510(k) Report

510(k) · Cleared

Cleared

Cochlear Baha SoundArc

K171088510(k) SpecialCleared Jun 7, 201756-day reviewClass II21 CFR 874.3302

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cochlear Americas · LXB Hearing Aid, Bone Conduction

Cleared on Jun 7, 2017 as substantially equivalent to K002913 (Entific Medical Systems, Inc., 2000) and 7 other predicates after a 56-day review. Its family tree reaches back 9 generations to 1986.

Decided
Jun 7, 2017
Review
56 days
Class
Class II
Product code
LXB
Recalls
None
Predicates
8

Its longest line of descent runs back 9 generations to K861971 (1986). 25 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 9 generations backK861971 Xomed Audiant Bone Conductor W/Improved AA CoilXomed, Inc. · 1986
  2. 5 generations between
  3. 3 generations backK132278 Cochlear Baha 4 Sound ProcessorCochlear Americas · 2013
  4. 2 generations backK142907 Baha 5 Sound ProcessorCochlear Americas · 2015 · Recalled
  5. 1 generation backK153245 Baha 5 Super Power Sound ProcessorCochlear Americas · 2016
  6. This deviceK171088 Cochlear Baha SoundArcCochlear Americas · 2017
Show all 9 generations
  1. Root · 9 generations backK861971 Xomed Audiant Bone Conductor W/Improved AA CoilXomed, Inc. · 1986
  2. 8 generations backK955713 Nobelpharma Bone Anchored Hearing AidNobelpharma USA, Inc. · 1996
  3. 7 generations backK984162 Branemark Bone-Anchored Hearing Aid (Baha) System for Pediatric UseNobelpharma USA, Inc. · 1999
  4. 6 generations backK992872 Branemark Bone-Anchored Hearing Aid (Baha) Cordelle IIEntific Medical Systems, Inc. · 1999
  5. 5 generations backK080363 Baha Cordelle IICochlear Bone Anchored Systems AB · 2008
  6. 4 generations backK090720 Cochlear Baha BP100Cochlear Bone Anchored Systems AB · 2009
  7. 3 generations backK132278 Cochlear Baha 4 Sound ProcessorCochlear Americas · 2013
  8. 2 generations backK142907 Baha 5 Sound ProcessorCochlear Americas · 2015 · Recalled
  9. 1 generation backK153245 Baha 5 Super Power Sound ProcessorCochlear Americas · 2016
  10. This deviceK171088 Cochlear Baha SoundArcCochlear Americas · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K861971Xomed Audiant Bone Conductor W/Improved AA CoilXomed, Inc.1986—
−8K955713Nobelpharma Bone Anchored Hearing AidNobelpharma USA, Inc.1996—
−7K984162Branemark Bone-Anchored Hearing Aid (Baha) System for Pediatric UseNobelpharma USA, Inc.1999—
−6K992872Branemark Bone-Anchored Hearing Aid (Baha) Cordelle IIEntific Medical Systems, Inc.1999—
−5K080363Baha Cordelle IICochlear Bone Anchored Systems AB2008—
−4K090720Cochlear Baha BP100Cochlear Bone Anchored Systems AB2009—
−3K132278Cochlear Baha 4 Sound ProcessorCochlear Americas2013—
−2K142907Baha 5 Sound ProcessorCochlear Americas20151
−1K153245Baha 5 Super Power Sound ProcessorCochlear Americas2016—
0 (this device)K171088Cochlear Baha SoundArcCochlear Americas2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199110
199310
199410
199510
199810
199910
200010
200110
200210
200410
200810
201110
201310
201510
201610
FIG. 1 — Lineage of K171088, 1986–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LXB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LXB device

