Cochlear Baha BP100

K090720510(k) TraditionalCleared Jun 17, 200991-day reviewClass II21 CFR 874.3302

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cochlear Bone Anchored Systems AB · LXB Hearing Aid, Bone Conduction

Cleared on Jun 17, 2009 as substantially equivalent to K042017 (Entific Medical Systems, Inc., 2004) and 6 other predicates after a 91-day review. Its family tree reaches back 5 generations to 1986. 33 later devices descend from it.

Decided
Jun 17, 2009
Review
91 days
Class
Class II
Product code
LXB
Recalls
None
Predicates
7

Its longest line of descent runs back 5 generations to K861971 (1986). 16 ancestors in all. 33 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2004: 1 other ancestor1999: 1 other ancestor1998: 1 other ancestor1994: 1 other ancestor1993: 1 other ancestor2002: 1 other ancestor2001: 1 other ancestor1991: 1 other ancestor1995: 1 other ancestor1987: 1 other ancestor2011: 4 descendants cleared2017: 1 descendant cleared2010: 1 descendant cleared2012: 1 descendant cleared2013: 4 descendants cleared2015: 3 descendants cleared2016: 5 descendants cleared2018: 2 descendants cleared2021: 5 descendants cleared2026: 4 descendants cleared2019: 1 descendant cleared2025: 2 descendants clearedK861971 · Xomed Audiant Bone Conductor W/Improved AA Coil · Xomed, Inc. · 1986K955713 · Nobelpharma Bone Anchored Hearing Aid · Nobelpharma USA, Inc. · 1996×2: 2 generations cleared close together, K984162 (1999), K992872 (1999) K080363 · Baha Cordelle II · Cochlear Bone Anchored Systems AB · 20081990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 6 other ancestors2 generations back: 3 other ancestors3 generations back: 2 other ancestors 1 generation forward: 7 descendants2 generations forward: 5 descendants3 generations forward: 8 descendants4 generations forward: 7 descendants5 generations forward: 5 descendants6 generations forward: 1 descendant K080363 · Baha Cordelle II · Cochlear Bone Anchored Systems AB · 2008K992872 · Branemark Bone-Anchored Hearing Aid (Baha) Cordelle II · Entific Medical Systems, Inc. · 1999K984162 · Branemark Bone-Anchored Hearing Aid (Baha) System for Pediatric Use · Nobelpharma USA, Inc. · 1999K955713 · Nobelpharma Bone Anchored Hearing Aid · Nobelpharma USA, Inc. · 1996K861971 · Xomed Audiant Bone Conductor W/Improved AA Coil · Xomed, Inc. · 1986−50+5
  1. Root · 5 generations backK861971 Xomed Audiant Bone Conductor W/Improved AA CoilXomed, Inc. · 1986
  2. 4 generations backK955713 Nobelpharma Bone Anchored Hearing AidNobelpharma USA, Inc. · 1996
  3. 3 generations backK984162 Branemark Bone-Anchored Hearing Aid (Baha) System for Pediatric UseNobelpharma USA, Inc. · 1999
  4. 2 generations backK992872 Branemark Bone-Anchored Hearing Aid (Baha) Cordelle IIEntific Medical Systems, Inc. · 1999
  5. 1 generation backK080363 Baha Cordelle IICochlear Bone Anchored Systems AB · 2008
  6. This deviceK090720 Cochlear Baha BP100Cochlear Bone Anchored Systems AB · 2009
  7. Descendant · cites it directly · 11 of its ownK132278 Cochlear Baha 4 Sound ProcessorCochlear Americas · 2013
  8. Descendant · cites it directly · 11 of its ownK103594 Ponto Pro PowerOticon Medical AB · 2011
  9. Descendant · 2 generations on · 10 of its ownK142907 Baha 5 Sound ProcessorCochlear Americas · 2015 · Recalled
  10. Descendant · 2 generations on · 10 of its ownK132775 Ponto Plus and Ponto Plus PowerOticon Medical AB · 2013
  11. Descendant · 3 generations on · 6 of its ownK161123 Baha 5 Power Sound ProcessorCochlear Americas · 2016

Showing 5 of 33 descendants. See all descendants

Descendants cleared per year

2010: 12011: 42012: 12013: 42015: 32016: 52017: 12018: 22019: 12021: 52025: 22026: 4
2010high 5 (2016) · low 0 (2014)2026

Most influential descendants

Showing 5 of 33 · See all
  • K132278Cochlear Baha 4 Sound ProcessorCochlear Americas · 2013 · cites it directly11descendants
  • K103594Ponto Pro PowerOticon Medical AB · 2011 · cites it directly11descendants
  • K142907Baha 5 Sound ProcessorCochlear Americas · 2015 · 2 generations on10descendants
  • K132775Ponto Plus and Ponto Plus PowerOticon Medical AB · 2013 · 2 generations on10descendants
  • K161123Baha 5 Power Sound ProcessorCochlear Americas · 2016 · 3 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K861971Xomed Audiant Bone Conductor W/Improved AA CoilXomed, Inc.1986—
−4K955713Nobelpharma Bone Anchored Hearing AidNobelpharma USA, Inc.1996—
−3K984162Branemark Bone-Anchored Hearing Aid (Baha) System for Pediatric UseNobelpharma USA, Inc.1999—
−2K992872Branemark Bone-Anchored Hearing Aid (Baha) Cordelle IIEntific Medical Systems, Inc.1999—
−1K080363Baha Cordelle IICochlear Bone Anchored Systems AB2008—
0 (this device)K090720Cochlear Baha BP100Cochlear Bone Anchored Systems AB2009—
+1K132278Cochlear Baha 4 Sound ProcessorCochlear Americas2013—
+1K103594Ponto Pro PowerOticon Medical AB2011—
+2K142907Baha 5 Sound ProcessorCochlear Americas20151
+2K132775Ponto Plus and Ponto Plus PowerOticon Medical AB2013—
+3K161123Baha 5 Power Sound ProcessorCochlear Americas2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199110
199310
199410
199510
199810
199910
200110
200210
200410
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201140
201210
201340
201530
201650
201710
201820
201910
202150
202520
202640
FIG. 1 — Lineage of K090720, 1986–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LXB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LXB device

