancestors1995: 5 other ancestors1994: 4 other ancestors1992: 1 other ancestor1991: 5 other ancestors1990: 3 other ancestors1989: 2 other ancestors1999: 1 other ancestor1998: 1 other ancestor1993: 1 other ancestor1984: 1 other ancestorK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK993531 · Passport 2 Vital Signs Monitor, Model 0998-00-0170-XX · Datascope Corp. · 2000 · recalledK032733 · PM-8000 Patient Monitor, Model 8000 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004K043348 · Modification to PM-8000 Patient Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005K060281 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K063055 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K072235 · PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · recalledK132038 · Accutorr 7 Vital Signs Monitor · Mindray North America · 2014K170712 · Accutorr 7/VS-900 Vital Signs Monitor · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · recalled19801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor4 generations back: 1 other ancestor5 generations back: 2 other ancestors6 generations back: 4 other ancestors7 generations back: 10 other ancestors8 generations back: 6 other ancestors9 generations back: 9 other ancestors10 generations back: 4 other ancestors K170712 · Accutorr 7/VS-900 Vital Signs Monitor · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · recalledK132038 · Accutorr 7 Vital Signs Monitor · Mindray North America · 2014K072235 · PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · recalledK063055 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K060281 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K043348 · Modification to PM-8000 Patient Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005K032733 · PM-8000 Patient Monitor, Model 8000 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004K993531 · Passport 2 Vital Signs Monitor, Model 0998-00-0170-XX · Datascope Corp. · 2000 · recalledK972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994−10−50
  1. Root · 11 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 7 generations between
  3. 3 generations backK072235 PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · Recalled
  4. 2 generations backK132038 Accutorr 7 Vital Signs MonitorMindray North America · 2014
  5. 1 generation backK170712 Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · Recalled
  6. This deviceK182821 Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019
Show all 11 generations
  1. Root · 11 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 10 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 9 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 8 generations backK993531 Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp. · 2000 · Recalled
  5. 7 generations backK032733 PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004
  6. 6 generations backK043348 Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005
  7. 5 generations backK060281 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  8. 4 generations backK063055 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  9. 3 generations backK072235 PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · Recalled
  10. 2 generations backK132038 Accutorr 7 Vital Signs MonitorMindray North America · 2014
  11. 1 generation backK170712 Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · Recalled
  12. This deviceK182821 Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−10K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−9K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−8K993531Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp.20001
−7K032733PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2004—
−6K043348Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2005—
−5K060281VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
−4K063055VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
−3K072235PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.20072
−2K132038Accutorr 7 Vital Signs MonitorMindray North America2014—
−1K170712Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd.20171
0 (this device)K182821Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198920
199030
199150
199210
199310
199440
199550
199640
199740
199810
199910
200020
200110
200620
FIG. 1 — Lineage of K182821, 1994–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MWI device

  1. #1 K230457 Empatica Health Monitoring Platform…score 28.0 · 39 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170712Accutorr 7/VS-900 Vital Signs Monitor
Summary page 5
…Clinical electronic thermometer, electronic, 880.2910 Hospital thermometer clinical Primary Predicate Device: K170712 - Accutorr 7/VS-900 Vital Signs Monitor; SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD Secondary Predicate Device: K011291 -…
Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd.20171
*K011291Temporalscanner Thermometer, Sensortouch
Summary page 5
…clinical Primary Predicate Device: K170712 - Accutorr 7/VS-900 Vital Signs Monitor; SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD Secondary Predicate Device: K011291 - TEMPORALSCANNER THERMOMETER, SENSORTOUCH; EXERGEN CORP. Indications for Use: The Accutorr 7/VS-900/VS-900c Vital Signs Mon…
Exergen Corp.2001—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 1, 2019Class IIZ-0236-2020No Marketing ApplicationNIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 1.6 years on median.
  • 8 of the 48 ancestors have a recall of any class.

How does its category compare?

All MWI devices

MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · MWI

4in 2025

2007: 92008: 92009: 62010: 172011: 112012: 132013: 202014: 162015: 192016: 122017: 132018: 112019: 142020: 82021: 92022: 52023: 82024: 102025: 42026 (partial year): 6
2007high 20 (2013) · low 4 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MWI

247in 2025

2007: 42 days median2008: 49 days median2009: 112 days median2010: 58 days median2011: 78 days median2012: 104 days median2013: 85 days median2014: 148 days median2015: 127 days median2016: 165 days median2017: 205 days median2018: 217 days median2019: 255 days median2020: 138 days median2021: 161 days median2022: 231 days median2023: 250 days median2024: 202 days median2025: 247 days median2026 (partial year): 239 days median
2007high 255 (2019) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MWI

9so far in 2026, the most since 2007

2007: 12008: 12009: 22010: 22011: 32012: 52013: 32014: 62015: 12017: 22019: 42022: 22024: 52026 (partial year): 9
2007high 9 (2026) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MWI

182in 2025

2007: 32008: 32009: 202010: 172011: 162012: 312013: 422014: 302015: 312016: 512017: 852018: 1582019: 1692020: 1102021: 1882022: 1682023: 1792024: 1202025: 182
2007high 188 (2021) · low 3 (2007)2025

MAUDE reports naming product code MWI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200794213
200894913
20096112220
20101758217
20111178316
201213104531
20132085342
201416148630
201519127131
201612165051
201713205285
2018112170158
2019142554169
202081380110
202191610188
202252312168
202382500179
2024102025120
202542470182
202662399—

What is it indicated for?

The Accutorr 7/VS-900/VS-900c Vital Signs Monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), Temperature (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.

• The monitor also provides a Modified Early Warning Score (MEWS) for clinical assessment in adult patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182821: Accutorr 7/VS-900/VS-900c Vital Signs Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182821

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.