OEC Elite

K170752510(k) TraditionalCleared May 19, 201767-day reviewClass II21 CFR 892.1650Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Oec Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on May 19, 2017 as substantially equivalent to K122234 (Ge Healthcare Surgery GE Oec Medical Systems, Inc., 2012) after a 67-day review. Its family tree reaches back 6 generations to 2002. 2 of the 15 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled 4 times.

Decided
May 19, 2017
Review
67 days
Class
Class II
Product code
OWB
Recalls
4
Predicates
1

Its longest line of descent runs back 6 generations to K021049 (2002). 7 ancestors in all. 7 later devices descend from it, 1 citing it directly. 2 of the 15 devices in this family, counting this one, have a Class I recall.

  1. Root · 6 generations backK021049 Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · Recalled (Class I)
  2. 5 generations backK041932 Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · Recalled
  3. 4 generations backK073543 Oec 9900 EliteGe Oec Medical Systems · 2008 · Recalled
  4. 3 generations backK082781 Oec 9900 EliteGe Healthcare Surgery · 2009 · Recalled
  5. 2 generations backK120613 Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · Recalled
  6. 1 generation backK122234 Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · Recalled
  7. This deviceK170752 OEC EliteOec Medical Systems, Inc. · 2017
  8. Descendant · cites it directly · 6 of its ownK171565 OEC EliteGe Oec Medical Systems, Inc. · 2017 · Recalled
  9. Descendant · 2 generations on · 5 of its ownK172550 OEC EliteGe Oec Medical Systems, Inc. · 2017 · Recalled
  10. Descendant · 3 generations on · 2 of its ownK192819 OEC EliteGe Oec Medical Systems, Inc. · 2019 · Recalled
  11. Descendant · 4 generations on · 1 of its ownK203346 OEC 3DGe Oec Medical Systems, Inc. · 2021 · Recalled
  12. Descendant · 3 generations on · 1 of its ownK200218 Digiscan FDXAllengers Medical Systems Limited · 2020

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2017: 22019: 12020: 12021: 12024: 12026: 1
2017high 2 (2017) · low 0 (2018)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K171565OEC EliteGe Oec Medical Systems, Inc. · 2017 · cites it directly6descendants
  • K172550OEC EliteGe Oec Medical Systems, Inc. · 2017 · 2 generations on5descendants
  • K192819OEC EliteGe Oec Medical Systems, Inc. · 2019 · 3 generations on2descendants
  • K203346OEC 3DGe Oec Medical Systems, Inc. · 2021 · 4 generations on1descendant
  • K200218Digiscan FDXAllengers Medical Systems Limited · 2020 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems2002Class I
−5K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.20047
−4K073543Oec 9900 EliteGe Oec Medical Systems20085
−3K082781Oec 9900 EliteGe Healthcare Surgery20094
−2K120613Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc.20127
−1K122234Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc.20126
0 (this device)K170752OEC EliteOec Medical Systems, Inc.2017—
+1K171565OEC EliteGe Oec Medical Systems, Inc.20174
+2K172550OEC EliteGe Oec Medical Systems, Inc.20174
+3K192819OEC EliteGe Oec Medical Systems, Inc.20194
+4K203346OEC 3DGe Oec Medical Systems, Inc.20213
+3K200218Digiscan FDXAllengers Medical Systems Limited2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201720
201910
202010
202110
202410
202610
FIG. 1 — Lineage of K170752, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122234Oec 9900 Elite
Summary page 4
…Salt Lake City, Utah 84116 Predicate Device: Device Name: OEC 9900 Elite 510(k) number: K122234 Manufacturer: GE OEC Medical Systems, Inc (same as proposed device) Regulation Name:…
Ge Healthcare Surgery GE Oec Medical Systems, Inc.20126
†K132904Ziehm Vision RFDZiehm Imaging GmbH20131
†K161976Ziehm Solo FDZiehm Imaging GmbH20161

Also cleared against K122234: 1 other device →

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171565OEC EliteGe Oec Medical Systems, Inc.20176

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 21, 2025Class IIZ-1819-2025Process controlOEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Nov 21, 2024Class IIZ-0704-2025Process controlDue to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
May 15, 2023Class IIZ-2278-2023Process controlImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries…
Jun 10, 2021Class IIZ-2161-2021Under Investigation by firmThere is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 5 companies; a new generation was cleared every 2.9 years on median.
  • 7 of the 7 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K171565 (Ge Oec Medical Systems, Inc., 2017) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K122234.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, critical care and emergency procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170752: OEC Elite.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170752

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026