  1. #1 K254296 Ossiclear (Ossiclear System)score 14.7 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042017Baha Divino
Summary page 2
…hree currently marketed sound processors in the family of processors available for use with the Baha® auditory osseointegrated-implant; namely, the Baha DivinoTM (510(k) K042017), Baha IntensoTM (510(k) K081606), and Baha Cordelle lI(510(k) K080363). Bench testing shows that the gain/output of the Baha BP1…
Entific Medical Systems, Inc.2004—
*K081606Baha Intenso
Summary page 2
…cessors in the family of processors available for use with the Baha® auditory osseointegrated-implant; namely, the Baha DivinoTM (510(k) K042017), Baha IntensoTM (510(k) K081606), and Baha Cordelle lI(510(k) K080363). Bench testing shows that the gain/output of the Baha BP100 are comparable to those for th…
Cochlear Americas2008—
*K080363Baha Cordelle II
Summary page 2
…ailable for use with the Baha® auditory osseointegrated-implant; namely, the Baha DivinoTM (510(k) K042017), Baha IntensoTM (510(k) K081606), and Baha Cordelle lI(510(k) K080363). Bench testing shows that the gain/output of the Baha BP100 are comparable to those for the marketed Baha Divino sound processor…
Cochlear Bone Anchored Systems AB2008—
*K973880Sonix Hearing Aid Fitting and Programming System
Summary page 2
…nded use and works in the same manner as that of many programmable hearing aids and devices. Predicate devices proposed for the fitting software system are 510(k) number K973880 for the Sonix hearing aid fitting and programming system, 510(k) number K93 1372 for the Resound® portable prescriptive programmi…
Sonix Technologies, Inc.1998—
*K931372Resound Portable Prescriptive Programming System
Summary page 2
…and devices. Predicate devices proposed for the fitting software system are 510(k) number K973880 for the Sonix hearing aid fitting and programming system, 510(k) number K93 1372 for the Resound® portable prescriptive programming system, and 510(k) number K942749 for the HI-PRO/NOAH universal hearing instru…
Resound Corp.1993—
*K942749HI-Pro/Noah Universal Hearing Instrument
Summary page 2
…ber K973880 for the Sonix hearing aid fitting and programming system, 510(k) number K93 1372 for the Resound® portable prescriptive programming system, and 510(k) number K942749 for the HI-PRO/NOAH universal hearing instrument by Madsen Electronics. Also unique to this new Baha sound processor is the abil…
Madsen Electronics (Canada), Ltd.1994—
*K984547Digital Lgob Procedure Using Resound, Digital 5000 Series Hearing Aids
Summary page 2
…e predicate for obtaining data to assist in more accurate fittings by obtaining measurements for pure tones produced by the hearing device can be found in 5 10(k) number K984547 for the LGOB test using the Resound® digital 5000 series hearing devices. Although that approach uses suprathreshold measures to…
Resound Corp.1999—

Also cleared against K042017: 4 other devices →

What descends from it? (33)

See all 33

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132278Cochlear Baha 4 Sound ProcessorCochlear Americas201311
K103594Ponto Pro PowerOticon Medical AB201111
K100193Otomag Bone Conduction Hearing SystemSophono, Inc.20105
K121228Ponto Bone Anchored Hearing SystemOticon Medical AB20124
K100649Soundbite Hearing System by SonitusSonitus Medical, Inc.20111
K171088Cochlear Baha SoundArcCochlear Americas20170
K110831Soundbite Hearing System by Sonitus MedicalSonitus Medical, Inc.20110

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 2.9 years on median.
  • None of the 16 ancestors has a recall.
  • Of its descendants, K132278 (Cochlear Americas, 2013) has gone furthest, with 11 descendants of its own.
  • 4 other devices also cite its predicate K042017.

How does its category compare?

All LXB devices

LXB Hearing Aid, Bone Conduction: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LXB

2in 2025

2008: 22009: 22010: 22011: 52012: 12013: 52015: 22016: 42017: 12018: 12019: 12021: 42025: 22026 (partial year): 4
2007high 5 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LXB

79in 2021

2011: 189 days median2013: 103 days median2016: 110 days median2021: 79 days median2026 (partial year): 164 days median
2007high 189 (2011) · low 79 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LXB

1in 2018

2012: 52018: 1
2007high 5 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LXB

261in 2025

2007: 42008: 1102009: 762010: 912011: 1232012: 3462013: 6402014: 5452015: 8242016: 6922017: 3792018: 3362019: 4972020: 5472021: 4902022: 3872023: 3482024: 2322025: 261
2007high 824 (2015) · low 4 (2007)2025

MAUDE reports naming product code LXB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—04
20082—0110
20092—076
20102—091
201151890123
20121—5346
201351030640
20140—0545
20152—0824
201641100692
20171—0379
20181—1336
20191—0497
20200—0547
20214790490
20220—0387
20230—0348
20240—0232
20252—0261
202641640—

What is it indicated for?

The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications: * Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz)…

Read the full statement (232 words)

The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications:

* Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.

* Bilateral fitting of the BP100 is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.

Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to .20 dB HL.

Baha for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K090720: Cochlear Baha BP100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K090720

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026