  1. #1 K254296 Ossiclear (Ossiclear System)score 14.7 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (8)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002913Headband for Baha
Summary page 4
…R §874.3300 Product Codes: LXB Panel: Ear Nose and Throat Devices Panel Primary System Predicate Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K…
Entific Medical Systems, Inc.2000—
*K080363Baha Cordelle II
Summary page 4
…Ear Nose and Throat Devices Panel Primary System Predicate Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Subm…
Cochlear Bone Anchored Systems AB2008—
*K090720Cochlear Baha BP100
Summary page 4
…se and Throat Devices Panel Primary System Predicate Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission Th…
Cochlear Bone Anchored Systems AB2009—
*K110996Baha 3 Power Sound Processor
Summary page 4
…m Predicate Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission The purpose of this Special 510(k) is to in…
Cochlear Americas2011—
*K132278Cochlear Baha 4 Sound Processor
Summary page 4
…te Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission The purpose of this Special 510(k) is to introduce a…
Cochlear Americas2013—
*K142907Baha 5 Sound Processor
Summary page 4
…Baha headband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission The purpose of this Special 510(k) is to introduce a new styl…
Cochlear Americas20151
*K153245Baha 5 Super Power Sound Processor
Summary page 4
…dband/Softband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission The purpose of this Special 510(k) is to introduce a new style of head…
Cochlear Americas2016—
*K161123Baha 5 Power Sound Processor
Summary page 4
…tband, K002913 Secondary System Predicates: Sound Processors of the Baha system: K080363, K090720, K110996, K132278, K142907, K153245, K161123 K171088 F. Purpose of Submission The purpose of this Special 510(k) is to introduce a new style of headband, kno…
Cochlear Americas2016—

Also cleared against K002913: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 1.5 years on median.
  • 2 of the 25 ancestors have a recall of any class.
  • 1 other device also cites its predicate K002913.

How does its category compare?

All LXB devices

LXB Hearing Aid, Bone Conduction: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LXB

2in 2025

2008: 22009: 22010: 22011: 52012: 12013: 52015: 22016: 42017: 12018: 12019: 12021: 42025: 22026 (partial year): 4
2007high 5 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LXB

79in 2021

2011: 189 days median2013: 103 days median2016: 110 days median2021: 79 days median2026 (partial year): 164 days median
2007high 189 (2011) · low 79 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LXB

1in 2018

2012: 52018: 1
2007high 5 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LXB

261in 2025

2007: 42008: 1102009: 762010: 912011: 1232012: 3462013: 6402014: 5452015: 8242016: 6922017: 3792018: 3362019: 4972020: 5472021: 4902022: 3872023: 3482024: 2322025: 261
2007high 824 (2015) · low 4 (2007)2025

MAUDE reports naming product code LXB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—04
20082—0110
20092—076
20102—091
201151890123
20121—5346
201351030640
20140—0545
20152—0824
201641100692
20171—0379
20181—1336
20191—0497
20200—0547
20214790490
20220—0387
20230—0348
20240—0232
20252—0261
202641640—

What is it indicated for?

The Cochlear Baha® SoundArc is intended for test situations and for patients who cannot or choose not to have an implant for the following indications for use: • Patients of any age who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3kHz) should be better than or equal to 45 dB HL for use with the BP100, Baha 4 and Baha 5 sound…

Read the full statement (265 words)

The Cochlear Baha® SoundArc is intended for test situations and for patients who cannot or choose not to have an implant for the following indications for use:

• Patients of any age who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3kHz) should be better than or equal to 45 dB HL for use with the BP100, Baha 4 and Baha 5 sound processors, 55 dB HL for use with the BP110 Power and Baha 5 Power sound processors, and better than or equal to 65 dB HL for use with the Cordelle II and Baha 5 SuperPower Sound Processors.

• Bilateral fitting is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conductive thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1,2, and 3 kHz), or less than a 15dB difference at individual frequencies.

• Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single-sided deafness: SSDTM). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.

• Baha for SSD is also indicated for any patient who is indicated for an air conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171088: Cochlear Baha SoundArc.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171088

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026